Literature DB >> 33380588

A narrative review of hydrogen-oxygen mixture for medical purpose and the inhaler thereof.

Hsin-Yung Lin1, Pei-Chen Lai1, Wei-Lin Chen1.   

Abstract

Recent development regarding mixture of H2 (concentration of ~66%) with O2 (concentration of ~34%) for medical purpose, such as treatment of coronavirus disease-19 (COVID-19) patients, is introduced. Furthermore, the design principles of a hydrogen inhaler which generates mixture of hydrogen (~66%) with oxygen (~34%) for medical purpose are proposed. With the installation of the liquid blocking module and flame arresters, the air pathway of the hydrogen inhaler is divided by multiple isolation zones to prevent any unexpected explosion propagating from one zone to the other. An integrated filtering/cycling module is utilized to purify the impurity, and cool down the temperature of the electrolytic module to reduce the risk of the explosion. Moreover, a nebulizer is provided to selectively atomize the water into vapor which is then mixed with the filtered hydrogen-oxygen mix gas, such that the static electricity of a substance hardly occurs to reduce the risk of the explosion. Furthermore, hydrogen concentration detector is installed to reduce the risk of hydrogen leakage. Result shows that the hydrogen inhaler implementing the aforesaid design rules could effectively inhibit the explosion, even ignition at the outset of the hydrogen inhaler which outputs hydrogen-oxygen gas (approximately 66% hydrogen: 34% oxygen).

Entities:  

Keywords:  COVID-19; anti-explosion; cancer; hydrogen; hydrogen concentration; hydrogen inhaler; medical purpose; nebulizer; respiratory disease; static electricity

Mesh:

Substances:

Year:  2020        PMID: 33380588      PMCID: PMC8092144          DOI: 10.4103/2045-9912.295226

Source DB:  PubMed          Journal:  Med Gas Res        ISSN: 2045-9912


INTRODUCTION OF HYDROGEN BIOMEDICINE

Hydrogen (H2) has been utilized in diving medicine for many years, and it is proved that hydrogen, unlike other medical gases such as nitric oxide, carbon monoxide, or hydrogen sulfide, is non-toxic for human body even at high pressure or concentration.12 However, hydrogen’s therapeutic effects against diseases were unknown or ignored in the past. Until 1975, hydrogen’s inhibitory effect against squamous cell carcinoma of mouse skin was reported by Dole et al.3 Roberts et al.4 also disclosed hydrogen’s prohibition against the growth of mouse tumor cells and leukemia cells in 1978. Especially, in 2007 Ohsawa et al.5 discovered that hydrogen is capable of selectively detoxifying hydroxyl radicals (• OH) and peroxynitrite (ONOO–) among active oxygens (such selectively detoxifying ability was further discussed by Liu et al.6 in 2010). Thereafter, the study and research of hydrogen biomedicine has been rapidly grown, including not limited to alleviation of inflammation,789 alleviation of cell apoptosis,1011 alleviation of side effects caused by anti-cancer therapy,1213 and cancer inhibition.14 Due to lack of safe and durable hydrogen inhalers which can generate inhaled gas with higher H2 concentration, conventionally hydrogen related animal studies or clinical trials were performed based on either inhaled gas with H2 concentration of less than 4%, or hydrogen-rich water with hydrogen concentration far less than 4%,15161718 nevertheless, the H2 concentration of less than 4% is not necessary for the optimum value in the human body. Especially, Graham’s Law of diffusion states that the rate of diffusion of one gas is inversely proportional to the square root of the density of the gas19 and the Bernoulli principle also states that an increase in the speed of a fluid occurs simultaneously with a decrease in pressure.20 The higher the H2 concentration in the inhaled gas, the lower the density of the inhaled gas. Therefore, the inhaled gas with higher H2 concentration then has the greater diffusion speed and lower pressure. Thus, it will reduce the driving pressure for human body to breathe the inhaled gas with higher H2 concentration. With the support of hydrogen inhaler AMS-H-01/AMS-H-03 provided by Asclepius Meditec Co., Ltd., tens of animal studies or pilot clinical trials based on inhaled gas which combines H2 (concentration of ~66%) with O2 (concentration of ~34%) were accomplished and published.1421222324252627282930313233343536373839404142434445464748495051525354 Additionally, recent formal clinic trials based on the same inhaler AMS-H-01/AMS-H-03 also revealed positive results in respiratory diseases, such as allergic rhinitis, asthma, chronic obstructive pulmonary disease, etc. Especially, it is recommended to use the mixed gas of 66.6% H2/33.3% O2 for treating coronavirus disease-19 (COVID-19) patients as described in the Diagnosis and Treatment Protocol for Novel Coronavirus Pneumonia, 7 Edition released by National Health Commission of China,55 and such guideline is also republished in English by the World Health Organization China Office.56 Moreover, a survey report regarding 82 cancer patients with the help of the same inhaler AMS-H-01 also demonstrated that in patients with advanced cancer, inhaling gas with H2 concentration of 64–67% can improve patients’ quality of life and control cancer progression.53 Therefore, how to provide safe and durable hydrogen-oxygen inhalers with higher H2 concentration is key factor to expedite and realize the hydrogen biomedicine. Search strategy is shown in .

