Mark V Sullivan1,2, William J Stockburn3, Philippa C Hawes4, Tim Mercer5, Subrayal M Reddy1. 1. Research Centre for Smart Materials, Department of Chemistry, School of Natural Sciences, University of Central Lancashire, Preston, PR1 2HE, United Kingdom. 2. Leicester School of Pharmacy, De Montford University, The Gateway, Leicester, LE1 9BH, United Kingdom. 3. Division of Forensic and Applied Sciences, School of Natural Sciences, University of Central Lancashire, Preston, PR1 2HE, United Kingdom. 4. The Pirbright Institute, Ash Road, Pirbright, Woking, Surrey, GU24 0NF, United Kingdom. 5. Jeremiah Horrocks Institute for Mathematics, Physics and Astronomy, School of Natural Sciences University of Central Lancashire, Preston, PR1 2HE, United Kingdom.
Abstract
We have developed a low-cost molecularly imprinted polymer (MIP)-based fluorometric assay to directly quantify myoglobin in a biological sample. The assay uses a previously unreported method for the development of microwave-assisted rapid synthesis of aldehyde functionalized magnetic nanoparticles, in just 20 min. The aldehyde functionalized nanoparticles have an average size of 7.5 nm ± 1.8 and saturation magnetizations of 31.8 emu g-1 with near-closed magnetization loops, confirming their superparamagnetic properties. We have subsequently shown that protein tethering was possible to the aldehyde particles, with 0.25 ± 0.013 mg of myoglobin adsorbed to 20 mg of the nanomaterial. Myoglobin-specific fluorescently tagged MIP (F-MIP) particles were synthesized and used within the assay to capture myoglobin from a test sample. Excess F-MIP was removed from the sample using protein functionalized magnetic nanoparticles (Mb-SPION), with the remaining sample analyzed using fluorescence spectroscopy. The obtained calibration plot of myoglobin showed a linear correlation ranging from 60 pg ml-1 to 6 mg ml-1 with the limit of detection of 60 pg ml-1. This method was successfully used to detect myoglobin in spiked fetal calf serum, with a recovery rate of more than 93%.
We have developed a low-cost molecularly imprinted polymer (MIP)-based fluorometric assay to directly quantify myoglobin in a biological sample. The assay uses a previously unreported method for the development of microwave-assisted rapid synthesis of aldehyde functionalized magnetic nanoparticles, in just 20 min. The aldehyde functionalized nanoparticles have an average size of 7.5 nm ± 1.8 and saturation magnetizations of 31.8 emu g-1 with near-closed magnetization loops, confirming their superparamagnetic properties. We have subsequently shown that protein tethering was possible to the aldehyde particles, with 0.25 ± 0.013 mg of myoglobin adsorbed to 20 mg of the nanomaterial. Myoglobin-specific fluorescently tagged MIP (F-MIP) particles were synthesized and used within the assay to capture myoglobin from a test sample. Excess F-MIP was removed from the sample using protein functionalized magnetic nanoparticles (Mb-SPION), with the remaining sample analyzed using fluorescence spectroscopy. The obtained calibration plot of myoglobin showed a linear correlation ranging from 60 pg ml-1 to 6 mg ml-1 with the limit of detection of 60 pg ml-1. This method was successfully used to detect myoglobin in spiked fetal calf serum, with a recovery rate of more than 93%.
A biomarker is a characteristic that is accurately measured and evaluated as an
indicator of normal biological processes, pathogenic processes, or pharmacologic
responses to a therapeutic intervention [1]. They are usually measured from tissue biopsies or liquid (blood, saliva,
and urine). Biomarkers are critical to the development of drug delivery and medical
devices [2] and are currently being
used in research and clinical practice for:Diagnosing diseases or
predicting risks of diseases.Monitoring healthy people to detect early signs of a
disease.Determining whether a treatment is efficient or
not.Targeting
groups of people for whom a certain drug may be
useful.Producing
safer drugs by predicting the potential adverse effects earlier [3, 4].Traditional biomarkers of myocardial necrosis are cardiac enzymes (kinase), myoglobin
(Mb) and troponin [5-7]. Myoglobin is the earliest biomarker
to appear, following acute myocardial infarction (AMI) [8]. It is an important oxygen carrying monomeric
heme-protein that is found in heart and skeletal muscles, with its main function being
the uptake of oxygen from oxyhemoglobin in the bloodstream [9]. When muscle cells are damaged Mb is rapidly released
into blood circulation. Mb that is determined in serum can be used as an important
biomarker for cardiac injury [10].
The normal range of Mb present in the serum of healthy patients is usually between 0 and
85 ng ml−1 [11]. The
concentration of Mb within serum can quickly become elevated in as little as one hour
after AMI and this correlates with the degree of myocardial injury [12, 13]. Current clinical practises use immunoassays, for
the detection of AMI biomarkers in blood serum, which have relatively low accuracy and
sensitivity, making the monitoring of disease treatment difficult. Work by Zaninotto,
et al show that the BNA and OPUS assay have relatively
high imprecision (>7.4%), as does the Access assay (6.0%–11.0%) [14]. There is, therefore scope for the
development of new and more accurate assays and diagnostics.Fluorescence assays, work by measuring emission intensity generated from a sample, the
emission intensity being directly proportional to the concentration of the fluorophore
present [15]. A typical method used
for biomarker detection in solution is to use a heterogeneous non-competitive assay.
This is when the analyte, within the sample, binds with labeled capture molecules.
Leftover unbound labeled capture molecules are then removed and then the remaining bound
labeled capture molecules are measured. The intensity of the signal is directly
proportional to the concentration of a known analyte [16]. Fluorescence methods are one of the most widespread
label-based approaches used in biomarker detection [17]. This is because of the availability of many
fluorescent probes with diverse spectral properties, use of visible light, and rapid
detection and signal generation. Affinity-based methods (immunoassays and hybridization
assays) that rely on the ability of biomolecules (antibodies and nucleic acids) to
specifically recognize and bind to target molecules, are used in conjunction with
label-based biomarker detection [18].
