BACKGROUND: Breast compression in mammography is important but is a source of discomfort and has been linked to screening non-attendance. Reducing compression has little effect on breast thickness, and likely little effect on image quality, due to force being absorbed in the stiff juxta thoracic area instead of in the central breast.
PURPOSE: To investigate whether a flexible compression plate can redistribute force to the central breast and whether this affects perceived pain.
MATERIAL AND METHODS: Twenty-eight women recalled from mammography screening were compressed with flexible and rigid plates while retaining force and positioning, 15 in the craniocaudal (CC) view and 13 in the mediolateral oblique (MLO) view. Pressure distribution was continuously measured using pressure sensors.
RESULTS: The flexible plate showed greater mean breast pressure in both views: 2.8 versus 2.3 kPa for CC (confidence interval [CI] = 0.2-0.8) and 1.0 versus 0.5 kPa for MLO (CI = 0.2-0.6). The percentage of applied force distributed to the breast was significantly higher with the flexible plate, both on CC (36% vs. 22%, CI = 1-11) and MLO (30% vs. 14%, CI = 4-13).
CONCLUSION: The flexible plate redistributes pressure to the central breast, achieving a better compression, particularly in the MLO view, though much applied force is still applied to the juxta thoracic region.

Entities:  

Keywords:  Mammography; breast; breast compression; pain

Mesh:

Year:  2020        PMID: 33280392     DOI: 10.1177/0284185120976925

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

1.  Finite element model of mechanical imaging of the breast.

Authors:  Rebecca Axelsson; Hanna Tomic; Sophia Zackrisson; Anders Tingberg; Hanna Isaksson; Predrag R Bakic; Magnus Dustler
Journal:  J Med Imaging (Bellingham)       Date:  2022-05-23

2.  Intra- and inter-rater reliability of compressed breast thickness, applied force, and pressure distribution in screening mammography.

Authors:  Martina Voigt; Anetta Bolejko; Magnus Dustler
Journal:  Acta Radiol Open       Date:  2021-12-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.