BASIS OF HYDROGEN INHALER

The currently available hydrogen inhalers can be classified as ion exchange membrane type or electrolytic solution type. shows an ion exchange membrane-based hydrogen inhaler which conventionally uses a platinum-coated ion exchange membrane to separate the electrolysis tank into an anode chamber and a cathode chamber. When the anode and cathode electrodes of the electrolysis tank are supplied with electrical power, oxygen is generated in the anode chamber, and the extra protons (H+) are transported from the anode chamber to the cathode chamber through the ion exchange membrane and combine with electrons on the cathode to form hydrogen therein. Since the ion exchange membrane is structured to separate the electrolytic tank, theoretically almost 100% hydrogen gas could be obtained from the cathode chamber. Ion exchange membrane-based hydrogen inhaler. However, 100% hydrogen gas is not suitable for direct inhalation of human body and shall be mixed with air.57 It is considered as oxygen-deficient gas when the O2 concentration of the inhaled gas is slower than 19.5%, and it is dangerous for human begins when the O2 concentration of the inhaled gas is slower than 16%.58 Mixing 100% hydrogen gas from the ion exchange membrane-based hydrogen inhaler with air for inhaling will inevitably dilute the original O2 concentration (~21%) in the air and likely poses negative impact to subject users. For example, a human body consumes approximately air of 6 L/min. Supposedly the ion exchange membrane-based hydrogen inhaler outputs 100% hydrogen gas of 1 L/min and only half of them (500 mL) is inhaled and mixed with the 5.5 L of air for human body, the oxygen concentration of the mixed gas (air of 5.5 L + 100% hydrogen gas of 500 mL) is approximately reduced to 19.25%. Therefore, it is not recommended to inhale 100% hydrogen gas for a long period of time. Moreover, deionized water is required for ion exchange membrane-based hydrogen inhaler to maintain the normal operation; otherwise impurities in regular water would block ion channels of the ion exchange membrane and then the function thereof fails. Additionally, the lifetime of such hydrogen inhaler is shorter due to the inevitable decay of the platinum catalyst coated on the ion exchange membrane. Furthermore, ion exchange membrane is a polymer membrane (such as perfluorosulfonic acid proton exchange membrane) mainly used in fuel cells and may not be suitable for medical purpose.59 To be worse, fluoride could be likely released from the perfluorosulfonic acid proton exchange membrane during operation.60 The traditional electrolytic solution-based hydrogen inhaler is illustrated as follows. As shown in , two electrodes of the electrolytic tank undergo an electrochemical reaction against water to produce oxygen and hydrogen which are then mixed within the electrolytic tank. The H2 concentration in the hydrogen-oxygen mixed gas is approximately equal to 66%. Unlike the ion exchange membrane-based hydrogen inhaler, deionized water is not necessary for the operation of electrolytic solution-based hydrogen inhaler, but it is recommended to use regular pure water or distilled water. Electrolytic solution-based hydrogen inhaler. Although the electrolytic solution-based hydrogen inhaler is a simple structure, electrolyte is required in the electrolytic solution to enhance electrolysis efficiency, and alkali mist generated during high temperature may irritate the human body if such alkali mist is not properly filtered and isolated. Moreover, heat released during the electrolysis process unavoidably raises the temperature of the electrolytic tank, and it is necessary to intermittently shut down the traditional electrolytic solution-based hydrogen inhaler for safety purpose. Furthermore, there exists risk of detonation due to the hydrogen (~66%)-oxygen (~34%) mixed gas in the event no suitable anti-explosion devices are designed in the solution-based hydrogen inhaler.