While, being well-established and highly efficient, this diagnostic method routinely
uses antibodies and enzymes, because of their sensitivity and affinity. Unfortunately,
they still suffer from short shelf-life, high manufacturing cost and relatively poor
stability (especially in organic solvents, and at extreme temperature and pH values)
[19]. Additionally, it can often
be difficult to immobilize antibodies on to the supports that are currently being used
in diagnostic assays [20].In order for a diagnostic test to be efficient, parameters such as specificity, cost,
limits of detection (LOD) and analysis time are very important [21]. As an alternative to the antibody approach,
molecularly imprinted polymers (MIPs) have been used in non-competitive assays and
biosensors, essentially as capture molecules. Molecular imprinting is a technique used
to synthesize template-shaped cavities within a polymer matrix with affinity to a chosen
‘template’ molecule. The process involves initiating the polymerization of monomers in
the presence of a template molecule [22-24]. The template is
then extracted, leaving a polymer with an affinity for the original template molecule
[25, 26]. While MIPs may currently lack sensitivity for
efficient application for early disease diagnosis they can be employed in the screening
and monitoring stages of a disease [27]. MIP-based biosensors have been reported for protein biomarkers,
including PSA, alpha-fetoprotein (AFP), and carcinoembryonic antigen (CEA), with lower
detection limits of 9.6 × 10–3, 1.25 × 10–3, and 8 ng
ml−1, respectively [28-30]. Erturk et al showed that while these LODs are good, they are yet to
compete with similar, antibody-based assays suggesting there is possibility for
improvement [28]. The use of the
whole protein as a template will increase the accuracy of the MIP, although the same
imprinted sites formed for these larger structures could potentially be attractive sites
for similar sized or smaller protein binding (due to other proteins native to a blood
sample), leading to unwanted cross-reactivity and reduced selectivity [31]. When compared with the current
protein detection methods (electrochemistry, ELISA, SPR, and colorimetric and
fluorescence assays), significant advances still need to be made with MIP-based sensors,
with regards to the sensitivity, selectivity and the capacity for multi-analyte
detection [21, 31]. Fluorescence materials can be incorporated into
sensors and fluorescence tagged MIP-based sensors offer a convenient solution due to
increased sensitivity and selectivity, while also offering many advantages such as short
analysis time, ease of use and small sample volume [32]. Adding a fluorescent monomer during MIP synthesis
is potentially an easy way to prepare a fluorescent MIP (F-MIP). For example, Ashley
et al (2018) used fluorescein-O-acrylate as the fluorescent tag, with fluorescent quenching used in the
detection of low molecular weight doxycycline [27, 33]. This suggests that using a fluorescent tag with a MIP imprinted for a
biomarker could be a successful approach.Superparamagnetic iron oxide nanoparticles (SPIONs) are particles of iron oxide with a
diameter between 1 and 100 nm [34,
35] and are available in the two
main forms of: magnetite (Fe3O4) and maghemite (γ-Fe2O3). They are created when
ferrimagnetic, multidomain samples of iron oxide (Fe3O4 and
γ-Fe2O3) particles are reduced to
sizes less than around 100 nm [35].
SPIONs are able to show robust paramagnetic nature, with high sensitivity under the
appliance of a magnetic field and complete disappearance of that nature once the
magnetic field is removed; thus, removing any chance of magnetic remanence and
coercivity. Due to their superparamagnetic properties and potential non-toxicity to
humans, there has been an extensive amount of interest in the use and synthesis of
SPIONs, especially within biological and medicinal applications. For biomedical
applications these include magnetic resonance imaging, magnetic fluid hyperthermia,
bioseparation (DNA, proteins and cells), biomolecule immobilization, and magnetic
targeted drug delivery; amongst others [36-45]. SPIONs offer
essential properties including magnetic susceptibility, low cytotoxicity, good
biocompatibility, and stability under physiological environments [46].In order for SPIONs to interact with bioactive substances, they are usually
functionalized with a reactive group, typically amino (–NH2), aldehyde (–CHO)
or carboxylic acid (–COOH) [47, 48]. These modified SPIONs can be used
to bind to biomolecules, such as proteins, through non-covalent (hydrogen bonding)
interactions but primarily through covalent linkages [47, 48]. They are preferred for use in biomedicine as they are biocompatible and
potentially non-toxic to humans [46].
Iron oxide is easily degradable and can therefore be useful when used for in vivo applications [49] such as cancer magnetic therapy [50, 51]. Here the iron oxide nanoparticles are loaded with
antitumor drugs and remotely directed via an external magnetic field. This allows for
more accurate, localized targeting systems, as well as further enhancing the antitumor
activity by moderate magnetic hyperthermia (below 40 °C) [40].There have been several different methods used to produce SPIONs of a controlled size
and shape with solvothermal methods proving particularly successful [52, 53]. However, solvothermal methods can be multi-step
processes and time consuming (>8 h), producing only small yields of functionalized
particles [48, 54, 55]. Recently, there has been a surge in interest in the application of
microwave radiation as a thermodynamic driving force, with the prospect of developing
environmentally conscious, simple and time efficient routes towards the synthesis of
nanomaterials [56].In this paper a heterogeneous non-competitive assay was developed for the detection of
the AMI biomarker myoglobin. A F-MIP was created for the target molecule (myoglobin),
with this MIP being used as a capture molecule within an assay for Mb. Previous work
showed that acrylamide was an effective monomer to use within a MIP for myoglobin,
because the monomer is water soluble, produces excellent rebinding, with minimal effect
on the secondary structure of the protein [57]. The fluorescent monomer, fluorescein O-acrylate was chemically bound during myoglobin MIP synthesis to allow for
fluorescence spectroscopy detection. The resulting F-MIP was added to a sample
containing the target molecule, whereby the target molecule would bind to the MIP.
