| Literature DB >> 33282854 |
Jialong Wang1, Xueqian Liu2, Chao Wang2, Dengren Liu1, Fang Li2, Li Wang1, Shufeng Liu1.
Abstract
A novel electrochemical biosensing method for protein kinase (PKA) activity was demonstEntities:
Keywords: electrochemical biosensor; polydopamine; protein kinase; reduced graphene oxide; silver nanoparticle
Year: 2020 PMID: 33282854 PMCID: PMC7691532 DOI: 10.3389/fbioe.2020.603083
Source DB: PubMed Journal: Front Bioeng Biotechnol ISSN: 2296-4185
FIGURE 1TEM images for GO (A), rGO-PDA (B) and rGO-PDA-AgNPs (C). (D) SEM image for rGO-PDA-AgNPs. (E) UV-vis spectroscopy of GO (black) and rGO-PDA-AgNPs (red). (F) XRD patterns of GO (black), rGO-PDA (red) and rGO-PDA-AgNPs (blue).
FIGURE 2Corresponding elemental mapping images of C, N, O, Ag, and Ti.
FIGURE 3Cyclic voltammetric (A) and differential pulse voltammetric (B) curves of the PKA biosensor toward blank (black curve) and 10 U/mL PKA (red curve) in 10 mM Tris-HCl (10 mM NaCl, pH 7.4). (C) Comparison of DPV responses toward 10 U/mL PKA obtained at different control conditions. The rGO-PDA and rGO-PDA-AgNPs were used as the substitutes of the rGO-PDA-AgNPs-Ti4+, respectively. (D) Electrochemical impedance spectroscopy for the differently assembled electrodes in 5 mM [Fe(CN)6]3– /4– of 10 mM PBS buffer (pH 7.4, 1 M KCl) including the bare gold electrode (a), kemptide-immobilized electrode (b), after MCH blocking (c), phosphorylation of kemptide (d), and recognition with the rGO-PDA-AgNPs-Ti4+ nanocomposite (e). EIS was operated with the frequency range from 0.1 Hz to 10 kHz and the amplitude of 5 mV.
FIGURE 4Optimization of experimental conditions. (A) The immobilization concentration of kemptide. Various kemptide concentrations (10, 50, 100, 200, 300, and 500 μM) were applied. (B) ATP concentration optimization. A series of ATP concentrations including 10, 20, 40, 60, 80, 100, and 120 μM were used. (C) The optimization of PKA catalytic time. The employed PKA catalytic time was 0, 10, 30, 60, 90, 120, and 150 min, respectively. (D) The optimization of AgNO3 concentration during the synthesis of rGO-PDA-AgNPs nanocomposite. Various AgNO3 concentration (1, 5, 10, 20, and 50 mM) were studied. The error bars were based on at least three repetitive experiment results. The used PKA concentration was 10 U/mL.
FIGURE 5(A) DPV responses obtained at different concentrations of PKA. The PKA concentrations from curve a to k were 0, 0.01, 0.05, 0.1, 0.2, 0.5, 1, 5, 10, 20, and 50 U/mL, respectively. (B) Calibration curve of DPV response vs. PKA concentration. (C) The linear plot between DPV peak current and PKA concentration. (D) Electrochemical response comparison toward thrombin (1 μM), bovine serum albumin (BSA, 1 μM), Exo III (50 U/mL), hemoglobin (1 mg/mL) and PKA (10 U/mL), respectively.
FIGURE 6(A) Electrochemical detection of PKA in buffer and 5% diluted FBS. (B) Electrochemical responses at different H-89 concentrations. The H-89 concentrations were 5, 10, 25, 50, 100, 200, and 500 nM, respectively. The concentrations for PKA and ATP were 10 U/mL and 100 μM, respectively.