DESIGN OF THE HYDROGEN INHALER FOR MEDICAL PURPOSE

Medical-grade hydrogen inhaler is the prerequisite for the development of hydrogen biomedicine. It should be safe, durable, and can stably generate higher flow rate (e.g., over 2 L/min) and higher hydrogen concentration (e.g., over 50%) over a long period (e.g., at least over 8 hours without any interrupt of operation). The lifetime of the hydrogen inhaler shall be more than 5000 operating hours, and no additional electrolyte shall be supplemented during the above lifetime of the solution-based hydrogen inhaler. Furthermore, all materials of the hydrogen inhaler used in contact with gas/liquid should be medical-grade and the hydrogen inhaler itself should be made in compliance with all international standards of medical device. Additionally, all impurity generated during the electrolysis process shall be filtered and isolated, ensuring that the outputted hydrogen-oxygen mixture or hydrogen meets the specifications of medical respiratory gas. The examples of available standards are shown in . Examples of available standards for medical device The following are the proposed design rules of a safe and durable hydrogen inhaler for medical purpose:

Electrolytic module and filtering module

As shown in , the electrolytic module is the kernel for the hydrogen inhaler, it will either generates almost 100% hydrogen gas (ion exchange membrane type) or hydrogen-oxygen mixed gas (H2 ~66%: O2 ~34%, electrolytic solution type) which is then combined with air for the breath of human body. It is important to reduce the space of the gas chamber in the electrolytic module, such that only few residuals of hydrogen gas or hydrogen-oxygen mixed gas would be kept in the electrolytic module to reduce the risk or impact of the unexpected explosion. Proposed structure of safe and durable hydrogen inhaler for medical purpose. Furthermore, filtering module shall be installed to purify the hydrogen gas or hydrogen/oxygen mixed gas in order to comply with available medical gas standards. Additionally, double or triple purification processes are recommended to filter various potential impurities, such as fluoride or electrolyte.

Cycling module

During electrolytic process, heat will be generated and eventually raises the temperature of the electrolytic module. Such continuously raised temperature could shorten the lifetime of the electrolytic module, breaks sealing of the electrolytic module and then causes the leakage of the liquid/gas therein, and likely triggers the explosion. Therefore, a cycling module to circulate and cool down the liquid in the electrolytic module is critical to maintain the lifetime and safety, as shown in . A temperature sensor coupled to the electrolytic module can monitor the temperature of the electrolytic module and activates the cycling module whenever necessary. Moreover, when the cycling module is utilized in the electrolytic solution type hydrogen inhaler, it could also circulate water and flush residual electrolyte in the air pathway of the hydrogen inhaler back to the electrolytic module. Flushing the residual electrolyte back to the electrolytic module could recycle the electrolyte and extend the operation hours of the hydrogen inhaler.

Liquid blocking module

A liquid blocking module in the hydrogen inhaler is an anti-explosion device. It will isolate the electrolytic module from the output of the liquid blocking module; therefore, to block the subject user from any impact of unexpected explosion happened in the electrolytic module. On the other hand, the liquid blocking module prohibits any accidental explosion propagating from the output of the liquid blocking module into the electrolytic module, and therefore to prevent any damage to the electrolytic module caused by any ignition happening outside the liquid blocking module. Thus, a first isolation zone is formed between the electrolytic module and the liquid blocking module, and any unexpected explosion will be restrained therein (). Besides, the liquid blocking module could lower the temperature of the hydrogen gas or hydrogen-oxygen mixed gas generated from the hot zone of the electrolytic module, and further dissolve potential impurity of the generated gas in the liquid as well.

Nebulizer

In the conventional hydrogen inhaler, a carbon dioxide absorbent is common used to partially remove the vapor in the hydrogen gas or hydrogen-oxygen mix gas. As a result, the mucus layer of the respiratory organ could not be kept in a normal state, and the respiratory airway could not be protected by the mucus layer. Therefore, it is necessary to provide appropriate humidity in the hydrogen gas or hydrogen-oxygen mixed gas for the breath of subject users. The nebulizer is provided to selectively atomize the water into vapor at the nebulizing chamber and the vapor will be mixed with the hydrogen gas or hydrogen-oxygen mixed gas in the mixing chamber, as shown in . Furthermore, the nebulizer could atomize suitable medicine for breath as well to assist treatment of human body. Moreover, when the humidity increases, the electrical resistance of a substance (gas or the subject user) decreases and static electricity hardly occurs. The disappearing time during which static electricity is eliminated also decreases exponentially as the absolute humidity increases.61 Therefore, the higher the relative humidity, the less likely static electricity causes explosion. With the help of vapor combined with the hydrogen gas or hydrogen/oxygen mix gas, the static electricity of a substance (gas or the subject user) hardly occurs, and the explosion incurred by the static electricity could be prevented.