Excess MIP would then be removed by the addition of myoglobin modified SPIONs,
synthesized using a previously unreported microwave-assisted one-step 20 min synthesis.
These particles would essentially bind to any excess MIP, which could then be separated
using a magnet leaving only the target protein bound MIPs in suspension. The F-MIPs in
suspension (bound to target protein) were then analyzed using fluorescence spectroscopy,
whereby the intensity of the signal is proportional to the amount of target protein in
the sample.
Experimental
Materials
7-[4-(Trifluoromethyl)coumarin]methacrylamide), acetonitrile, acrylamide (AAm),
ammonium persulphate (APS), anhydrous sodium acetate (NaOAc), bovine serum albumin
(BSA) ethanol, ethyl acetate, ethylene glycol, ferric chloride hexahydrate
(FeCl3 · 6H2O), fetal calf serum (FCS), fluorescein O-acrylate, glacial acetic acid (AcOH), glutaraldehyde,
myoglobin (Mb) (from equine skeletal muscle), N N,
N′-methylenebisacrylamide (mBAm), phosphate buffered saline (PBS), sodium
dodecyl sulfate (SDS), tetramethylethyldiamide (TEMED), were all purchased and used
without purification from Sigma-Aldrich, Poole, Dorset, UK.
Instrumentation
BioDrop μLITE UV/visible spectrometer was purchased
from Biochrom Ltd Cambridge, UK. Horiba JY FluorMax – 4 spectrofluorometer was
purchased from Horiba UK Limited, Northampton, UK. The 6 kOe vibrating sample
magnetometer (VSM) was built in-house at the University of Central Lancashire,
Preston, UK. Nicolet AVATAR 330 FTIR spectrophotometer with Pike MIRacle accessory
and FEI Tecnai 12 TEM at 100 kV with a Tietz F214 2k × 2k CCD camera were purchased
from Thermo Fisher Scientific, Loughborough, UK. Bruker D2 Phaser Bench top X-ray
diffractometer was purchased from Bruker UK Limited, Coventry, UK. Anton Paar
monowave 200 microwave oven was purchased from Anton Paar Ltd Hertfordshire, UK. SLS
Lab basics centrifuge was purchased from Scientific Laboratory Supplies, Nottingham,
UK.
Methods
Synthesis of fluorophore tagged MIPs
Solution preparation
A solution of 10% (w/v):10% (v/v) SDS:AcOH was prepared for use in the washing
(protein elution) stages before the template reloading stage. SDS (10 g) and
AcOH (10 ml) was dissolved in 990 ml of deionised (DI) water, to produce 1 l of
the washing solution.
MIP preparation
Bulk MIP hydrogels were produced, using an optimized methodology [58], where a 10% cross-linking
monomer/N,N′-methylenebisacrylamide hydrogel was found to produce the optimal
imprint for Mb, in terms of specificity and rebinding efficiency of the MIP,
compared with the non-imprinted polymer (NIP) [58].
Fluorescein MIP synthesis
Into an Eppendorf tube, 12 mg of myoglobin template was dissolved in 435
μl of deionised water vortexed for 1 min,
followed by the addition of 135 μl
(7.6 × 10–4 mol) of 40% AAm solution, 300 μl (3.9 × 10–5 mol) mBAm (cross-linker), and 100
μl of 1 × 10–3 mMol fluorescein
O-acrylate solution (monomer: crosslinker:
f-monomer molar ratio was 760:39:1), then vortexed for a further minute).
Finally, 10 μl of a 5% TEMED (v/v) solution and
20 μl 5% APS (w/v) solution were added and the
mixture was vortexed for 1 min. Solutions were purged with nitrogen for 5 min
and polymerization occurred overnight at room temperature (∼20 °C).
Corresponding NIPs were produced using the same method, but in the absence of a
protein template.After polymerization, the gels were granulated separately, using a 35 μm sieve. The refined gels were washed with five 1 ml
volumes of deionised water followed by five 1 ml volumes of 10% (w/v):10% (v/v)
SDS:AcOH eluent; this allowed for the removal of the template protein from the
MIP cavities. Following this, the gels were washed with five 1 ml volumes of
deionised water to remove all residual 10% (w/v):10% (v/v) SDS:AcOH from the
MIP gels. Each wash step was followed by centrifugation, whereby the gels were
vortexed then centrifuged (using SLS Lab Basics centrifuge) for 5 min at 15 000
rpm (RCF: 15 100 × g). Corresponding NIPs were synthesized using the same
procedure as the MIPs, but in the absence of the template molecule.
MIP rebinding studies
The subsequent protein rebinding ability of the conditioned and equilibrated
MIPs and NIPs were characterized using the BioDrop μLITE UV/visible spectrometer. Hydrogels (200 mg) were then each
treated with 400 μl of a 3 mg ml−1
myoglobin (template protein) solution. The polymer/protein solutions were mixed
on a rotary vortex mixer and allowed to associate at room temperature (∼20 °C),
then washed four times with 1 ml of deionized water. Each reload and wash step
for hydrogels was followed by centrifugation for 5 min at 15 000 rpm (RCF: 15
100 × g). All supernatants were collected for analysis by spectrophotometry (at
405 nm for myoglobin).Selectivity studies were performed on the MIP using the protein, bovine serum
album (BSA). This performed using the previous method, but substituting the 400
μl of a 3 mg ml−1 myoglobin
(template protein) solution, with a 400 μl of a 3
mg ml−1 bovine serum albumin (non-template protein) solution.
