Do Kyoung Han1, Jeonghyun Oh2, Jaehyeon Lee3, Yong Gon Cho3, Ju Seong Park2, Jong Soon Choi4, Dal Sik Kim5, Joseph Kwon6.   

Abstract

We report a new paper-based multiplex analytical device (mPAD) for simultaneous screening of three analytes (glutamate dehydrogenase, toxin A, and toxin B) known as biomarkers for Clostridioides difficile infection (CDI). To overcome the limitation of common rapid assays (e.g. lateral flow immunochromatographic and enzyme immunoassays) in terms of multiplexing, sensitivity, simplicity, and ease-of-use, the mPAD is constructed with a three dimensional (3D) configuration of paper components with a multi-channel design. Multiple fluidic paths developed with wax-patterned paper allow the simultaneous detection of glutamate dehydrogenase, toxin A, and toxin B without any cross-reactivity. The 3D fluidic network on the mPAD facilitates a self-operating test procedure for the mixing and addition of amplification reagents with a one-step sliding operation. The results of the multiplex CDI assay can be easily interpreted by the naked eye within 10 min, and are visually intensified over time resulting in up to 3-fold signal amplification. Our device exhibited remarkable analytical performances for the simultaneous detection of three CDI biomarkers, providing a sensitivity of 97%, specificity of 88%, accuracy of 95%, and limits of detection for glutamate dehydrogenase, toxin A, and toxin B of 0.16 ng mL-1, 0.09 ng mL-1, and 0.03 ng mL-1, respectively. These results indicate the high applicability and feasibility of mPAD for multiplex testing for CDI with the advantages of being simple, sensitive, inexpensive, user-friendly, and equipment-free.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Antibiotic-resistant bacteria; Clostridioides difficile infection; Multiplex assay; Paper-based devices; Point-of-care testing; Signal amplification

Mesh:

Substances:

Year:  2020        PMID: 33358286     DOI: 10.1016/j.bios.2020.112894

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  3 in total

1.  Simple, fast, and instrumentless fabrication of paper analytical devices by novel contact stamping method based on acrylic varnish and 3D printing.

Authors:  Tatiane Alfonso de Araujo; Natália Canhete de Moraes; Jacqueline Marques Petroni; Valdir Souza Ferreira; Bruno Gabriel Lucca
Journal:  Mikrochim Acta       Date:  2021-11-27       Impact factor: 5.833

2.  Rapid Quantification of C. difficile Glutamate Dehydrogenase and Toxin B (TcdB) with a NanoBiT Split-Luciferase Assay.

Authors:  Hope Adamson; Modupe O Ajayi; Kate E Gilroy; Michael J McPherson; Darren C Tomlinson; Lars J C Jeuken
Journal:  Anal Chem       Date:  2022-05-28       Impact factor: 8.008

Review 3.  Toward Next Generation Lateral Flow Assays: Integration of Nanomaterials.

Authors:  Amadeo Sena-Torralba; Ruslan Álvarez-Diduk; Claudio Parolo; Andrew Piper; Arben Merkoçi
Journal:  Chem Rev       Date:  2022-09-06       Impact factor: 72.087

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.