Using a pulsed-beam transmission electron microscope, we discover a reduction in damage to methylammonium lead iodide (MAPbI3) as compared to conventional beams delivered at the same dose rates. For rates as low as 0.001 e·Å-2·s-1, we find up to a 17% reduction in damage at a total dose of 10 e·Å-2. We systematically study the effects of number of electrons in each pulse and the duration between pulse arrival. Damage increases for both, though the number of electrons per pulse has a larger effect. A crossover is identified, where a pulsed beam causes more damage than a conventional one. Although qualitatively similar to previous findings, the degree to which damage is reduced in MAPbI3 is less than that observed for other materials (e.g., C36H74), supporting the hypothesis that the effects are material- and damage-mechanism-dependent. Despite this, the observation here of damage reduction for relatively large electron packets (up to 200 electrons per pulse) suggests that MAPbI3 is in fact less susceptible to irradiation than C36H74, which may be related to reported self-healing effects. This work provides insights into damage processes and durability in hybrid perovskites and also illustrates the viability of using pulsed-beam TEM to explore the associated molecular-level routes to degradation, analogous to laser-accelerated energetic pulsed electron beams and the study of damage to biomolecules, cells, and tissues in radiobiology.
Using a pulsed-beam transmission electron microscope, we discover a reduction in damage to methylammonium lead iodide (MAPbI3) as compared to conventional beams delivered at the same dose rates. For rates as low as 0.001 e·Å-2·s-1, we find up to a 17% reduction in damage at a total dose of 10 e·Å-2. We systematically study the effects of number of electrons in each pulse and the duration between pulse arrival. Damage increases for both, though the number of electrons per pulse has a larger effect. A crossover is identified, where a pulsed beam causes more damage than a conventional one. Although qualitatively similar to previous findings, the degree to which damage is reduced in MAPbI3 is less than that observed for other materials (e.g., C36H74), supporting the hypothesis that the effects are material- and damage-mechanism-dependent. Despite this, the observation here of damage reduction for relatively large electron packets (up to 200 electrons per pulse) suggests that MAPbI3 is in fact less susceptible to irradiation than C36H74, which may be related to reported self-healing effects. This work provides insights into damage processes and durability in hybrid perovskites and also illustrates the viability of using pulsed-beam TEM to explore the associated molecular-level routes to degradation, analogous to laser-accelerated energetic pulsed electron beams and the study of damage to biomolecules, cells, and tissues in radiobiology.
Comprehensive understanding of high-power-conversion
efficiencies
of hybrid organic–inorganic perovskite (HOIP) photovoltaic
cells, as well as apparent self-healing properties and the associated
implications for durability under illumination, requires elucidation
of atomic and nanoscale properties and behaviors of these materials.[1,2] Structural, chemical, electronic, and dynamic properties at this
scale are accessible with transmission electron microscopy (TEM),
but the stability of HOIPs is such that electron-beam damage can be
significant, while connections to photon-induced damage can be drawn.[3−5] Indeed, sensitivity of HOIPs to even low dose rates (∼1 e·Å–2·s–1) may limit what can be
learned about fundamental structure/function relationships and illustrates
the importance of developing new methods for probing degradation and
uncovering new behaviors.[6−8] For example, damage during electron-beam
irradiation is thought to occur through a combination of charging,
ionic excitation, and heating leading to ion migration and separation
of organic and inorganic constituents.[8,9] As such, a
number of methods, such as cryo-electron microscopy and low-dose imaging
and diffraction, have been used to reduce deleterious beam effects.[10,11]Femtosecond (fs) laser-driven and picosecond chopped-beam
sources
in modified TEMs offer additional means for mitigating and studying
damage.[12−16] These methods employ temporally-modulated beams, where electrons
are delivered to the specimen in discrete pulses with well-defined
durations between each, rather than in the random fashion of conventional
sources (e.g., thermionic). Indeed, time-dependent
aspects of charging and thermal energy, as well as the dynamic self-healing
properties of HOIPs, suggest that there may be insights into these
temporal mechanisms that can be achieved by using well-timed pulses.[3,17−19] Accordingly, here we explore fs laser-driven pulsed
TEM for studying damage in HOIPs, specifically methylammonium lead
iodide (CH3NH3PbI3 or MAPbI3). Our main goal was to determine—all else being equal—if
a pulsed beam leads to a reduction in MAPbI3 damage compared
to a conventional beam. We focused on quantitatively comparing damage
caused by pulsed beams to that caused by conventional thermionic beams at the same dose rates and the same total doses. We also
studied the effects of the number of electrons per pulse (e/p) and
the duration between pulse arrival at the specimen (f–1, where f is the laser repetition
rate). We find a clear reduction in damage for pulsed beams compared
to that for random (thermionic) beams, as well as an apparent exacerbation
of damage with increasing instantaneous dose (i.e., with more electrons per pulse) even with a longer time elapsed
between the arrival of each. We also observe a crossover effect, where
pulsed beams become more detrimental than random beams. Interestingly,
the degree to which damage is reduced in MAPbI3 is lower
than that for C36H74, supporting the hypothesis
that the pulsed-beam effect is material- and mechanism-dependent.
