BACKGROUND: Data on the arrhythmic burden of women at risk for sudden cardiac death are limited, especially in patients using the wearable cardioverter-defibrillator (WCD).
OBJECTIVE: We aimed to characterize WCD compliance, atrial and ventricular arrhythmic burden, and WCD outcomes by sex in patients enrolled in the Prospective Registry of Patients Using the Wearable Cardioverter Defibrillator (WEARIT-II U.S. Registry).
METHODS: In the WEARIT-II Registry, we stratified 2000 patients by sex into women (n = 598) and men (n = 1402). WCD wear time, ventricular and atrial arrhythmic events during WCD use, and implantable cardioverter-defibrillator (ICD) implantation rates at the end of WCD use were evaluated.
RESULTS: The mean WCD wear time was similar in women and men (94 days vs 90 days; P = .145), with longer daily use in women (21.4 h/d vs 20.7 h/d; P = .001). Burden of ventricular tachycardia or ventricular fibrillation was higher in women, with 30 events per 100 patient-years compared with 18 events per 100 patient-years in men (P = .017), with similar findings for treated and non-treated ventricular tachycardia/ventricular fibrillation. Recurrent atrial arrhythmias/sustained ventricular tachycardia was also more frequent in women than in men (167 events per 100 patient-years vs 73 events per 100 patient-years; P = .042). However, ICD implantation rate at the end of WCD use was similar in both women and men (41% vs 39%; P = .448).
CONCLUSION: In the WEARIT-II Registry, we have shown a higher burden of ventricular and atrial arrhythmic events in women than in men. ICD implantation rates at the end of WCD use were similar. Our findings warrant monitoring women at risk for sudden cardiac death who have a high burden of atrial and ventricular arrhythmias while using the WCD.
Copyright © 2020 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ICD; Outcomes; Sex; Ventricular arrhythmias; Wearable cardioverter-defibrillator

Mesh:

Year:  2020        PMID: 33248269     DOI: 10.1016/j.hrthm.2020.11.025

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  Real life experience with the wearable cardioverter-defibrillator in an international multicenter Registry.

Authors:  Ibrahim El-Battrawy; Boldizsar Kovacs; Tobias C Dreher; Norbert Klein; Stephanie Rosenkaimer; Susanne Röger; Jürgen Kuschyk; Ardan Muammer Saguner; Jacqueline Kowitz; Julia W Erath; Firat Duru; Ibrahim Akin
Journal:  Sci Rep       Date:  2022-02-25       Impact factor: 4.379

Review 2.  Perspectives on Sex- and Gender-Specific Prediction of New-Onset Atrial Fibrillation by Leveraging Big Data.

Authors:  Sven Geurts; Zuolin Lu; Maryam Kavousi
Journal:  Front Cardiovasc Med       Date:  2022-07-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.