Benchmarking different brands of perfluorocarbon liquids.

Joachim H Dresp1.   

Abstract

PURPOSE: To compare the analytical quality characteristics of currently available CE-marked perfluorocarbon liquids (PFCL) applied intraoperatively during vitreoretinal surgery.
METHODS: Twenty-one samples of 8 brands of perfluorooctane (PFO) and 25 samples of 13 brands of perfluorodecalin (PFD) were analysed. Gas chromatography coupled with mass spectroscopy (GC/MS) was used to determine the content of the main product. The amount of reactive and underfluorinated impurities was analysed and expressed as an H-value using fluoride selective potentiometry after a chemical transformation reaction to detect impurities that triggered both acute and latent toxic effects. UV-active substances were determined in order to draw conclusions on the integrity of primary packaging components. Moreover, we controlled for any 1H-PFO contamination in PFO, as it is known to modify PFO's surface characteristics.
RESULTS: Significant differences in the tested products' purity profiles were detected. The PFO batches revealed H-values ranging from < 10 to 1230 ppm and 1H-PFO concentrations ranging from < 1 to 376 ppm. Leachable substances from packaging components cause UV absorption in the 0.1 to > 3 AU range. The PFD batches revealed H-values ranging from < 10 to 70 ppm and leachables from packaging components resulting in absorbances in the 0 AU to 3.2 AU range.
CONCLUSION: The quality characteristics of the analysed PFCL vary significantly, not only among different brands but among batches from the same manufacturer as well. Manufacturers should communicate the purity of their products in an understandable and clear manner. This would require providing a complete certificate of analysis focussing especially on quality characteristics to enable vitreoretinal surgeons to differentiate between the effects from the PFCL itself and those from impurities.

Entities:  

Keywords:  Heavy liquids; Impurity; Retina; Vitrectomy

Year:  2020        PMID: 33245429     DOI: 10.1007/s00417-020-04964-6

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  2 in total

1.  Fibroblast behavior at aqueous interfaces with perfluorocarbon, silicone, and fluorosilicone liquids.

Authors:  J R Sparrow; R Ortiz; P R MacLeish; S Chang
Journal:  Invest Ophthalmol Vis Sci       Date:  1990-04       Impact factor: 4.799

2.  How to Ward Off Retinal Toxicity of Perfluorooctane and Other Perfluorocarbon Liquids?

Authors:  Dirk-Henning Menz; Nicolas Feltgen; Helge Menz; Bernd-Kristof Müller; Thorsten Lechner; Joachim Dresp; Hans Hoerauf
Journal:  Invest Ophthalmol Vis Sci       Date:  2018-10-01       Impact factor: 4.799

  2 in total
  2 in total

Review 1.  Acute intraocular toxicity caused by perfluorocarbon liquids: safety control systems of medical devices.

Authors:  Girish K Srivastava; Vivekanandan Kalaiselvan; Cristina Andrés-Iglesias; Shatrunajay Shukla; Rohit Saxena; Jose Carlos Pastor
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-05       Impact factor: 3.535

2.  Toxicity Threshold of Perfluorocarbon Liquids for Intraocular Use: Dose-Response Assessment of In Vitro Cytotoxicity of Possible Contaminants.

Authors:  Mario Rosario Romano; Claudio Gatto; Laura Giurgola; Eugenio Ragazzi; Jana D'Amato Tóthová
Journal:  Transl Vis Sci Technol       Date:  2021-05-03       Impact factor: 3.283

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.