| Literature DB >> 33380619 |
PremNandhini Satgunam1, Monika Thakur1, Virender Sachdeva2, Sneha Reddy3, Padmaja Kumari Rani4.
Abstract
PURPOSE: The objective of this study was to identify and validate smartphone-based visual acuity (VA) apps that can be used in a teleophthalmology portal.Entities:
Keywords: Smartphone; tele-ophthalmology; tele-optometry; vision testing; visual acuity Apps
Mesh:
Year: 2021 PMID: 33380619 PMCID: PMC7933864 DOI: 10.4103/ijo.IJO_2333_20
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1Study flowchart
Figure 2Examiner with personal protective equipments testing the distance visual acuity with visual acuity App for a participant
Overall Median logMAR visual acuity with range (minimum, maximum) for COMPlog, SmartOptometry, Peek Acuity and Reduced Snellen near vision chart
| Participants | Median age [Min, Max] | Charts | Median Visual Acuity [Min, Max] | |
|---|---|---|---|---|
| All ( | 31 [20, 60] | COMPlog | 0.01 [0.00,1.09] | 0.31 |
| Peek Acuity | 0.00 [0.00,1.30] | |||
| SmartOptometry | 0.00 [0.00,1.00] | 0.002* | ||
| Reduced Snellen near vision | 0.00 [0.00,0.80] | |||
| Young ( | 27 [20, 36] | COMPlog | 0.00 [0.00,1.09] | 0.98 |
| Peek Acuity | 0.00 [0.00,1.30] | |||
| SmartOptometry | 0.00 [0.00,1.00] | 0.004* | ||
| Reduced Snellen near vision | 0.00 [0.00,0.78] | |||
| Presbyopia ( | 50 [37, 60] | COMPlog | 0.10 [0.00,0.83] | 0.17 |
| Peek Acuity | 0.15 [0.00,1.30] | |||
| SmartOptometry | 0.39 [0.00,0.60] | 0.007 | ||
| Reduced Snellen near vision | 0.30 [0.00,0.80] | |||
| Myopia ( | 29 [20, 56] | COMPlog | 0.36 [0.00,1.09] | 0.90 |
| Peek Acuity | 0.00 [0.00,1.30] | |||
| SmartOptometry | 0.30 [0.00,1.00] | 0.74 | ||
| Reduced Snellen near vision | 0.00 [0.00,0.80] | |||
| Hyperopia ( | 50 [22, 60] | COMPlog | 0.06 [0.00,0.58] | 0.74 |
| Peek Acuity | 0.10 [0.00,0.60] | |||
| SmartOptometry | 0.40 [0.00,0.60] | 0.11 | ||
| Reduced Snellen near vision | 0.30 [0.00,0.78] | |||
| Emmetropia ( | 29 [21, 47] | COMPlog | 0.00 [0.00,0.04] | 0.71 |
| Peek Acuity | 0.00 [0.00,0.10] | |||
| SmartOptometry | 0.00 [0.00,0.40] | 0.04* | ||
| Reduced Snellen near vision | 0.00 [0.00,0.18] |
*indicates statistically significant P-values (P < 0.05)
Figure 3Bland-Altman plot for limits of agreement between visual acuity tested with COMPlog and Peek Acuity (a) and between visual acuity tested with SmartOptometry and Reduced Snellen near vision chart (b)
Mean difference ± standard deviation (SD) of visual acuity measured on two different days with COMPlog, Peek Acuity and SmartOptometry
| Charts | Mean difference ± SD | 95% CI of mean difference | 95% CI of limit of agreement | Coefficient of repeatability |
|---|---|---|---|---|
| COMPlog | -0.006 ± 0.02 | -0.03 to -0.01 | -0.04 to 0.03 | 0.04 |
| Peek Acuity | -0.01 ± 0.03 | -0.02 to 0.08 | -0.07 to 0.05 | 0.03 |
| SmartOptometry | 0.03 ± 0.07 | -0.02 to 0.01 | -0.10 to 0.16 | 0.14 |
Reduced Snellen near vision chart is not tabulated since all the participants had 0 logMAR acuity on both the days and their Coefficient of repeatability was 0 (i.e., 100% repeatable)