Flame arresters

It is known that flame will be extinguished if the dimension of the air pathway is 0.20 mm or less at any hydrogen gas concentration.62 Based on the 0.20 mm air pathway criteria, a flame arrester made of stainless steel meshes which are randomly compressed and overlapped together can set up pathway the dimension of which is less than 0.2 mm. Therefore, the flame arrester is also an anti-explosion device to extinguish flame and to prevent explosion from propagating. The flame arrester could be installed at the input of the nebulizer such that a second isolation zone is formed between the nebulizer and the liquid blocking module, and any unexpected explosion in the second isolation zone will be confined therein. For the convenience of breath of the subject user, a comfortable nasal tube is necessary. An additional flame arrester could be installed in the nasal tube such that a third isolation zone is formed between the nasal tube and the nebulizer, and again any unexpected explosion in the third isolation zone will be confined therein. Moreover, such flame arrester is preferred to be installed closed to the output of the nasal tube to reduce the space between such flame arrester and the outlet of the nasal tube (e.g., within the range of 1–2 cm3). In the event any static electricity generated from the subject user or intentional ignition happens at the outlet of the nasal tube, the small space of 1–2 cm3 hardly causes explosion, and therefore prevent damage to the subject user.

Hydrogen concentration detection

It is well discussed that the H2 concentration of the mixed gas higher than 4% may cause explosion. Nevertheless, even the hydrogen inhaler outputs the hydrogen-oxygen mixed gas and the H2 concentration of which is around 66%, H2 concentration around the subject user breathing the hydrogen-oxygen mixed gas is less than 1% (as discussed later). Thus, during normal operation it is almost unlikely to incur explosion even there is unexpected flame or ignition close to the subject user. However, in the event there is any abnormal situation happened during the operation of the hydrogen inhaler, such as leakage in the hydrogen inhaler, H2 concentration around the subject user may be higher than 4% and there still exits risk of detonation. Therefore, a detector could be installed to monitor the H2 concentration and the hydrogen inhaler should be immediately shut down if the detected H2 concentration is approaching 4%.

RESULTS AND DISCUSSION

Pursuant to the aforesaid design principles, AMS-H-03 hydrogen inhaler is developed and built. It provides inhaled gas with H2 concentration of ~66% and O2 concentration of ~34%, at output rate of 2–3 L/min. Since O2 concentration of the inhaled gas is higher than 21%, any combination of the inhaled gas and air for the breath of human body will not be considered as oxygen-deficient gas. During the operation, the electrolytic module of the AMS-H-03 hydrogen inhaler generates hydrogen-oxygen mixed gas (H2 ~66%: O2 ~34%, electrolytic solution type) which then passes through the filtering module and the liquid blocking module for purification. One additional filter is installed after the liquid blocking module to further percolate the hydrogen-oxygen mixed gas. Such triple purification (including the filtering module, the liquid blocking module, and the other filter) will assure that the outputted hydrogen-oxygen mixed gas is complied with available medical gas standards. When the temperature of the electrolytic module raises due to the heat generated by the electrolytic process, one cycling module of AMS-H-03 then activated to circulate and cool down the liquid in the electrolytic module. Furthermore, in the event the water in the electrolytic module is consumed or exhausted due to the electrolytic process, another cycling module of AMS-H-03 will supply water into the electrolytic module through the filtering module, and the residual electrolyte originally blocked by the filter module will then be flushed back to the electrolytic module simultaneously. Flushing the residual electrolyte back to the electrolytic module could recycle the electrolyte and extend the operation hours of the hydrogen inhaler. The gas pathway of AMS-H-03, beginning from the water surface of the electrolytic module to the outlet of the nasal tube, is designed as short as possible. Moreover, such gas pathway is further divided into three isolation zones, one is between the electrolytic module and the liquid blocking module, another is between the liquid blocking module and a flame arrester coupled to the nebulizer, and the other is between the flame arrester coupled to the nebulizer and anther flame arrester close to the outlet of the nasal tube worn by a user. With the installation of the liquid blocking module and flame arresters, any unexpected explosion generated in one isolation zone will be prevented from propagating to the other isolation zone. Moreover, the nebulizer is provided to selectively atomize the water into vapor which is then mixed with the filtered hydrogen/oxygen mix gas, not only to maintain appropriate humidity in the hydrogen gas or hydrogen/oxygen mix gas for the respiratory system of subject users, but also to reduce the risk of explosion incurred by the static electricity. A multicenter randomized clinical trial that verifies the efficacy and safety regarding inhalation of H2 (~66%) and O2 (~34%) in patients with COVID-19 was reported based on utility of hundreds of AMS-H-03 hydrogen inhalers. It is found that inhalation of H2 (~66%) and O2 (~34%) significantly improves disease severity of patients with COVID-19, such as improvement of chest distress, chest pain, resting oxygen saturation, and dyspnea scale, etc.63 As previously mentioned, the inhaled gas with higher H2 concentration has the greater diffusion speed and lower pressure. Thus, mixture of H2 (~66%) and O2 (~34%) could decrease the inspiratory efforts due to the significantly lower resistance when passing through the respiratory tract,63 and could be an option for relieving respiratory symptoms in patients with COVID-19.

