Emily F Dunn1, Adil Akhtar1, Adam Dunn2, Scott Lacey1, Ethan Pauley3, Colton Powers3, James McKee4, Daniel Petereit4.
Abstract
PURPOSE: The supply of N95 masks and filtering facepiece respirators (FFRs) has been limited nationally owing to the coronavirus disease 2019 pandemic. Ultraviolet C (UVC) light has been suggested as a potential option for decontamination of FFRs by the Centers for Disease Control. There has been a lack of publications characterizing UVC dose distribution across FFRs. METHODS AND MATERIALS: A UVC light box and FFR rack system was assembled using low-pressure mercury lamps peaked at 254 nm and aluminum flashing to reduce shadowing effect. Dose was characterized with the use of ultraviolet (UV) intensity labels and an ultraviolet germicidal irradiation (UVGI) National Institute of Standards and Technology traceable meter. Ozone production was evaluated after extended bulb run time.Entities:
Year: 2020 PMID: 33283072 PMCID: PMC7705333 DOI: 10.1016/j.adro.2020.100636
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
Figure 1Physical dimensions of ultraviolet C light sources and mask rack. Completed structure includes 2 identical source racks, which contain 6 lamps in pairs, fastened to the structure at heights of 75, 112.5, and 153.5 cm, respectively. The mask rack is sandwiched between the 2 light sources and was manufactured to suspend approximately 12 FFRs. There are 4 rows, each able to suspend 3 FFRs at heights of 68.5, 99.5, 131.5, and 162 cm from the base, respectively. When bases of the garment racks are abutted, the distance from light sources to the rack is 54 cm. Abbreviation: FFR = filtering facepiece respirators.
Figure 2Light sources with 6 paired bulbs laterally and mask rack centrally within aluminum flashing box that is fully enclosed to increase ultraviolet C reflection and reduce shadowing. Abutment of bases inferiorly ensures identical geometry for each use.
Figure 3Locations of UVGI meter adherence to 5 positions on the convex (A) and concave (B) locations of the FFR to evaluate exposure dose. Dose output is demonstrated in Table 2. Abbreviations: FFR = filtering facepiece respirators; UVGI = ultraviolet germicidal irradiation.
Quantitative measurement of exposure dose in mW/cm2 at 5 locations on convex (A) and concave (B) surface of the FFR, as demonstrated in Figure 3
| Mask height | Rack column | Position on mask | Irradiance (mW/cm2) | Irradiance normalized to position 3A | Time (sec) to deliver 1000 (mJ/cm2) |
|---|---|---|---|---|---|
| 68.5 (cm) | 1 | 1A | 1.868 | 1.24 | 535 |
| 2A | 0.637 | 0.42 | 1570 | ||
| 4A | 0.745 | 0.50 | 1343 | ||
| 5A | 0.746 | 0.50 | 1340 | ||
| 1B | 1.212 | 0.81 | 825 | ||
| 2B | 1.129 | 0.75 | 886 | ||
| 3B | 1.205 | 0.80 | 830 | ||
| 4B | 1.396 | 0.93 | 716 | ||
| 5B | 1.574 | 1.05 | 635 | ||
| 162.0 (cm) | 3 | 1A | 1.150 | 0.72 | 870 |
| 2A | 0.966 | 0.61 | 1036 | ||
| 4A | 0.775 | 0.49 | 1290 | ||
| 5A | 1.684 | 1.06 | 594 | ||
| 1B | 1.505 | 0.94 | 664 | ||
| 2B | 1.126 | 0.71 | 888 | ||
| 3B | 1.342 | 0.84 | 745 | ||
| 4B | 1.241 | 0.78 | 806 | ||
| 5B | 0.675 | 0.42 | 1481 |
Abbreviation: FFR = filtering facepiece respirator.
Measurements obtained via ultraviolet germicidal irradiation meter. The most divergent locations were selected as examples of dose distribution across the FFR. Substantial variation in dose is demonstrated and dependent on distance from bulb, proximity to aluminum flashing, and arced surface of FFR. The irradiance at each position is normalized to position 3A (bold). Additional measurements from other rack locations are demonstrated in Table E1.
Exposure dose en face of each bulb source to 12 different locations along the respirator rack, as denoted in Figure 1
| Height (cm) | Irradiance (mW/cm2) | Exposure dose (mJ/cm2) | ||||
|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 1 | 2 | 3 | |
| Light source 1 | ||||||
| 162.0 | 0.955 | 1.068 | 0.979 | 1261 | 1410 | 1292 |
| 131.5 | 1.236 | 1.406 | 1.292 | 1632 | 1856 | 1705 |
| 99.5 | 1.349 | 1.526 | 1.380 | 1781 | 2014 | 1822 |
| 68.5 | 1.120 | 1.182 | 1.046 | 1478 | 1560 | 1381 |
| Light source 2 | ||||||
| 162.0 | 1.012 | 1.085 | 0.941 | 1336 | 1432 | 1242 |
| 131.5 | 1.284 | 1.393 | 1.216 | 1695 | 1839 | 1605 |
| 99.5 | 1.208 | 1.311 | 1.128 | 1595 | 1731 | 1489 |
| 68.5 | 0.873 | 0.908 | 0.784 | 1153 | 1198 | 1035 |
| Goal exposure dose (mJ/cm2) = | 1000 | |||||
| Time (seconds) = | 1320 | |||||