Myoglobin functionalized SPIONs
Preparation of functionalized SPION
The functionalized SPIONs were prepared through a one-pot solvothermal
microwave method. FeCl3ˑ6H2O was used as a single iron
source and glutaraldehyde as a modification ligand. With stirring, 0.5 g of
FeCl3ˑ6H2O and 1.8 g of NaOAc were dissolved in 15 ml
of ethylene glycol in a 30 ml Anton Parr G30 microwave reaction vial (MRV). The
functionalization agent (glutaraldehyde (3.5 ml) was then added to the
resulting solution and stirring continued for a further 5 min. The magnetic
stirrer bar was then removed and the MRV was placed into an Anton Paar monowave
200 microwave oven and the reaction was heated up to a temperature of 200 °C
with a ramp time of 2 min. The reaction was held at 200 °C for 20 min under
pressure (9 bar). The resulting composite products were washed five times with
deionised water followed by two washes of ethanol, and then collected with a
magnet and finally dried for further use. The method was repeated, but in the
absence of a functionalization agent, to produce bare SPIONs.
Protein conjugation to functionalized SPION
The protein adsorption capacity of the obtained magnetic particles was
investigated using Mb as the model protein. Five Mb solutions of 0.2, 0.3, 0.4,
0.5, 0.6 mg ml−1 were prepared and 1 ml of each protein solution was
added to 20 mg of the magnetic material. The mixture was vortexed followed by
shaking. We determined that the optimum time for this step was 30 min.
Afterwards, the mixture was then centrifuged (SLS Lab basics centrifuge) for 5
min at 15 000 rpm (RCF: 15 100 × g). The amount of protein adsorbed onto the
SPIONs (functionalized and bare) was calculated through comparing the initial
and final concentrations of protein remaining in the supernatant. The
concentration of the non-adsorbed protein was measured by spectrophotometry
(405 nm for Mb) using a BioDrop μLITE UV/visible
spectrometer.
SPION characterization
Functionalized nanoparticles were suspended in ultra-pure water (0.1 g in 50
μl water) and a 5 μl droplet was deposited onto a Formvar/carbon coated 200 mesh
copper TEM grid (Agar Scientific, UK). After 1 min the grid was blotted, washed
for 30 s in ultra-pure water, blotted again and allowed to dry. Images were
collected using a FEI Tecnai 12 TEM at 100 kV with a Tietz F214 2k × 2k CCD
camera. Fourier transform infrared (FTIR) spectra were obtained on a Nicolet
AVATAR 330 FTIR spectrophotometer fitted with a Pike MIRacle accessory. The
FTIR was recorded using transmission mode and the spectrum was collected at
room temperature in 4000–400 cm−1 region with a resolution of 4
cm−1 using 32 scans. The crystal structure information was
recorded using a Bruker D2 Phaser Bench top x-ray diffractometer system
equipped with a LynxEye detector and using Cu Kα
12 radiation (λ = 1.5418 Å). Data were
collected between 5° and 80° 2-theta using a step size of 0.02019°. Pawley fits
were performed using the TOPAS program. Initial lattice parameters for ETS-10
and quartz were taken from Anderson et al [59] and le Page and Donnay
[60], respectively. A
12-term Chebyshev background function and a pseudo-Voight peak profile function
were all used for refinements. Magnetization curves were measured at room
temperature using an in-house 6 kOe VSM. The samples were first crushed with a
pestle and mortar, to break-up the large agglomerations of the dry state. This
enabled them to be packed into rectangular glass microslides (Camlab) of
internal dimensions 0.40 ± 0.04 mm by 4.0 ± 0.1 mm that were cut in lengths of
10.5 ± 0.5 mm. In this geometry the demagnetization factor (N), was kept low and within the range
0.030 < N < 0.068 [61].
F-MIP assay
A MIP-based non-competitive assay was designed for the detection of the AMI
biomarker myoglobin. The assay uses an excess amount of F-MIP as a capture
molecule, to bind to all of the myoglobin within the sample. Mb-SPIONs were then
used to bind to all the remaining unbound MIP, which was then subsequently removed
with magnetism. The MIP left in the sample was then removed and analyzed using
fluorescent spectroscopy, with fluorescent intensity directly related to the
amount of myoglobin in the sample.
Non-competitive assay calibration
Solutions containing known concentrations of Mb (6, 0.6, 6 × 10–2,
6 × 10–3, 6 × 10–4, 6 × 10–5,
6 × 10–6, 6 × 10–7, 6 × 10–8, and
6 × 10–9 mg) were prepared in 1 ml of PBS, followed by the
addition of 100 mg of F-MIP/NIP to the sample. This was then vortexed for 30 s
and left for 5 min to allow the protein to associate with the imprinted or
non-imprinted gel. Next, 30 mg (optimum amount) of Mb-SPION was added, followed
by further vortexing (30 s) and protein association (5 min). MIP/NIP particles
bound to the Mb-SPION were removed using a magnet, and the supernatant then
analyzed using a Horiba JY FluorMax – 4 spectrofluorometer with an excitation
at 490 nm and an emission scan range of 200–900 nm. A calibration curve was
created by plotting the log of the amount of protein loaded (mg
ml−1) versus log intensity (CPS). The calibration was repeated, but
with FCS used instead of PBS.To test the assay, 1 ml samples of FCS were spiked with known concentrations of
Mb (10, 80 and 160 ng ml−1). These values were chosen because it
would allow for the detection of Mb within the normal range of healthy
patients, as well as showing the range on and above the borderline for elevated
levels (above 85 ng ml−1) seen in patients [11]. Next, 100 mg of F-MIP/NIP was added to the
sample and vortexed for 30 s and left for 5 min to allow the protein to
associate with the imprinted gel. Next, 30 mg (optimum amount) of Mb-SPION was
added, followed by further vortexing (30 s) and protein association (5 min).
MIP particles bound to the Mb-SPION were removed using a magnet, and the
supernatant then analyzed using fluorescence spectroscopy with excitation at
490 nm and an emission scan range of 200–900 nm. The results were compared with
the calibration plot in order to evaluate the error within the assay.