Acid/alkaline measurement and hydrogen concentration measurement

The inhaled gas of the AMS-H-01/AMS-H-03 is subject to the tests set forth in GB8982-2009 (CN) regarding standards of medical gas and aviation gas. With the help of filtering module and cycling module, the inhaled gas of the AMS-H-01/AMS-H-03 passes the section 5.4 of GB8982-2009 (CN) regarding the residual amount of acid material and alkaline material. Therefore, the alkali mist is totally removed from the inhaled gas of the AMS-H-01/AMS-H-03. As shown in , during the normal operation, even H2 concentration at the output of the nebulizer and nasal tube is around 64–67%, the H2 concentration around the subject user is less than 1% and the H2 concentration of exhaled breath of the subject user is less than 4%, both of which are less than the critical explosive 10% H2 concentration in the mixture of hydrogen and air (or oxygen).57646566 Therefore, during normal operation, in the event there is unexpected flame or ignition close to the subject user, it is almost unlikely to incur explosion or detonation due to the fact that the H2 concentration surrounding the subject user is less than 4%. H2 concentration measured at different locations Note: Furthermore, H2 concentration is measured by New Cosmos Electric: XP-3140 (Osaka, Japan). First, when the AMS-H-01 is tuned on, the hydrogen concentration at: (1) ambient environment which is one meter away from the AMS-H-01 (“VA”); (2) output of the nebulizer (“OA”); and (3) output of special nasal tube or nasal mask connected to the nebulizer (“ONT”) are measured separately. Secondly, when each subject user is equipped with the nasal tube, H2 concentration around the subject user is detected at the location of 1 cm away from: (1) right ear (“A”); (2) left ear (“B”); (3) top of head (“C”); and (4) back of head (“D”) of the subject user after 5-minute, 60-minute, and 120-minute operation of the AMS-H-01 respectively. Last, the exhaled breath of the subject was collected using a gas sample collection bag immediately after the nasal tube was removed from the subject user, and then the hydrogen concentration of the exhaled breath was measured.

Ignition experiments

The ignition experiments at flow rate of 3 L/min, 2.5 L/min, and 2 L/min of hydrogen-oxygen gas (approximately 66% H2: 34% O2) generated from the AMS-H-03 are performed. The AMS-H-03 is ignited at output of nasal tube, as shown in . Ignited at output of nasal tube of AMS-H-03. As shown in the following , it is found that only clean ignition noises could be detected for three different flow rates. The results are reasonable since, as previously discussed, the volume between the flame arrester in the nasal tube and the outlet of the nasal tube is quite small, almost no explosive sound is detected when ignition is made at the output of the nasal tube. Therefore, during the operation of AMS-H-03, it is likely that the user who wears the nasal tube for hydrogen inhaling would not be subject to damage even there is ignition caused by static electricity of human body. Ignition results for three different flow rates

Conclusion

Hydrogen-oxygen mixture (approximately 66% H2: 34% O2) for medical purpose is an oncoming and developing research topic. A safe and durable hydrogen inhaler for generating such hydrogen-oxygen mixture is proposed based on reliable design principles. Experiment and measurement result shows that the hydrogen inhaler implementing the proposed design rules could effectively reduce the explosion, and that the hydrogen-oxygen mixture generated from the hydrogen inhaler is suitable for medical purpose.

