| Literature DB >> 33387130 |
Robert G Hauser1,2, Susan A Casey3, Christopher B Gitter3, Chuen Y Tang3, Raed H Abdelhadi3, Charles C Gornick3, Larissa Stanberry3, Jay D Sengupta3.
Abstract
PURPOSE: We hypothesized that data in manufacturers' product performance reports (PPRs) can provide clinically valuable ICD and cardiac resynchronization defibrillator (CRT-D) reliability and longevity information.Entities:
Keywords: Battery; Cardiac resynchronization; Implantable defibrillator; Longevity; Malfunction; Reliability
Mesh:
Year: 2021 PMID: 33387130 PMCID: PMC8645539 DOI: 10.1007/s10840-020-00920-w
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.900
Study population
| Manufacturer, models, battery type (A-h capacity) | # Pulse generators (%) |
|---|---|
| CRT-D | 493,478 (43) |
| Abbott | 192,510 (39) |
Quadra Assura, Unify Assura, Unify Quadra, Unify LiSVO/CFx (1.94 A-h) | |
| Biotronik* | 13,898 (3) |
Iperia 7, Ilivia 7, Intica 7, Itrevia 7, Ilesto 7 LiSVO/CFx (1.6 A-h) | |
| Boston Scientific | 100,617 (20) |
Resonate, Charisma, Momentum, Vigilant, Dynagen, Inogen, Origen, Incepta, Energen, Punctua LiMnO2 (1.9 A-h) | |
| Medtronic | 186,453 (38) |
Viva, Brava, Amplia, Claria, Compia LiSVO/CFx (1.0–1.2 A-h) | |
| DR-ICD | 384,183 (33) |
| Abbott | 123,370 (32) |
Ellipse, Fortify LiSVO/CFx (1.75–1.94 A-h) | |
| Biotronik | 34,528 (9) |
Ilesto 7, Iforia 7, Ilivia 7, Intica 7, Itrevia 7, Iperia 7, Lumax 740, Inventra 7 DX LiSVO/CFx or LiMnO2 (1.39–1.6 A-h) | |
| Boston Scientific | 77,170 (20) |
Resonate, Charisma, Momentum, Vigilant, Dynagen, Inogen, Origin, Incepta, Energen, Punctua LiMnO2 (1.8–1.9 A-h) Dynagen, Inogen, Origen Mini LiMnO2 (1.0 A-h) | |
| Medtronic | 149,115 (39) |
Evera, Primo, Mirro LiSVO/CFx (1.0–1.2 A-h) | |
| VR-ICD | 272,142 (24) |
| Abbott | 87,440 (32) |
Ellipse, Fortify LiSVO/CFx (1.75–1.94 A-h) | |
| Biotronik | 4453 (2) |
Ilesto 7, Itrevia 7, Lumax 740 LiSVO/CFx (1.39–1.6 A-h) | |
| Boston Scientific | 91,949 (34) |
Resonate, Charisma, Momentum, Vigilant, Dynagen, Inogen, Origen, Incepta LiMnO2 (1.8–1.9 A-h) Emblem, SQ-Rx LiMnO2 (5.67W) Dynagen, Inogen, Origin Mini LiMnO2 (1.0 A) | |
| Medtronic | 88,400 (32) |
Evera, Visia AF, Primo, Mirro LiSVO/CFx (1.0–1.2 A) | |
Reported as numbers and percent
A-h capacity, amperes per hour; CERT-D, cardiac resynchronization therapy defibrillator; DR-ICD, dual-chamber implantable cardioverter defibrillator; VR-ICD, single-chamber implantable cardioverter defibrillator; LiMnO2, lithium manganese dioxide; LiSVO/CFx, lithium silver vanadium oxide/carbon monofluoride
Fig. 1Reliability of CRT-D, DR-ICD, and VR-ICD pulse generators by manufacturer. Note: Biotronik plot is under the Medtronic plot
Fig. 2a Impact of malfunctions on therapy availability. b Causes of malfunctions that compromised therapy
Fig. 3Cumulative Abbott versus Medtronic battery malfunctions. The Abbott batteries were subject to an advisory in October 2016 (arrow) that alerted physicians to the potential risk of rapid battery depletion
Fig. 4Normal battery longevity of CRT-D, DR-ICD, and VR-ICD pulse generators by the manufacturer
Fig. 5Performance of CRT-D, DR-ICD, and VR-ICD pulse generators by the manufacturer
Fig. 6Comparison of subcutaneous and transvenous implantable cardioverter defibrillators
Fig. 7Longevity of lithium manganese dioxide and hybrid lithium silver vanadium oxide/carbon monofluoride batteries (NEW) compared to lithium silver vanadium oxide batteries (OLD). Data were obtained from 2019 Abbott, Biotronik, Boston Scientific, and Medtronic product performance reports