Literature DB >> 33352917

Comparison of Phakic Intraocular Lens Vault Using Conventional Nomogram and Prediction Formulas.

Wakako Ando1, Kazutaka Kamiya2, Hideki Hayakawa1, Masahide Takahashi1, Nobuyuki Shoji1.   

Abstract

This study aimed to compare the achieved vault using a manufacturer's nomogram and the predicted vault using the currently available prediction formulas after posterior chamber phakic intraocular lens (EVO Implantable Collamer Lens; ICL, STAAR Surgical) implantation. We included 200 eyes of 100 consecutive patients (mean age ± standard deviation, 34.3 ± 7.8 years) undergoing ICL implantation with a central hole. Three months postoperatively, we quantitatively measured the actual vault, and we compared it with the predicted vault using anterior segment optical coherence tomography (CASIA 2, Tomey). The agreement rate of the recommended ICL size using the manufacturer's nomogram, the NK formula, and the KS formula was 50.0%. The achieved vault was 477.1 ± 263.7 µm, which was significantly smaller than the predicted vaults of 551.2 ± 335.1 and 606.4 ± 212.2 µm, using the NK and KS formulas, respectively (Dunnett test, p = 0.014, p < 0.001). The achieved vault was not significantly different from the predicted vault using the NK or KS formula (p = 0.386, p = 0.157) when selecting a 12.1 mm ICL size. It was not significantly different from the predicted vault using the NK formula (p = 0.962), but it was significantly smaller than that using the KS formula (p = 0.033) when selecting a 12.6 mm size. It was significantly smaller than the predicted vault using the NK and KS formulas (p < 0.001) when selecting 13.2 mm size. The total agreement rate of the recommended ICL size was approximately 50%. The predicted ICL vault tended to overestimate the actual ICL vault, especially when selecting a larger ICL size.

Entities:  

Keywords:  ICL; nomogram; phakic IOL; predictability; sizing; vault

Year:  2020        PMID: 33352917      PMCID: PMC7765914          DOI: 10.3390/jcm9124090

Source DB:  PubMed          Journal:  J Clin Med        ISSN: 2077-0383            Impact factor:   4.241


  3 in total

1.  Special Issue on Refractive Surgery.

Authors:  Majid Moshirfar; Yasmyne C Ronquillo
Journal:  J Clin Med       Date:  2022-01-28       Impact factor: 4.241

2.  Development of a Web-Based Ensemble Machine Learning Application to Select the Optimal Size of Posterior Chamber Phakic Intraocular Lens.

Authors:  Eun Min Kang; Ik Hee Ryu; Geunyoung Lee; Jin Kuk Kim; In Sik Lee; Ga Hee Jeon; Hojin Song; Kazutaka Kamiya; Tae Keun Yoo
Journal:  Transl Vis Sci Technol       Date:  2021-05-03       Impact factor: 3.283

3.  Vertically Fixated Posterior Chamber Phakic Intraocular Lens Implantation Through a Superior Corneal Incision.

Authors:  Kazutaka Kamiya; Wakako Ando; Hideki Hayakawa; Satoshi Gotoda; Nobuyuki Shoji
Journal:  Ophthalmol Ther       Date:  2022-02-04
  3 in total