Additional file

: Additional file 1.
Table 1

Examples of available standards for medical device

International standard numberTitle of standard
IEC 60601-1:2005 Medical electrical equipment—Part 1: General requirements for safety
IEC 60601-1-11:2015 Medical electrical equipment – Part 1–11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-1-2:2007 Medical electrical equipment-Part 1–2: General requirements for safety-Collateral standards: Electromagnetic compatibility-Requirements and tests
ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications-Parts 2: Tests for emissions of particulate matter
ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications-Parts 3: Tests for emissions of volatile organic compounds
ISO 18562-4:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications-Parts 4: Tests for leachable in condensate
ISO 10993-5:2016 Biological evaluation of medical devices—Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2009 Biological evaluation of medical devices—Part 10: Tests for irritation and skin sensitization
Table 2

H2 concentration measured at different locations

Subject codeVA (%)OA (%)ONT (%)Test time (min)A (%)B (%)C (%)D (%)Exhaled breath (%)
DUYI 0 65 65 5 0 0.4 0 0 2.5
60 0.4 0 0 0.3
120 0 0 0 0
LLDI 0 66 66 5 0.3 0.3 0.3 0.3 1.9
60 0.3 0.3 0.4 0.3
120 0.3 0.5 0.5 0.3
YSFA 0 64 64 5 0.4 0.3 0.5 0.3 0.9
60 0.4 0.3 0 0
120 0 0.7 0 0
MLRO 0 64 64 5 0.3 0.3 0.3 0.4 2.8
60 0 0 0 0
120 0 0 0 0
ZCSH 0 64 64 5 0 0 0 0 2
60 0.3 0.3 0.7 0.6
120 0.7 0.3 0.6 0
DUYI 0 64 64 5 0.5 0 0 0 2.3
60 0 0 0 0
120 0 0 0 0
LLDI 0 66 66 5 0.4 0.5 0.6 0.4 1.3
60 0.4 0.4 0.4 0.4
120 0.8 0.7 0.5 0.4
YSFA 0 67 67 5 0.6 0.5 0.5 0.4 2
60 0.4 0.5 0.3 0.4
120 0.7 0.5 0.4 0.4
SXYU 0 64 64 5 0.4 0.3 0.3 0.3 2.2
60 0.5 0.4 0.4 0.4
120 0.4 0.4 0.4 0.4
HDFE 0 64 64 5 0 0 0 0 3.3
60 0 0 0 0
120 0.3 0.3 0 0
CDXI 0 66 66 5 0.3 0.3 0.3 0.3 3.8
60 0.4 0.4 0.4 0.4
120 0.4 0.4 0.4 0.4
ZDHU 0 64 64 5 0.7 0.4 0.7 0.4 1.4
60 0.4 0.5 0.7 0.5
120 0.5 0.4 0.6 0.5
MLRO 0 64 64 5 0.5 0.5 0.5 0 2.5
60 0.5 0 0 0
120 0.5 0.5 0.5 0.5
ZCSH 0 64 64 5 0.7 0.5 0.3 0 1.9
60 0.5 0.3 0.3 0.3
120 0 0.7 0.5 0

Note: Furthermore, H2 concentration is measured by New Cosmos Electric: XP-3140 (Osaka, Japan). First, when the AMS-H-01 is tuned on, the hydrogen concentration at: (1) ambient environment which is one meter away from the AMS-H-01 (“VA”); (2) output of the nebulizer (“OA”); and (3) output of special nasal tube or nasal mask connected to the nebulizer (“ONT”) are measured separately. Secondly, when each subject user is equipped with the nasal tube, H2 concentration around the subject user is detected at the location of 1 cm away from: (1) right ear (“A”); (2) left ear (“B”); (3) top of head (“C”); and (4) back of head (“D”) of the subject user after 5-minute, 60-minute, and 120-minute operation of the AMS-H-01 respectively. Last, the exhaled breath of the subject was collected using a gas sample collection bag immediately after the nasal tube was removed from the subject user, and then the hydrogen concentration of the exhaled breath was measured.