In addition to observing damage reduction with pulsed-beam TEM in
an electronic material, the dose rates and doses used here are comparable
to those typically used in conventional low-dose TEM. The combination
of relatively low repetition rates in laser-driven pulsed-beam TEM
(thus maximizing time between electron arrival and specimen relaxation
time) at low-dose beam currents enables quantitative studies of damage
mechanisms and the opportunity for exploring improvements in data
quality, particularly when combined with cryogenic methods.
Results
and Discussion
Beam-damage mechanisms in TEM are numerous
and often synergistic,
necessitating detailed design and systematic execution of experiments.[5,20] Conveniently, fs pulsed lasers in a stable lab environment afford
high levels of control, enabling accurate and precise quantification
of pulsed TEM beam damage.[14] An overview
of the fs laser-driven approach used here and the method for quantifying
damage to MAPbI3 specimens are shown in Figure . Temporal regularity of the
pulsed electron beam was accomplished using a 300-fs pulsed laser
(PHAROS, Light Conversion), which confines the electron emission to
a train of 300-fs windows evenly spaced in time by f–1. The pulse train was generated via the photoelectric effect in a modified 200 kV TEM (Tecnai Femto,
Thermo Fisher) using ultraviolet laser pulses (hν
= 4.8 eV) and a LaB6 electron source (φ = 2.4 eV; Figure a). The dose rate
was controlled with both the laser-pulse fluence (which dictates e/p)
and f (which dictates the number of pulses per second
and the specimen relaxation time between each pulse, f–1).
Figure 1
Laser-driven pulsed-beam TEM and quantification
of damage to MAPbI3. (a) Simplified schematic of the electron-source
region with
pertinent aspects and components labeled. Adapted with permission
from VandenBussche, E. J.; Flannigan, D. J. Nano Lett.2019, 19, 6687–6694. Copyright
2019 American Chemical Society. (b) Structure of MAPbI3.[21] Crystallographic axes are shown in
the lower left of the panel. (c) Low-magnification bright-field image
of the TEM specimens. The diffraction pattern (inset; scale bar =
5 nm–1) was obtained from the red circled region.
(d) Intensity plot from azimuthally averaging the pattern in (c).
Red peaks are those used for monitoring beam-induced intensity changes.
The space group used for indexing was I4/mcm, though the structure is not yet universally agreed
upon.[22,23] (e) Bragg-intensity fading curve for a pulsed
electron beam (dose rate = 0.001 e·Å–2·s–1; beam size = 245 ± 3 μm2; 50.4 ± 1.0 e/p; f–1 = 2 μs). The red curve is used to determine ΔI at 10 e·Å–2, which here is
−19.2%.
Laser-driven pulsed-beam TEM and quantification
of damage to MAPbI3. (a) Simplified schematic of the electron-source
region with
pertinent aspects and components labeled. Adapted with permission
from VandenBussche, E. J.; Flannigan, D. J. Nano Lett.2019, 19, 6687–6694. Copyright
2019 American Chemical Society. (b) Structure of MAPbI3.[21] Crystallographic axes are shown in
the lower left of the panel. (c) Low-magnification bright-field image
of the TEM specimens. The diffraction pattern (inset; scale bar =
5 nm–1) was obtained from the red circled region.
(d) Intensity plot from azimuthally averaging the pattern in (c).
Red peaks are those used for monitoring beam-induced intensity changes.
