BACKGROUND: Posterior-stabilized and ultracongruent designs are 2 common examples of posterior cruciate ligament-substituting total knee arthroplasty (TKA) prostheses. There is a lack of evidence regarding joint perception differences between ultracongruent and posterior-stabilized TKA. The purpose of this study was to compare the postoperative knee stability, function, and joint perception between ultracongruent and posterior-stabilized TKA in the same patients in order to determine whether there was a difference based on design.
METHODS: This prospective, randomized controlled trial included 50 patients who underwent same-day bilateral TKA utilizing Persona (Zimmer Biomet) posterior-stabilized implants in 1 knee (posterior-stabilized group) and ultracongruent implants in the other knee (ultracongruent group). At 2 years postoperatively, patients were evaluated for knee stability, function, and joint perception. Postoperative static knee anteroposterior stability and femoral rollback were compared. Knee function was evaluated with use of the Knee Society Score and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC). Joint perception was compared by evaluating noise generation, side preference, satisfaction, and the Forgotten Joint Score.
RESULTS: The ultracongruent group had more knee anteroposterior laxity (4.5 mm compared with 1.8 mm in the posterior-stabilized group; p < 0.001) and less femoral rollback (4.2 mm compared with 9.0 mm in the posterior-stabilized group; p < 0.001). Knee Society Score and WOMAC scores were not significantly different between groups (all p > 0.05). Although the ultracongruent group showed less noise generation than the posterior-stabilized group (18% compared with 38%, respectively; p = 0.026), there were no differences in side preference, satisfaction, or Forgotten Joint Score (all p > 0.05).
CONCLUSIONS: Both groups had similar joint perception 2 years postoperatively, with obvious differences in noise generation and more sagittal laxity with less femoral rollback in the ultracongruent group. LEVEL OF EVIDENCE: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.
Copyright © 2020 by The Journal of Bone and Joint Surgery, Incorporated.

Entities:  

Mesh:

Year:  2021        PMID: 33259430     DOI: 10.2106/JBJS.20.00279

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  4 in total

1.  Comparative Analysis of Contemporary Fixed Tibial Inserts: A Systematic Review and Network Meta-analysis of Randomized Controlled Trials.

Authors:  John Krumme; Roma Kankaria; Madana Vallem; John Cyrus; Peter Sculco; Gregory Golladay; Niraj Kalore
Journal:  Orthop Rev (Pavia)       Date:  2022-06-27

2.  Comparison of patient reported outcomes after robotic versus manual total knee arthroplasty in the same patient undergoing staged bilateral knee arthroplasty.

Authors:  Praharsha Mulpur; A B Suhas Masilamani; Mrinal Prakash; Adarsh Annapareddy; Kushal Hippalgaonkar; A V Gurava Reddy
Journal:  J Orthop       Date:  2022-08-19

Review 3.  Comparison of Functional Outcomes, Femoral Rollback and Sagittal Stability of Anterior-Stabilized Versus Posterior-Stabilized Total Knee Arthroplasty: A Systematic Review and Meta-analysis of Randomized Controlled Trials.

Authors:  Balgovind S Raja; Aditya K S Gowda; Sajid Ansari; Arghya Kundu Choudhury; Roop Bhushan Kalia
Journal:  Indian J Orthop       Date:  2021-08-23       Impact factor: 1.033

4.  Similar outcomes including maximum knee flexion between mobile bearing condylar-stabilised and fixed bearing posterior-stabilised prosthesis: a case control study.

Authors:  Jobe Shatrov; Elliot Sappey-Marinier; Moussa Kafelov; Stanislas Gunst; Cécile Batailler; Elvire Servien; Sébastien Lustig
Journal:  J Exp Orthop       Date:  2022-02-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.