Table 3

Ignition results for three different flow rates

Flow rate (L/min)Results
3 Ignition noise detected, no explosion
2.5 Ignition noise detected, no explosion
2 Ignition noise detected, no explosion
  49 in total

1.  Hydrogen gas inhibits lung cancer progression through targeting SMC3.

Authors:  Dongchang Wang; Lifei Wang; Yu Zhang; Yunxia Zhao; Gang Chen
Journal:  Biomed Pharmacother       Date:  2018-05-29       Impact factor: 6.529

2.  Inhalation of high concentrations of hydrogen ameliorates liver ischemia/reperfusion injury through A2A receptor mediated PI3K-Akt pathway.

Authors:  He Li; Ouyang Chen; Zhouheng Ye; Rongjia Zhang; Huijun Hu; Ning Zhang; Junlong Huang; Wenwu Liu; Xuejun Sun
Journal:  Biochem Pharmacol       Date:  2017-02-08       Impact factor: 5.858

3.  Hydrogen inhalation attenuated bleomycin-induced pulmonary fibrosis by inhibiting transforming growth factor-β1 and relevant oxidative stress and epithelial-to-mesenchymal transition.

Authors:  Li Gao; Dingyuan Jiang; Jing Geng; Run Dong; Huaping Dai
Journal:  Exp Physiol       Date:  2019-10-23       Impact factor: 2.969

4.  Oral hydrogen water prevents chronic allograft nephropathy in rats.

Authors:  Jon S Cardinal; Jianghua Zhan; Yinna Wang; Ryujiro Sugimoto; Allan Tsung; Kenneth R McCurry; Timothy R Billiar; Atsunori Nakao
Journal:  Kidney Int       Date:  2009-11-11       Impact factor: 10.612

5.  Drinking hydrogen water ameliorated cognitive impairment in senescence-accelerated mice.

Authors:  Yeunhwa Gu; Chien-Sheng Huang; Tota Inoue; Takenori Yamashita; Torao Ishida; Ki-Mun Kang; Atsunori Nakao
Journal:  J Clin Biochem Nutr       Date:  2010-04-23       Impact factor: 3.114

6.  Effects of Hydrogen Gas Inhalation on Endometriosis in Rats.

Authors:  Y He; J Z Shi; R J Zhang; D X Zhai; D Zhang; C Q Yu; Y H Liu
Journal:  Reprod Sci       Date:  2016-09-27       Impact factor: 3.060

7.  Hydrogen gas inhalation enhances alveolar macrophage phagocytosis in an ovalbumin-induced asthma model.

Authors:  Peikai Huang; Shushan Wei; Weihua Huang; Penghui Wu; Shuyu Chen; Ailin Tao; Hongyu Wang; Zhenyu Liang; Rongchang Chen; Jie Yan; Qingling Zhang
Journal:  Int Immunopharmacol       Date:  2019-06-11       Impact factor: 5.714

Review 8.  Molecular hydrogen: a preventive and therapeutic medical gas for various diseases.

Authors:  Li Ge; Ming Yang; Na-Na Yang; Xin-Xin Yin; Wen-Gang Song
Journal:  Oncotarget       Date:  2017-09-21

9.  Protective effects of hydrogen gas in a rat model of branch retinal vein occlusion via decreasing VEGF-α expression.

Authors:  Pan Long; Weiming Yan; Mengshan He; Qianli Zhang; Zhe Wang; Manhong Li; Junhui Xue; Tao Chen; Jing An; Zuoming Zhang
Journal:  BMC Ophthalmol       Date:  2019-05-16       Impact factor: 2.209

10.  Molecular hydrogen suppresses glioblastoma growth via inducing the glioma stem-like cell differentiation.

Authors:  Meng-Yu Liu; Fei Xie; Yan Zhang; Ting-Ting Wang; Sheng-Nan Ma; Peng-Xiang Zhao; Xin Zhang; Tyler W Lebaron; Xin-Long Yan; Xue-Mei Ma
Journal:  Stem Cell Res Ther       Date:  2019-05-21       Impact factor: 6.832

View more
  1 in total

1.  Guidelines for the selection of hydrogen gas inhalers based on hydrogen explosion accidents.

Authors:  Yusuke Ichikawa; Shin-Ichi Hirano; Bunpei Sato; Haru Yamamoto; Yoshiyasu Takefuji; Fumitake Satoh
Journal:  Med Gas Res       Date:  2023 Apr-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.