The space group used for indexing was I4/mcm, though the structure is not yet universally agreed
upon.[22,23] (e) Bragg-intensity fading curve for a pulsed
electron beam (dose rate = 0.001 e·Å–2·s–1; beam size = 245 ± 3 μm2; 50.4 ± 1.0 e/p; f–1 = 2 μs). The red curve is used to determine ΔI at 10 e·Å–2, which here is
−19.2%.Specimens were synthesized by
spin-coating a 0.3 M solution of
MAPbI3 with 10% molar excess methylammonium iodide in a
4:1 volume ratio of DMF/DMSO onto holey amorphous-carbon grids (Quantifoil,
Electron Microscopy Sciences) (Figure b,c) inside an inert atmosphere glove box. First, the
grid was reversibly adhered to a silicon support using a drop of toluene,
followed by heating at 100 °C for 1 min. Next, 0.2 mL of MAPbI3 solution was dropped onto the supported grid while spinning
at 3000 rpm for 2 min. Finally, the grid was annealed at 100 °C
for 1 h. This produced free-standing, polycrystalline MAPbI3 islands spanning the 2.5 μm holes of the grids (Figure c,d). Accordingly, each island
was determined from the bright-field TEM images to be 4.8 ± 0.2
μm2 in area (error is one standard deviation from
the mean). Damage was quantified using a fading-curve method, where
the reduction in Bragg-beam intensities (ΔI: indicative of the destruction of MAPbI3 crystallinity)
was tracked as a function of accumulated dose (Figure e).[24] Peaks arising
from planes with d < 6.3 Å were used to monitor
the damage effects for both beam types (i.e., pulsed
and conventional). Note that ΔI = (I – Io)/Io, which is the normalized change in intensity relative
to that at nominally zero dose (Io). A
total accumulated dose of 10 e·Å–2 was
used as the reference point throughout.Comparison of damage
caused to MAPbI3 by a pulsed beam
to that caused by a conventional thermionic beam is shown in Figure . For a common dose
rate (0.001 e·Å–2·s–1 for an illuminated area of 245 ± 3 μm2), there
is a clear reduction in the extent to which the intensities fade for
the pulsed beam under the conditions used for this experiment (50.4
± 1.0 e/p and f–1 = 2 μs).
At a total accumulated dose of 10 e·Å–2, the intensity change for the pulsed beam is −19.2%, while
that of the thermionic beam is −23.2%. This shows that providing
temporally-regular pauses in electron-beam irradiation, and thus providing
regular periods of specimen relaxation and a reduction in exacerbating
effects (e.g., multi-electron impact within the damage
radius), leads to enhanced preservation of MAPbI3 structural
order compared to that of conventional random-beam low-dose methods.
Note that for pulsed-beam TEM experiments performed on C36H74 and bacteriorhodopsin, the extent to which damage
was reduced was greater than that seen here, though evidence indicates
that the degree of reduction also depends on f–1.[14,16] This suggests that the extent
to which damage is reduced is material- and mechanism-dependent (perhaps
strongly so) and that the overall effect of pulsed beams with respect
to damage reduction is a general one.
Figure 2
Bragg-peak intensity fading curves for
MAPbI3 for pulsed
(blue; also shown in Figure e) and thermionic (red) beams delivered at the same dose rate
(0.001 e·Å–2·s–1). Fits to the data are for determining ΔI at 10 e·Å–2, which here is −19.2%
for the pulsed beam and −23.2% for the conventional thermionic
beam (as indicated by the blue and red horizontal dot-dashed lines,
respectively). The beam was spread over an area of 245 ± 3 μm2 for both types. The pulsed beam consisted of 50.4 ±
1.0 e/p and f–1 = 2 μs (i.e., a 500-kHz repetition rate).
