Comparison of design, metallurgy, mechanical performance and shaping ability of replica-like and counterfeit instruments of the ProTaper Next system.

J N R Martins1,2,3, E J N L Silva4,5, D Marques3,6, F Belladonna5, M Simões-Carvalho5, E Camacho7, F M Braz Fernandes7, M A Versiani8.   

Abstract

AIM: To compare the ProTaper Next (PTN) system with a replica-like and a counterfeit system regarding design, metallurgy, mechanical performance and shaping ability.
METHODOLOGY: Replica-like (X-File) and counterfeit (PTN-CF) instruments were compared to the PTN system regarding design (microscopy), phase transformation temperatures (differential scanning calorimetry), nickel-titanium ratio (energy-dispersive X-ray spectroscopy), cyclic fatigue, torsional resistance, bending strength, and untouched canal areas in extracted mandibular molars (micro-CT). anova, post hoc Tukey's and Kruskal-Wallis tests were used according to normality assessment (Shapiro-Wilk test) with the significance level set at 5%.
RESULTS: Overall similarities in design and nickel-titanium (Ni/Ti) ratio were observed amongst instruments with the X-File having a smoother surface finish. PTN and PTN-CF had mixed austenite plus R-phase (R-phase start approximately at 45 ºC and near 30 ºC, respectively), whilst X-File instruments were austenitic (R-phase started at approximately at 17 ºC) at room temperature (20 ºC). PTN-CF had the greatest inconsistency in the phase transformation temperatures. Time to fracture of PTN-CF X2 and X3 was significantly shorter than PTN and X-File instruments (P < 0.05), whilst no difference was noted in maximum torque to fracture amongst the tested systems (P > 0.05). X-Files and PTN-CF had a stress-induced phase change during bending load. Mean unprepared surface areas of root canals were 25.8% (PTN), 31.1% (X-File) and 32.5% (PTN-CF) with no significant difference amongst groups (P > 0.05).
CONCLUSION: Similarities amongst the systems were noted in the Ni/Ti ratio and maximum torque to fracture, whilst differences were observed in the design, phase transformation temperatures and mechanical behaviour. The ProTaper Next counterfeit instruments could be considered as the less secure system considering its low-cyclic fatigue resistance. Apart from these differences, the unprepared canal surface areas, obtained with the tested systems, were similar.
© 2020 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  bending load; counterfeit instrument; cyclic fatigue; differential scanning calorimetry; micro-CT; nickel-titanium alloy; replica-like system; torsional strength

Year:  2021        PMID: 33300121     DOI: 10.1111/iej.13463

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  Multimethod Assessment of Design, Metallurgical, and Mechanical Characteristics of Original and Counterfeit ProGlider Instruments.

Authors:  Jorge N R Martins; Emmanuel J N L Silva; Duarte Marques; Sofia Arantes-Oliveira; António Ginjeira; João Caramês; Francisco M Braz Fernandes; Marco A Versiani
Journal:  Materials (Basel)       Date:  2022-06-02       Impact factor: 3.748

2.  Mechanical and Metallurgical Evaluation of 3 Different Nickel-Titanium Rotary Instruments: An In Vitro and In Laboratory Study.

Authors:  Alessio Zanza; Paola Russo; Rodolfo Reda; Paola Di Matteo; Orlando Donfrancesco; Pietro Ausiello; Luca Testarelli
Journal:  Bioengineering (Basel)       Date:  2022-05-20

3.  Nickel-Titanium Rotary Instruments: Mechanical and Metallurgical Characteristics.

Authors:  Alessio Zanza; Rodolfo Reda; Luca Testarelli
Journal:  Clin Pract       Date:  2022-02-14

4.  Design, Metallurgical Features, and Mechanical Behaviour of NiTi Endodontic Instruments from Five Different Heat-Treated Rotary Systems.

Authors:  Jorge N R Martins; Emmanuel J N L Silva; Duarte Marques; Mário Rito Pereira; Victor T L Vieira; Sofia Arantes-Oliveira; Rui F Martins; Francisco Braz Fernandes; Marco Versiani
Journal:  Materials (Basel)       Date:  2022-01-28       Impact factor: 3.623

5.  A Multimethod Assessment of a New Customized Heat-Treated Nickel-Titanium Rotary File System.

Authors:  Emmanuel J N L Silva; Jorge N R Martins; Natasha C Ajuz; Henrique S Antunes; Victor T L Vieira; Francisco M Braz Fernandes; Felipe G Belladonna; Marco A Versiani
Journal:  Materials (Basel)       Date:  2022-07-31       Impact factor: 3.748

6.  Can Type of Instrumentation and Activation of the Final Irrigant Improve the Obturation Quality in Oval Root Canals? A Push-Out Bond Strength Study.

Authors:  Ajinkya M Pawar; Anda Kfir; Zvi Metzger; Anuj Bhardwaj; Yeyen Yohana; Dian Agustin Wahjuningrun; Alexander Maniangat Luke; Bhaggyashri A Pawar
Journal:  Biology (Basel)       Date:  2022-01-01

Review 7.  Metallurgical Tests in Endodontics: A Narrative Review.

Authors:  Alessio Zanza; Marco Seracchiani; Rodolfo Reda; Gabriele Miccoli; Luca Testarelli; Dario Di Nardo
Journal:  Bioengineering (Basel)       Date:  2022-01-12
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.