Flow Diverter Apposition in Patients with Large or Giant Intracranial Aneurysms Evaluated on Three-Dimensional Fusion Images Acquired by High-Resolution Cone-Beam Computed Tomography and Digital Subtraction Angiography.

Naoki Kato1, Ichiro Yuki2, Katharina Otani3, Toshihiro Ishibashi4, Shota Kakizaki4, Gota Nagayama4, Fumiaki Maruyama4, Ayako Ikemura4, Issei Kan4, Tomonobu Kodama4, Yukiko Abe5, Yuichi Murayama4.   

Abstract

OBJECTIVE: To investigate flow diverter (FD) apposition on fused images acquired by high-resolution cone-beam computed tomography (CBCT) and 3-dimensional (3D) digital subtraction angiography.
METHODS: Patients with large or giant internal carotid artery aneurysms treated with Pipeline FDs who underwent CBCT imaging at our institution between October 2016 and May 2019 were included. Two neurosurgeons measured the maximum malapposition between FDs and vessel walls on 3D fusion images of high-resolution CBCT images displaying the FD and 3D digital subtraction angiography images displaying the vessels. Associations between the relative malapposition, FD diameter, vessel diameter, proximal and distal vessel diameter discrepancy, and siphon angle were evaluated by linear regression analysis. Inter-rater and intermethod (3D and 2D image) agreements of the malapposition measurements were assessed by Bland-Altman analysis and by interclass correlation coefficients.
RESULTS: Images of 2 patients were excluded because of image artifacts or fusion errors, and 3D fusion images were acquired in the remaining 26 patients. Our results did not suggest that relative malapposition was associated with vessel diameter (P = 0.12), vessel diameter discrepancy (P = 0.60), or syphon angle (P = 0.34), but relative malapposition increased by an estimated 13% (95% confidence interval: 4%-23%, P = 0.006) for each 1 mm increase in FD diameter. Inter-rater and intermethod agreements for apposition measurements were excellent and good, respectively.
CONCLUSIONS: Three-dimensional fusion images provided clear visualization of structures of both the stent and parent artery with excellent diagnostic reliability. Careful deployment may be needed for FDs with larger diameters, as they tended to have larger relative malapposition.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Aneurysm; Angiography; CT; Flow diverter

Year:  2021        PMID: 33359518     DOI: 10.1016/j.wneu.2020.12.068

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  1 in total

1.  Cone-beam CT angiography to assess the microvascular anatomy of intracranial arterial dissections.

Authors:  Francesco Diana; Eytan Raz; Francesco Biraschi; Daniele G Romano; Giada Toccaceli; Maksim Shapiro; Simone Peschillo
Journal:  Neuroradiol J       Date:  2021-09-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.