Results and discussion
MIP synthesis and rebinding studies
Fluorophoric MIPs (F-MIPs) and their corresponding NIPs were successfully synthesized
using the optimized method [58].
Rebinding studies were performed on these and the results are presented in table
1. This confirmed that there was
no detrimental effect in the ability of target molecule to bind to MIP when
integrating the fluorophore into the MIP/NIP synthesis.
Table 1.
Percentage of the myoglobin target protein (Mb) rebind to the acrylamide-based
MIP and NIP with incorporated fluorophore and the corresponding imprinting
factor as well as percentage of bovine serum albumin non-target protein (BSA)
bind to MIP.
Fluorophore
MIP percentage of Mb rebind
(%)
NIP percentage of protein bind
(%)
Imprinting factor (IF)
MIP percentage of BSA bind
(%)
Fluorescein O-acrylate
85.8 ± 2.9
43.9 ± 2.7
1.95
42.4 ± 2.3
Percentage of the myoglobin target protein (Mb) rebind to the acrylamide-based
MIP and NIP with incorporated fluorophore and the corresponding imprinting
factor as well as percentage of bovine serum albumin non-target protein (BSA)
bind to MIP.Our previous work, showed that a polyacrylamide-based MIP for the target protein
myoglobin had a rebinding capacity , of 85.4% with an IF value of 1.80, suggesting
that the MIP has a good degree of affinity with a high degree of selectivity compared
to the corresponding NIP and other acrylamide-based functional monomers [57]. When the fluorophore was
integrated into a polyacrylamide-based MIP, the amount of myoglobin target that
rebound was 85.8%. This shows that the fluorophore integrated MIPs produce rebinding
characteristics which are very similar to that of the polyacrylamide MIP (±1.6%)
[57], indicating that there is
little to no effect on rebinding of the target protein, when integrating the
fluorophore into the MIP. These results offered assurance that a fluorophore
integrated MIP can be used to develop a fluorometric MIP-based assay for myoglobin.
Table 1 also shows that the Mb
templated MIP will bind 42.4% of non-specific BSA. This value is very similar to the
binding capacity of the NIP, confirming that the Mb MIP has selectivity towards the
target protein.
Rapid microwave-assisted synthesis of functionalized SPIONs
Microwave-assisted synthesis is based on the accelerated heating of materials due to
dielectric heating effects. The microwave energy produced is only transferred
directly to those reaction components, which are susceptible to microwave
polarization. This improves the energy efficiency, by only heating the reaction
mixture and reduces the need to heat any reaction vessels, unlike conventional
heating. As a result of this directed heating, heating the reagents is much faster
than with conventional methods. This, in turn, minimizes the time it takes for the
reaction to reach its activation energy and can radically reduce the reaction time,
with the added benefit that this method reduces unwanted side reactions and
by-products [62].Ethylene glycol was the solvent used in this reaction and is known to have a high
dissipation factor (tan δ = 1). The solvent has a
higher capacity to absorb microwaves and convert to thermal energy, making it ideal
for microwave synthesis [63].
Producing functionalized SPIONs by a conventional solvothermal method [48] takes 480 min for the reaction to
be complete, whereas our microwave-assisted method only takes 20 min. Different
synthesis times were trialed, before settling upon 20 min; 10 min did not yield
nanoparticles. Whereas 30 min was successful in producing nanoparticles, we selected
20 min synthesis for further studies.The TEM images for SPION@CHO, and SPION (bare) are shown in figures 1(A), and (B), respectively. The obtained
nanoparticles are nearly spherical and dispersive with a diameter, ranging from 1.6
to 11.6 nm. The microwave method, generally produces particles that are smaller (7.5
and 7.4 nm, for SPION@CHO and SPION (bare) respectively) than the particles produced
using a solvothermal method (19.3 nm) [48], with a summary of all the particles shown in table 2. The SPION@CHO (figure 1(A)) appears to show small agglomerates
of SPIONs embedded within a polymeric matrix, which could account for the larger size
of the individual particles within each agglomerate. The bare SPIONs (figure 1(B)) shows particles that appear
consistently spherical shape and similar in size.
Figure 1.
TEM images of SPION@CHO (A) and SPION (bare) (B). Images were collected using a
FEI Tecnai 12 TEM at 100 kV with a Tietz F214 2k × 2k CCD camera.
Table 2.
A comparison for the reaction methods to produce functionalized SPION between a conventional solvothermal method
versus our microwave method.
Reaction time (min)
Average particle size (nm)
Standard deviation (nm)
Solvothermal method [48]
480
19.3
1.8
Microwave method (CHO)
20
7.5
1.8
Microwave method (bare)
20
7.4
1.4
TEM images of SPION@CHO (A) and SPION (bare) (B). Images were collected using a
FEI Tecnai 12 TEM at 100 kV with a Tietz F214 2k × 2k CCD camera.A comparison for the reaction methods to produce functionalized SPION between a conventional solvothermal method
versus our microwave method.The FTIR spectra of the functionalized SPIONs are shown in figures 2(A) and (B), where the main functional
groups of the predicted structure can be observed. Figure 2(A) shows the FTIR spectrum for the aldehyde
functionalized SPION (SPION@CHO). The absorption band at 520 cm−1 is
assigned to the Fe–O stretching vibration. The peak at 1635 cm−1 is
assigned to the stretching vibration of carbonyl, while the typical peak at 2956
cm−1 is due to the asymmetric stretching of C–H vibration. These
characteristic peaks indicate that iron oxide magnetic particles containing aldehyde
groups on the surface have been obtained through our one-pot microwave facilitated
solvothermal method [48]. Figure
3(B) shows an absence of any
functional groups peaks, with the only peak observed being the absorption band at 538
cm−1, which is assigned to the Fe–O stretching vibration. The absence
of any other peaks again highlights that fuctionalization has taken place in figure
2(A).
