Non-contact specular microscopy with Topcon instruments to assess central corneal thickness of healthy human eyes - A 20 year review.

M J Doughty1.   

Abstract

BACKGROUND: The purpose of this review was to evaluate the consistency of central corneal thickness (CCT) values reported with use of Topcon SP-2000 P and SP-3000 P non-contact specular microscopes since their introduction in 1999 with the two microscopes having been commonly used in a wide range of studies.
METHODS: As a primary resource, PubMed was used to search for peer-reviewed articles in any language that included CCT values obtained with non-contact specular microscopy reported for humans with nominally healthy corneas. Relevant articles were obtained and any cited publications also checked.
RESULTS: A total of 76 articles were identified which reported CCT on different small-to-moderate sized groups of individuals, published between 1999 and 2019. From these, an overall group mean CCT value of 0.525 ± 0.013 mm (median 0.525 mm) can be calculated. An estimated 95 % confidence interval (CI, based on 1.96 SD) would be between 0.500 and 0.550 mm. For the two Topcon models, the group mean ± SD values were 0.529 ± 0.013 mm and 0.517 ± 010 mm respectively. An assessment of the CCT data sets in relation to the reported average age indicated no statistically significant effect (p = 0.289, r = -0.129). Very similar average CCT values were also encountered in 4 other reports where these microscopes were used in large-scale population studies as well as in 2 other reports using the newer Topcon SP-1 P model.
CONCLUSIONS: The Topcon stand-alone non-contact specular microscopes have yielded consistent and predictable corneal thickness measures over many years. Crown
Copyright © 2020. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Cornea; Human; Non-contact specular microscopy; Pachymetry

Year:  2020        PMID: 33341365     DOI: 10.1016/j.clae.2020.11.005

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  1 in total

1.  A Liquid Hydrogel to Restore Long Term Corneal Integrity After Perforating and Non-Perforating Trauma in Feline Eyes.

Authors:  Alejandro Juarez; Mohamed Djallali; Marilyse Piché; Mathieu Thériault; Marc Groleau; Sharifa Beroual; Christopher D McTiernan; Grace Lin; Pierre Hélie; Michel Carrier; May Griffith; Isabelle Brunette
Journal:  Front Bioeng Biotechnol       Date:  2021-12-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.