Comparison of the diagnostic value of virtual touch tissue quantification and virtual touch tissue imaging quantification in infants with biliary atresia.

Yong-Fang Liu1, Xian-Wei Ni1, Yin Pan1, Hong-Xia Luo1.   

Abstract

OBJECTIVE: This study aimed to compare the diagnostic performance of vital touch tissue quantification (VTQ) and virtual touch tissue imaging quantification (VTIQ) in diagnosing infants with biliary atresia (BA) from jaundiced infants.
METHODS: In this study, 26 jaundiced infants with BA, 33 jaundiced infants without BA, and 40 normal infants were enrolled. The hepatic shear wave velocity (SWV) of each infant was determined by VTQ and VTIQ examinations, respectively. Then, the receiver operating characteristic (ROC) curves were drawn and the area under the curve (AUC) and optimal cut-off values were calculated to evaluate the sensitivities and specificities of VTIQ and VTQ for BA.
RESULTS: The mean values of SWV of the control group measured by VTQ and VTIQ were (1.09 ± 0.18) m/s and (1.36 ± 0.21) m/s, respectively. The mean values of SWV of the non-BA group measured by VTQ and VTIQ were (1.30 ± 0.28) m/s and (1.52 ± 0.29) m/s, respectively. The mean values of SWV of the BA group measured by VTQ and VTIQ were (2.36 ± 0.36) m/s and (2.43 ± 0.29) m/s, respectively. The diagnostic threshold of VTQ and VTIQ to diagnose BA was 1.77 and 1.92 m/s. The sensitivities of VTQ and VTIQ to diagnose BA were 90.9% and 95.5%. The specificities of VTQ and VTIQ to diagnose BA were 68.4% and 78.9%.
CONCLUSION: Vital touch tissue quantification and VTIQ could help distinguish infants with BA from jaundiced infants by measuring the liver SWV values. VTIQ has higher sensitivity and specificity than VTQ.
© 2020 John Wiley & Sons Ltd.

Entities:  

Mesh:

Year:  2020        PMID: 33236451     DOI: 10.1111/ijcp.13860

Source DB:  PubMed          Journal:  Int J Clin Pract        ISSN: 1368-5031            Impact factor:   2.503


  3 in total

1.  Ultrasound elastography in the diagnosis of biliary atresia in pediatric surgery: a systematic review and meta-analysis of diagnostic test.

Authors:  Yanqiang Li; Jinghua Jiang; Hong Wang
Journal:  Transl Pediatr       Date:  2022-05

2.  Use of shear wave elastography for the diagnosis and follow-up of biliary atresia: A meta-analysis.

Authors:  Ellen S Wagner; Hussien Ahmed H Abdelgawad; Meghan Landry; Belal Asfour; Mark B Slidell; Ruba Azzam
Journal:  World J Gastroenterol       Date:  2022-08-28       Impact factor: 5.374

Review 3.  Ultrasound for the Diagnosis of Biliary Atresia: From Conventional Ultrasound to Artificial Intelligence.

Authors:  Wenying Zhou; Luyao Zhou
Journal:  Diagnostics (Basel)       Date:  2021-12-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.