Figure 2.
FTIR spectra of SPION@CHO (A), and SPION (bare) (B), obtained using a Nicolet
AVATAR 330 FTIR spectrophotometer fitted with a Pike MIRacle accessory.
Figure 3.
XRD pattern of SPION@CHO (A) and SPION (bare) (B), obtained using a Bruker D2
Phaser Bench top x-ray diffractometer system equipped with a LynxEye detector
and using Cu Kα
12 radiation (λ = 1.5418 Å).
FTIR spectra of SPION@CHO (A), and SPION (bare) (B), obtained using a Nicolet
AVATAR 330 FTIR spectrophotometer fitted with a Pike MIRacle accessory.XRD pattern of SPION@CHO (A) and SPION (bare) (B), obtained using a Bruker D2
Phaser Bench top x-ray diffractometer system equipped with a LynxEye detector
and using Cu Kα
12 radiation (λ = 1.5418 Å).The XRD patterns for the synthesized SPIONs and functionalized SPIONs produce the
same six strong diffraction peaks, which is to be expected as XRD only determines the
atomic and molecular structure of a crystal and does not include any information on
the functionalization. These XRD patterns are illustrated in figure 3, where the six strong diffraction peaks
for the sample are observed in the 2θ range of 20°–80°,
which are indexed as (220), (311), (400), (422), (511), and (440), respectively.
Figures 3(A) and (B) can be seen to be
very similar in nature and this is shown in the analysis of the XRD data. The
crystallite size of the of the SPIONs was estimated using the Scherrer equation
(equation (1)), which
relates the crystallite size (D) and a specific diffraction peak broadening [64, 65]:where, D
is the average crystallite size, K is the
Scherrer constant (0.94), β
311 is line broadening (full-width at half maximum, in radians) and
θ
311 is the Bragg angle (radian). The estimated crystallite sizes for the
Fe3O4 NPs are 7.47 nm and 7.45 nm for the SPION@CHO and
SPION (bare), respectively, and is consistent with the size estimates using TEM
(table 2). The lattice parameters
‘a’ were determined to be 8.3656 and 8.3547 for the
SPION@CHO and SPION (bare), respectively, with these values being similar to those
found in literature (a = 8.399) [66]. These results are consistent with iron oxide
found in the inorganic crystal structure database (ICSD Collection Code 5247) [67], and confirms that we have
produced Fe3O4 (magnetite) [68, 69].The magnetic properties of the functionalized SPIONs were studied using a VSM at room
temperature (figures 4(A) and (B)).
The magnetization curves with S-like shape are symmetrical to the origin and there is
near-zero hysteresis, with both remanence and coercivity approximately zero,
indicating that the samples are predominantly superparamagnetic. The saturation
magnetization values of 38 emu g−1 (SPION@CHO) and 75 emu g−1
(SPION (bare)) are lower than the ferrimagnetic bulk value of 92 emu g−1
[70] and is expected when in a
superparamagnetic state at particle sizes <100 nm, as is the case here [71]. However, the values are still
high enough for easy and rapid separation when under an external magnetic field. The
TEM image of the functionalized SPIONs (figure 2(A)) shows the magnetic particles are distributed
within the nonmagnetic functionalized material and hence this dilutes the magnetic
content fraction within the sample volume when compared to the bare SPIONs and will
contribute to the lower saturation magnetization value observed [72].
Figure 4.
VSM spectra of SPION@CHO (A) and SPION (bare) (B), measured at room temperature
using an in-house 6 kOe Vibrating Sample Magnetometer.
VSM spectra of SPION@CHO (A) and SPION (bare) (B), measured at room temperature
using an in-house 6 kOe Vibrating Sample Magnetometer.We further investigated the ability of these particles to covalently bind protein.
This was with a view to use protein tethered SPIONs in the development of our F-MIP
based protein diagnostic. Protein adsorption onto the SPION@CHO was investigated by
adding 1 ml of known concentrations of a myoglobin solution. Prior to this, we had
determined that a minimum of 30 min was required for optimum adsorption of 0.4 mg
ml−1 of protein (figure 5(A)). The results showed that on average, 20 mg of SPION@CHO was able to
adsorb 0.247 ± 0.013 mg of myoglobin, respectively (figure 5(B)). In contrast, binding of protein on
non-functionalized SPIONs, showed only limited protein adsorption (0.012 ± 0.004 mg).
This suggests that the mechanism for protein attachment to the aldehyde
functionalized SPION is primarily through covalent binding; glutaraldehyde has been
used extensively in the immobilization of proteins [73]. The high reactivity of glutaraldehyde towards
proteins is based on the presence of several reactive residues in proteins and
molecular forms of glutaraldehyde in aqueous solution, which may lead to many
different possible reaction mechanisms, including, but not limited to, aldol
condensation or Michael-type addition [74].
Figure 5.
Optimum adsorption time for Mb binding to 20 mg of SPION@CHO particles (A) and
adsorption capacity (B) of the SPIONs. Measured using spectrophotometry (405 nm
for Mb) using a BioDrop μLITE UV/visible
spectrometer.
Optimum adsorption time for Mb binding to 20 mg of SPION@CHO particles (A) and
adsorption capacity (B) of the SPIONs. Measured using spectrophotometry (405 nm
for Mb) using a BioDrop μLITE UV/visible
spectrometer.
Optimization of protein-SPION level for assay development
The fluorescently tagged MIP and NIP were explored using fluorescent spectroscopy to
determine if protein binding had an effect on the fluorescent emission of the sample.
