Literature DB >> 33267671

Study of haemolysis interference limit on serum potassium assay on Roche® Cobas 8000 and evaluation of corrected potassium.

Claudio Ilardo1, Amandine Lancien1, Joël Barthes1.   

Abstract

Roche Diagnostics® decided a more restrictive haemolysis index (≤20 HI) for approval the release of serum potassium results. This study examined the risk of overestimating serum potassium results related to the HI according to the Ricos total error (±5.6%) and evaluated the approach of Martinez-Morillo and Alvarez to determine a corrected potassium. According to Ricos' criteria, our study showed compliant potassium results with HI less than or equal to 75. Between 90 and 100 HI, the results did not show a significant overestimation. The equation to obtain Kcorrected was: Kmeasured - (0.004 × HI). The use of corrective formulas for adjusting results of potassium could help the laboratory to identify patients at increased risk and to repeat the test as soon as possible.

Entities:  

Keywords:  Potassium; analytical interference; correcting results; haemolysis index; overestimation

Mesh:

Substances:

Year:  2020        PMID: 33267671     DOI: 10.1080/00365513.2020.1855363

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  1 in total

1.  The Impact of Hemolysis-Index Thresholds on Plasma and Serum Potassium Measurements.

Authors:  Tiffany Yin; Adrianna Z Herskovits
Journal:  J Appl Lab Med       Date:  2022-05-04
  1 in total