Bragg-peak intensity fading curves for
MAPbI3 for pulsed
(blue; also shown in Figure e) and thermionic (red) beams delivered at the same dose rate
(0.001 e·Å–2·s–1). Fits to the data are for determining ΔI at 10 e·Å–2, which here is −19.2%
for the pulsed beam and −23.2% for the conventional thermionic
beam (as indicated by the blue and red horizontal dot-dashed lines,
respectively). The beam was spread over an area of 245 ± 3 μm2 for both types. The pulsed beam consisted of 50.4 ±
1.0 e/p and f–1 = 2 μs (i.e., a 500-kHz repetition rate).A number of effects and conditions—in addition to beam-induced
damage to crystalline order—can produce an apparent dose-dependent
drop in Bragg-beam intensities. Thus, drawing conclusions from direct
comparisons made across multiple measurements and specimens requires
control of myriad variables that might influence, interfere with,
and overwhelm the intrinsic beam-damage behavior (e.g., specimen thickness and lateral dimensions, specimen bending under
the beam, lab- and instrument-temperature stability, and specimen
and beam drift). Accordingly, sources of error and artifacts were
identified and accounted for via control experiments
and monitoring of experimental conditions routinely employed during
these studies. Thus, detailed descriptions of the measures taken here
are reported elsewhere.[14] As an example
of the requirements that need to be met for data to be accepted, experiments
were rejected and repeated when the beam current, as well as the beam
size, differed by more than 1% before and after acquisition of a data
series. Experiments were conducted after observable directional specimen
drift ceased. Furthermore, experiments were rejected if the specimen
was found to have directionally drifted more than 1% of the substrate
hole diameter (i.e., by 25 nm) between the start
and the finish of data acquisition. Pre-irradiation was eliminated
by navigating the specimen movement systematically and ensuring that
no two experiments were conducted within two beam diameters of one
another. Data presented here consist of multiple experimental trials
conducted over several days.To better understand the origins
of damage reduction when using
a pulsed beam, effects of varying e/p and f–1 were systematically explored (Figure ). Depending upon the mechanisms at work, one might
intuitively expect an increase in damage with increased e/p, due to
more inelastic electron/specimen collisions occurring within a given
pulse, and also with decreased f–1, due to there being less time between electron arrival at the specimen
(thus reducing the relaxation and recovery time and increasing the
probability of simultaneous or near-simultaneous impact within a particular
damage radius).[14−16] However, regimes do exist wherein the benefits gained
by increasing f–1 are lost when
simultaneously increasing e/p, as observed for C36H74.[14] This indeed is the case for
MAPbI3. For dose rates of 0.001 and 0.01 e·Å–2·s–1 (Figure a,b, respectively), damage increases with
increasing e/p despite an accompanying modest increase in f–1 from 2 μs to 4 μs for
the larger electron pulses. Such an effect was also observed in single
crystals of C36H74, where going from 1 e/p to
5 e/p had a substantially larger effect on damage than did decreasing f–1 from 20 μs to 5 μs.[14] Taken altogether, these results suggest that
specimen relaxation and recovery processes at work during the few
microseconds between pulses can be overwhelmed by exacerbating effects
of multi-electron impact within a given damage radius. That is, additional
energy deposited into the already-excited specimen region causes further
structural damage that otherwise would have recovered—or would
have been exponentially less significant (owing to the exponential
relationships for reaction kinetics, diffusion rates, and thermal
energy)—during a single electron event isolated in space and
time.
Figure 3
Effect of e/p and f–1 on damage
to MAPbI3 for a total dose of 10 e·Å–2. (a) Bragg-intensity reduction (ΔI) for pulsed
(blue) and thermionic (therm., red) beams administered at a dose rate
of 0.001 e·Å–2·s–1. The time between electron pulses in microseconds (f–1) is noted below the number of electrons per
pulse (e/p). Error bars for 50 e/p, 100 e/p, and thermionic are one
standard deviation over 2, 4, and 2 separate experiments, respectively.
(b) Bragg-intensity reduction (ΔI) for pulsed
(blue) and thermionic (therm., red) beams administered at a dose rate
of 0.01 e·Å–2·s–1. Error bars for 200 e/p, 400 e/p, and thermionic are one standard
deviation over 2, 2, and 3 separate experiments, respectively.
Effect of e/p and f–1 on damage
to MAPbI3 for a total dose of 10 e·Å–2. (a) Bragg-intensity reduction (ΔI) for pulsed
(blue) and thermionic (therm., red) beams administered at a dose rate
of 0.001 e·Å–2·s–1. The time between electron pulses in microseconds (f–1) is noted below the number of electrons per
pulse (e/p). Error bars for 50 e/p, 100 e/p, and thermionic are one
standard deviation over 2, 4, and 2 separate experiments, respectively.