As can been seen in figure S1, in the supplementary data (available online at
stacks.iop.org/NANO/32/095502/mmedia), the results show that the F-MIP
has an emission intensity value which is lower than that of the F-NIP. This is to be
expected as a F-MIP molecule contains cavities, resulting in less polymer matrix
within the same volume of particle (35 μm particle),
meaning there is less polymer within the F-MIP for the light to be absorbed,
therefore resulting in less photons being emitted; consequently, lowering the
emission intensity. Polymer matrix differences between MIP and NIP were explored in
the work of Larpant et al (2019) who showed that the
MIP contained a matrix that was highly absorbent, with strong affinity of the target
molecule [75]. The resulting
changes in polymer matrix explains the changing in absorbance, with work by Pelras
et al (2017) showing changes to a hydrogel matrix
can increase transparency [76].
The fluorescence spectra (figure S2), for all concentrations of the fluorescein
O-acrylate monomer, show a peak at 482 nm, which is
due to excitation of the sample. The energy from the emitted photon is responsible
for the peak seen at 514 nm. As the concentration of the fluorescein monomer is
diluted further the intensity of the peak at 514 nm steadily decreases as to be
expected, with the fluorescence intensity being proportional to the concentration of
the fluorophore.In constructing the non-competitive MIP-based assay for the detection of the AMI
biomarker Mb, aldehyde functionalized SPIONs (SPION@CHO) were used, with myoglobin
(Mb) bound to the nanoparticle (Mb-SPION). These nanoparticles are nearly spherical
and dispersive with a diameter, averaging 7.5 nm, with a saturation magnetization
value of 38 emu g−1, meaning they can be easily and rapidly separated from
suspension under an external magnetic field.The amount (30 mg) of the Mb-SPION conjugate needed to bind to all of the F-MIP was
first determined. This would then give a value for the maximum amount of Mb-SPION
required per assay, and would simulate the situation of there being zero target Mb in
the test solution. Masses (0–50 mg) of Mb-SPION were loaded into 1 ml of water
containing 100 mg of fluorescein O-acrylate
(1 × 10–3 mmol) MIP. A magnet was used to remove the F-MIP bound to the
Mb-SPION complex. Any remaining unbound MIP left in solution was analyzed by
fluorescence spectroscopy. The limiting amount of Mb-SPION needed to bind to all of
the F-MIP was determined as the point when the measured fluorescence emission
intensity petered out and remained constant (at about the same intensity as when
measuring water only). Figure 6 shows
that as the amount of Mb-SPION is increased, the fluorescence emission intensity of
the sample decreases. This pattern is of exponential decay resulting from an initial
rapid decrease in intensity, which steadily reaches a minimum at and above 30 mg of
Mb-SION with an intensity of 734 490 CPS, which is of similar intensity of a sample
of water only. It was determined that 30 mg of Mb-SPION was sufficient to bind all of
the F-MIP in the solution, so this amount was selected be used within the assay.
Figure 6.
A plot showing how much Mb-SPION is needed to bind and remove all of the
fluorescently tagged MIP from a 1ml 1:10 solution of fluorescein O-acrylate
(1 × 10–3 mMol) MIP:H2O.
A plot showing how much Mb-SPION is needed to bind and remove all of the
fluorescently tagged MIP from a 1ml 1:10 solution of fluorescein O-acrylate
(1 × 10–3 mMol) MIP:H2O.
Non-competitive assay calibration
Using a non-competitive assay format, a calibration was undertaken by dissolving
known concentrations of target protein Mb (6, 0.6, 6 × 10–2,
6 × 10–3, 6 × 10–4, 6 × 10–5,
6 × 10–6, 6 × 10–7, 6 × 10–8, and
6 × 10–9 mg) into FCS (1 ml). FCS was used, with known Mb
concentrations dissolved, in order to determine the selectivity and biocompatibility
of the F-MIP, investigating whether non-target molecules present within the serum
would interfere with the assay, leading to the production of false positives. FCS
contains many different proteins (BSA, Immunoglobulin G (IgG), fibrinogen,
transferrin, amongst others), with the main constituent being albumin, [77]. The F-MIP was added to the
sample, with the target protein being allowed to associate with the MIP for 10 min.
Next, 30 mg of Mb-SPION was added to the sample and further associated with any
unbound MIP was allowed for 10 min MIP particles bound to the Mb-SPIONs were removed
using a magnet, and the supernatant then analyzed using fluorescence spectroscopy.
Calibration plots were created by plotting the log10 of the amount of
protein loaded (mg ml-1) versus log10 intensity (CPS) and are
shown in figure 7. The calibration
plots show that the assay has a linear range with a lower LOD of 60 pg
ml−1. This is a vastly wider range than that needed for the detection
of Mb in patient samples (0–80 ng ml−1). That we can use this MIP-based
assay to measure protein levels down to pg ml−1, makes it highly suitable
for the potential development of a suite of assays for a range of biomarkers which
require a low limit of detection including cancer markers such as PSA (<4 ng
ml−1), CEA (<2.5 ng ml−1), and AFP (<10 ng) [11, 78]. Calibrations were produced in PBS buffer and FCS
showing that the assay is specific only for the target protein Mb and shows minimal
interference from other proteins (native to serum).
Figure 7.
Calibration plots for the non-competitive assay, by comparing the log amount of
protein loaded (mg ml−1) in fetal calf serum (FCS) (blue) and
phosphate buffered saline (PBS) (orange), with the log fluorescence emission
intensity (CPS).
Calibration plots for the non-competitive assay, by comparing the log amount of
protein loaded (mg ml−1) in fetal calf serum (FCS) (blue) and
phosphate buffered saline (PBS) (orange), with the log fluorescence emission
intensity (CPS).
F-MIP-based assay for analysis of samples
The MIP-based assay was further tested by spiking 3 × 1 ml samples of FCS with known
concentrations of Mb (10, 80 and 160 ng ml−1). The results of the spike
serum test are shown in table 3.
Table 3.
Analysis of Mb spiked serum concentration using the developed non-competitive
assay.
