Li Peng1, Jing Zhao2, Zu-Hua Wang3, Yu-Chun Sun4, Yu-Hong Liang2,3.
Abstract
BACKGROUND/Entities:
Keywords: 3D printing; Accuracy; Digital guide; Endodontic surgery; Root-end resection
Year: 2020 PMID: 33384777 PMCID: PMC7770308 DOI: 10.1016/j.jds.2020.06.024
Source DB: PubMed Journal: J Dent Sci ISSN: 1991-7902 Impact factor: 2.080
Figure. 1Virtual model of the digital guide for endodontic surgery. The red arrow indicates the location of the axial plane 3 mm from the apex.
Figure. 2Guided surgical procedure. (A, B) The experimental scene to mimic the clinical environment. (C, D) Initial guide placement. (E) Exposure of the periapical lesion. (F) Removal of the periapical lesion. (G) The depth stopper (red arrow) on the bur. (H, I) Guided root end removal. (J) The tooth after surgery.
Differences in length/angle deviation of root-end resection and operation time between groups. The operation time indicates that consumed in procedures including root-end resection and osteotomy.
| Length deviation (mm) | Angle deviation (°) | Operation time (s) | ||||
|---|---|---|---|---|---|---|
| Guide | No guide | Guide | No guide | Guide | No guide | |
| Experienced | 0.31 ± 0.19a | 0.99 ± 0.58b | 5.04 ± 2.99a | 16.74 ± 10.61b | 155.71 ± 75.13 | 187.67 ± 77.65 |
| Inexperienced | 0.31 ± 0.12a | 1.18 ± 1.18b | 6.79 ± 3.25a | 15.06 ± 10.17b | 189.75 ± 85.08 | 260.08 ± 153.68 |
a, b: The difference between the groups was significant (P < 0.01).
Figure. 3Differences in length deviation of root-end resection performed with/without guides. The negative value indicates the length of root-end resection is less than 3 mm, otherwise greater than 3 mm. Ex, the experienced; In, the inexperienced. There were no significant differences between the groups (P > 0.05).