(b) Bragg-intensity reduction (ΔI) for pulsed
(blue) and thermionic (therm., red) beams administered at a dose rate
of 0.01 e·Å–2·s–1. Error bars for 200 e/p, 400 e/p, and thermionic are one standard
deviation over 2, 2, and 3 separate experiments, respectively.In addition to the overall trends shown in Figure , a type of crossover
or threshold region
exists, where more—rather than less—damage is caused
by the pulsed beam as compared to a conventional beam delivered at
the same dose rate (Figure b). Here, such a crossover is seen when increasing from 200
e/p separated by 2 μs to 400 e/p separated by 4 μs. This
behavior lends support to the hypothesis that it is multi-electron
impact leading to compounding effects that is driving the dramatic
increase in damage with increasing e/p.[14] In essence, this is due to additional energy being deposited into
already-excited specimen regions prior to full relaxation back to
ground-state conditions. Interestingly, the instantaneous dose rate
for a single pulse of 400 electrons confined to a 300-fs window (more
likely a few picoseconds due to Coulombic expansion)[25] is 5 × 104 e·Å–2·s–1, assuming uniform illumination across
the 245-μm2 beam area. It is therefore perhaps surprising
that such a beam does not produce substantially more damage than observed,
though the number of electrons per Å2 per pulse is
only 2 × 10–8. Though the time-averaged reduction
in damage for pulsed beams is clear, pulse-to-pulse behaviors and
the resulting specimen effects are largely unknown and require additional
investigation.Damage reduction (and increase at the crossover)
to MAPbI3 when using pulsed-beam TEM is likely driven by
temporal processes
that are active on the pulse-to-pulse timescales.[12−16] Such processes include thermal effects which, while
not yet well understood for electron-beam excitation and damage in
MAPbI3, are important in similar low-thermal-conductivity
materials.[20,24,26] The effect of thermal processes in damage reduction can be appreciated
by noting timescales of thermal diffusion and relaxation in MAPbI3 relative to f–1. Using
known constants,[27−30] it is estimated that thermal energy deposited into a 2.5-μm
diameter MAPbI3 crystal would largely dissipate into the
carbon substrate within several microseconds. Note, however, that
temperature from the perspective of the entire specimen is likely
too coarse a view when considering pulsed-beam damage mechanisms and
temporal aspects of molecular-scale excitations. Nevertheless, this
estimate indicates that thermal-dissipation times are comparable to f–1, suggesting complete relaxation of
increased vibrational energies prior to a subsequent inelastic collision
is a plausible source of reduced damage when electron delivery to
the specimen is precisely temporally controlled.Timescales
of electron–phonon coupling and lattice thermalization
are also worth noting. Electron–phonon coupling in MAPbI3 films occurs on the order of hundreds of femtoseconds, while
lattice thermalization takes a few picoseconds.[31] Owing to the electron-pulse duration, most or all will
arrive within this time frame. Furthermore, increasing e/p generally
causes an increase in pulse duration,[25] thus creating an environment where the likelihood of exacerbating
effects contributing to damage is increased. Systematic pulse-duration
experiments may shed light on such effects. Indeed, similar arguments
can be made regarding specimen charging, charge dissipation, and electron-pulse
durations and timing.
Conclusions
In conclusion, for common
total doses and dose rates, we have discovered
a regime where using a pulsed-beam TEM leads to reduced damage to
MAPbI3. The degree of reduction is enhanced by using pulses
with fewer electrons, while the duration between pulses appears to
have a smaller but still non-trivial effect. We also discover a crossover
effect, in which pulsed beams cause more damage than an otherwise
identical conventional beam. This constitutes the first example of
these effects for direct-comparison experiments at dose rates commonly
used in low-dose TEM. The findings also support the hypothesis that
the behavior is material- and mechanism-dependent owing to differences
when compared to other materials systems. These results have fundamental
and practical implications, in that fs-laser-driven pulsed TEM offers
a combination of stability and tunability that affords studying specific
damage mechanisms and durability of MAPbI3, as well as
providing structural, chemical, electronic, and dynamic information
from less-damaged specimens obtained at low repetition rates compared
to RF-modulated chopped electron beams.
Authors: Andrea Pisoni; Jaćim Jaćimović; Osor S Barišić; Massimo Spina; Richard Gaál; László Forró; Endre Horváth Journal: J Phys Chem Lett Date: 2014-07-08 Impact factor: 6.475
Authors: Yi Yu; Dandan Zhang; Christian Kisielowski; Letian Dou; Nikolay Kornienko; Yehonadav Bekenstein; Andrew B Wong; A Paul Alivisatos; Peidong Yang Journal: Nano Lett Date: 2016-11-07 Impact factor: 11.189
Authors: Tiarnan A S Doherty; Andrew J Winchester; Stuart Macpherson; Duncan N Johnstone; Vivek Pareek; Elizabeth M Tennyson; Sofiia Kosar; Felix U Kosasih; Miguel Anaya; Mojtaba Abdi-Jalebi; Zahra Andaji-Garmaroudi; E Laine Wong; Julien Madéo; Yu-Hsien Chiang; Ji-Sang Park; Young-Kwang Jung; Christopher E Petoukhoff; Giorgio Divitini; Michael K L Man; Caterina Ducati; Aron Walsh; Paul A Midgley; Keshav M Dani; Samuel D Stranks Journal: Nature Date: 2020-04-15 Impact factor: 49.962