Randomized Controlled Comparison of Valveless Trocar (AirSeal) vs Standard Insufflator with Ultralow Pneuomoperitoneum During Robotic Prostatectomy.

Ronney Abaza1,2, Oscar Martinez1, Christopher Murphy1.   

Abstract

Objective: To compare valveless insufflation (AirSeal®) with a conventional insufflation system (CIS) during robotic prostatectomy (RP) and the ability to use ultralow pneumoperitoneum at 6 mm Hg with each system as well as comparison of physiologic outcomes and pain scores. Patients and
Methods: We conducted a prospective study of 100 patients randomized to AirSeal or CIS during RP. The frequency of need for increasing pneumoperitoneum was assessed as well as arterial blood gases, respiratory/hemodynamic parameters, pain scores, and analgesic requirements. Quality of smoke evacuation and scope cleaning frequency were also measured.
Results: All procedures were completed at 6 mm Hg without needing to increase pressures with either insufflator. There were no statistically significant differences in partial pressure of carbon dioxide (PaCO2), partial pressure of oxygen (PaO2), HCO3, pH, carbon dioxide (CO2) elimination, or end-tidal carbon dioxide pressure (EtCO2) between groups. The AirSeal group had a lower maximum peritoneal pressure (7.9 vs 9.9 mm Hg, p < 0.001) but without differences in pain scores or analgesics. Surgeon-assessed smoke evacuation was poorer using CIS with more laparoscope cleanings in nonobese patients than with AirSeal (2.1 vs 3.0, p = 0.026).
Conclusion: Valveless-trocar insufflation provided more stable pressure but without benefits in physiologic or pain parameters. Previously identified benefits may have been negated by being able to complete all procedures at ultralow pressure with either insufflator, although an expert bedside assistant moderating suction may have contributed to feasibility of maintaining low pressure with CIS. A randomized trial of 6 vs 15 mm Hg is currently underway. The ClinicalTrials.gov Identifier: NCT02114164.

Entities:  

Keywords:  insufflation; prostate cancer; robotic surgery; ultralow pneumoperitoneum pressure; valveless trocars

Mesh:

Substances:

Year:  2021        PMID: 33349135     DOI: 10.1089/end.2020.1025

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  1 in total

1.  Intraoperative Measures to Reduce the Risk of COVID-19 Transmission During Minimally Invasive Procedures: A Systematic Review and Critical Appraisal of Societies' Recommendations.

Authors:  Giovanni A Tommaselli; Philippe Grange; Crystal D Ricketts; Jeffrey W Clymer; Raymond S Fryrear
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2021-07-28       Impact factor: 1.719

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.