Literature DB >> 33279905

One-year anatomical and functional outcomes of the Argus II implantation in Korean patients with late-stage retinitis pigmentosa: A prospective case series study.

Young Hee Yoon, Mark Humayun, Yoon Jeon Kim.   

Abstract

PURPOSE: To report the anatomical and functional outcomes of Argus II retinal prosthesis implantation in Korean patients
Methods: We included five consecutive patients with end-stage retinitis pigmentosa (RP) who underwent Argus II retinal prosthesis implantation and followed for at least 12 months. The transcorneal electrical evoked response was utilized for patient selection. We used intraoperative optical coherence tomography (OCT) for optimal placement of the array and provided specialized vision rehabilitation training. Morphological evaluation using SD-OCT and functional evaluation using computer-based visual function tests (CVFT), a letter reading ability test, and the Functional Low-Vision Observer Rated Assessment (FLORA) were conducted.
RESULTS: Postoperatively, the array was completely apposed to the retinal surface in all eyes except for one eye that had a preexisting macular concavity. Fibrosis-like tissues of ≥ 50-μm thickness developed at the interface in two eyes. All patients showed improvement in CVFT, and could read ETDRS letters at a distance of 50 cm. Three patients could read Korean words. FLORA was improved in all patients, mainly in tasks of visual mobility, daily activities, and social interactions.
CONCLUSIONS: Along with good anatomical outcomes and specialized rehabilitation practices, recipients of the Argus II implant showed profound improvements in functional vision and mobility.
© 2020 S. Karger AG, Basel.

Entities:  

Year:  2020        PMID: 33279905     DOI: 10.1159/000513585

Source DB:  PubMed          Journal:  Ophthalmologica        ISSN: 0030-3755            Impact factor:   3.250


  4 in total

1.  Functional Vision in the Real-World Environment With a Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis.

Authors:  Lewis Karapanos; Carla J Abbott; Lauren N Ayton; Maria Kolic; Myra B McGuinness; Elizabeth K Baglin; Samuel A Titchener; Jessica Kvansakul; Dean Johnson; William G Kentler; Nick Barnes; David A X Nayagam; Penelope J Allen; Matthew A Petoe
Journal:  Transl Vis Sci Technol       Date:  2021-08-12       Impact factor: 3.283

Review 2.  Clinical Progress and Optimization of Information Processing in Artificial Visual Prostheses.

Authors:  Jing Wang; Rongfeng Zhao; Peitong Li; Zhiqiang Fang; Qianqian Li; Yanling Han; Ruyan Zhou; Yun Zhang
Journal:  Sensors (Basel)       Date:  2022-08-30       Impact factor: 3.847

3.  Interobserver Agreement of Electrode to Retina Distance Measurements in a Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis.

Authors:  Carla J Abbott; Elizabeth K Baglin; Maria Kolic; Myra B McGuinness; Samuel A Titchener; Kiera A Young; Jonathan Yeoh; Chi D Luu; Lauren N Ayton; Matthew A Petoe; Penelope J Allen
Journal:  Transl Vis Sci Technol       Date:  2022-09-01       Impact factor: 3.048

4.  Stage-Dependent Changes of Visual Function and Electrical Response of the Retina in the rd10 Mouse Model.

Authors:  Seongkwang Cha; Jungryul Ahn; Yurim Jeong; Yong Hee Lee; Hyong Kyu Kim; Daekee Lee; Yongseok Yoo; Yong Sook Goo
Journal:  Front Cell Neurosci       Date:  2022-07-19       Impact factor: 6.147

  4 in total