Mb spiked concentration (ng
ml−1)
Fluorescent emission intensity
(CPS)
Amount of Mb calculated in sample
(ng)
Percentage recovery (%)
10
2164570 ± 1180
9.8
97.79
80
2782603 ± 1125
85.0
93.66
160
2970330 ± 1210
149.0
93.31
Analysis of Mb spiked serum concentration using the developed non-competitive
assay.The results in table 3 show that
when FCS is spiked with known concentrations of the AMI biomarker Mb, the MIP-based
non-competitive assay was able to detect the Mb to within an error of 6.7%. This is
an acceptable percentage error as it is similar to the percentage error of other
assays which are commercially available and currently being used [14]. A comparison of these assays
with the one developed is shown in table 4.
Table 4.
A comparison of different assays available against the developed MIP-based
assay.
Assay
Percentage error (%)
Developed MIP Assay
2.2–6.7
Accessa
6.0–11.0
Hitachia
3.8–5.8
Stratusa
3.4–6.5
Data for these assay kits supplied by the work of Zaninotto et al (2000) [14].
A comparison of different assays available against the developed MIP-based
assay.Data for these assay kits supplied by the work of Zaninotto et al (2000) [14].Table 4 shows that the developed MIP
assay performs slightly better (with a percentage error range of 2.2%–6.7%) than the
Access assay (with a percentage error range of 6.0–11.0). The results also show that
the developed MIP assay performs very similarly to the Hitachi (with a percentage
error range of 3.8%–5.8%) and Stratus (with a percentage error range of 3.4–6.5).
This shows that the developed MIP assay is competitive with those that are currently
on the market. All the market assays (Access, Hitachi, and Stratus) use Mb antibodies
as the capture molecule [79].This uniquely developed MIP assay uses a fluorescence spectroscopy for detection of
myoglobin from a spiked sample. Fluorescently tagged MIP is added to the sample
containing the target molecule, whereby the target molecule would bind to the MIP.
Excess MIP would then be removed by the addition of the myoglobin modified SPIONs,
produced using the unreported microwave method that synthesized aldehyde
functionalized SPIONs in 20 min. Previous studies synthesized these particles using
non-microwave solvothermal synthesis, by either a one-pot synthesis or multi-step
synthesis, often taking in excess of 8 h to produce [48, 52, 53]. This novel
synthesis dramatically decreases the time needed to produce functionalized SPIONs
(SPION@CHO), thus using less energy and reagents, and making this synthesis much
greener than traditional methods. These particles would bind to the excess MIP, which
could then be separated using a magnet leaving only the target protein bound MIPs in
suspension. The F-MIPs in suspension could then be analyzed using fluorescence
spectroscopy, where by the intensity of the signal is proportional to the amount of
target protein in the sample, as shown in figure 8.
Figure 8.
A schematic showing how the non-competitive assay works: (A) is the sample of
target protein. (B) Excess amount of the fluorescently tagged MIP is added and
binds to all the target protein. (C) Excess amount of target protein-SPION is
added, this binds to all of the unbound MIP in the sample. (D) MIP bound to
target protein-SPION is removed by magnetism. (E) After removal of MIP bound to
the target protein-SPION, MIP bound to the target protein in the sample is left
to be analyzed.
A schematic showing how the non-competitive assay works: (A) is the sample of
target protein. (B) Excess amount of the fluorescently tagged MIP is added and
binds to all the target protein. (C) Excess amount of target protein-SPION is
added, this binds to all of the unbound MIP in the sample. (D) MIP bound to
target protein-SPION is removed by magnetism. (E) After removal of MIP bound to
the target protein-SPION, MIP bound to the target protein in the sample is left
to be analyzed.
Conclusion
In this work, we have demonstrated a microwave facilitated, one-pot solvothermal method
for the synthesis of functionalized SPIONs. This method allowed for the rapid synthesis
of functionalized SPIONs with time to production being reduced from many hours to a mere
20 min This significant reduction in synthesis time, shows that functionalized SPION
particles can be produced using much less energy than previous solvothermal synthesis
methods. A non-competitive MIP-based fluorometric assay was subsequently developed for
the AMI biomarker Mb. The assay used a fluorophore tagged MIP as a capture molecule,
binding specifically to myoglobin, enabling its detection and quantitative analysis.
Calibration of the assay was conducted in FCS showing that the assay was specific only
to the target protein. The assay has a low LOD, capable of detecting down to 60 pg
ml−1 of Mb, suggesting the assay would be suitable for detection of Mb
within patients where 0–80 ng ml−1 is the normal range for Mb levels in
healthy patients and elevated levels (>80 ng ml−1) are seen in patients
where AMI has occurred Analyzing FCS spiked with known concentrations of Mb demonstrated
that our MIP-based assay can detect Mb within a biological sample with an error range of
only 2.2%–6.7%. This is a similar range to antibody-based assay kits that are available
on the market [14] and demonstrating
that our developed MIP-based assay is able to compete with assay kits that are already
available, but at a fraction of the cost. MIP-based protein assays can be a real
alternative to antibody-based assays with major implications in the future development
of disease diagnostics, especially in instances where antibodies are either difficult to
produce or do not currently exist.
Authors: Alessandro Poma; Antonio Guerreiro; Michael J Whitcombe; Elena V Piletska; Anthony P F Turner; Sergey A Piletsky Journal: Adv Funct Mater Date: 2013-06-13 Impact factor: 18.808
Authors: Felismina T C Moreira; M Judite M S Ferreira; José R T Puga; M Goreti F Sales Journal: Sens Actuators B Chem Date: 2016-02 Impact factor: 7.460
Authors: H F El Sharif; S R Dennison; M Tully; S Crossley; W Mwangi; D Bailey; S P Graham; S M Reddy Journal: Anal Chim Acta Date: 2022-04-01 Impact factor: 6.911