Literature DB >> 33279215

Coagulation complications after conversion from roller to centrifugal pump in neonatal and pediatric extracorporeal membrane oxygenation.

Özge Erdem1, Jan Willem Kuiper2, Robert Jan Houmes1, Cornelia Heleen van Ommen3, Joost van Rosmalen4, Dick Tibboel1, Enno Diederik Wildschut1.   

Abstract

BACKGROUND/
PURPOSE: Coagulation complications are frequent, unwanted occurrences in extracorporeal membrane oxygenation (ECMO) treatment, possibly influenced by the pump in the ECMO-circuit. We hypothesized that fewer complications would occur with a smaller, heparin-coated ECMO system with a centrifugal pump (CP) than with one with a roller pump (RP) and that after conversion, complication rates would decrease over time.
METHODS: This single-center, retrospective chart study included all first neonatal and pediatric ECMO runs between 2009 and 2015. Differences between groups were assessed with Mann-Whitney U tests and Kruskal-Wallis tests. Determinants of complication rates were evaluated through Poisson regression models. The CP group was divided into three consecutive groups to assess whether complication rates decreased over time.
RESULTS: The RP group comprised 90 ECMO runs and the CP group 82. Hemorrhagic complication rates were significantly higher with the CP than with the RP, without serious therapeutic consequences, while thrombotic complications rates were unaffected. Intracranial hemorrhage rates and coagulation-related mortality rates were similar. Gained experience with the CP did not improve complication rates or survival over time.
CONCLUSIONS: Although the CP seems safe, it does not seem beneficial over the RP. Further research is warranted on how pump type affects coagulation, taking into account the severity and implications of coagulation complications. LEVEL OF EVIDENCE: Level III.
Copyright © 2020. Published by Elsevier Inc.

Entities:  

Keywords:  Centrifugal pump; Coagulation; Extracorporeal membrane oxygenation; Neonates; Pediatrics; Roller pump

Year:  2020        PMID: 33279215     DOI: 10.1016/j.jpedsurg.2020.11.018

Source DB:  PubMed          Journal:  J Pediatr Surg        ISSN: 0022-3468            Impact factor:   2.545


  2 in total

1.  Computational fluid dynamics analysis and experimental hemolytic performance of three clinical centrifugal blood pumps: Revolution, Rotaflow and CentriMag.

Authors:  Dong Han; Joshua L Leibowitz; Lu Han; Shigang Wang; Ge He; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Nov Technol Devices       Date:  2022-06-19

2.  Use of rotational thromboelastometry to predict hemostatic complications in pediatric patients undergoing extracorporeal membrane oxygenation: A retrospective cohort study.

Authors:  Joppe G Drop; Özge Erdem; Enno D Wildschut; Joost van Rosmalen; Moniek P M de Maat; Jan-Willem Kuiper; Robert Jan M Houmes; C Heleen van Ommen
Journal:  Res Pract Thromb Haemost       Date:  2021-07-14
  2 in total