Abstract
OBJECTIVES: Thanks to its lack of allergic reactions and renal toxicity, CO2 represents an alternative to iodine as a contrast medium for peripheral subtraction angiography. Since CO2 has a lower and negative contrast than iodine, postprocessing DSA and stacking are mandatory. So, it seems that higher doses than traditional iodine angiography are required. We addressed the dosimetric aspects of CO2 angiography for two different commercial DSA-apparatus.Entities:
Keywords: Carbon dioxide contrast medium; X-ray spectrum; digital subtraction angiography
Year: 2020 PMID: 33273772 PMCID: PMC7694728 DOI: 10.4103/ijri.IJRI_247_20
Source DB: PubMed Journal: Indian J Radiol Imaging ISSN: 0970-2016
Figure 1The fluoroscopic suites during the tests (from left to right: ZIEHM and GE)
Specifications of the fluoroscopic suites
| Manufacturer | Name | Power | Focal spot size | Anode angle | Total filtration (mmAl) | Detector type and size |
|---|---|---|---|---|---|---|
| ZIEHM | Vision RFD | 20 kW | 0.3/0.6 mm | 10° | 5 | Flat Panel 30 × 30 cm |
| GE | Innova IGS | 100 kW | 0.3/0.6/1 mm | 11° | 1.8 + 0.2 mmCu | Flat Panel 31 × 31 cm |
Figure 2Examples of dose rate waveforms for ZIEHM (25 fps, iodine protocols and CO2 protocols)
Comparisons between protocols for ZIEHM
| Protocols | kV | mA | Pulse rate (fps) | Pulse length (ms) |
|---|---|---|---|---|
| Iodine | 67 | 58.8 | 25,12.5,8,4 | 20 |
| CO2 | 67 | 58.8 | 25,12.5,8,4 | 20 |
Figure 3Examples of dose rate waveforms for GE Innova (7.5 fps, iodine protocols and CO2 protocols)
Comparisons between protocols for GE Innova
| Protocols | kV | mA | Pulse rate (fps) | Pulse length (ms) |
|---|---|---|---|---|
| Iodine | 83 | 144 | 7.5 | 42 |
| CO2 | 83 | 144 | 7.5 | 42 |
| Iodine | 81 | 160 | 4 | 42 |
| CO2 | 81 | 160 | 4 | 42 |