Literature DB >> 33227272

Toward a Standard Practice to Quantify von Willebrand Factor Degradation During Left Ventricular Assist Device Support.

Samson Hennessy-Strahs1, Christian A Bermudez1, Michael A Acker1, Carlo R Bartoli2.   

Abstract

BACKGROUND: Continuous-flow left ventricular assist devices (LVADs) cause degradation of von Willebrand factor (VWF) multimers and bleeding. Multiple techniques exist to characterize VWF deficiency. However, a standard methodology has not been established in LVAD patients. Toward this goal, we evaluated 4 methods to quantify VWF multimers.
METHODS: We collected paired blood samples from patients (n = 48) before and after 1 week of LVAD support. After 652 ± 59 days of support, patients were classified as bleeders (≥1 bleeding episode) or nonbleeders. VWF multimers were resolved with electrophoresis and immunoblotting, the gold-standard to evaluate VWF multimers. We evaluated 4 quantification methods.
RESULTS: Each method demonstrated significant VWF degradation during LVAD support vs a paired, pre-LVAD sample (method 1, VWF length: 48 of 48 patients, -10% ± 1%, P < .0001; method 2, VWF density: 40 of 48, -34% (interquartile range, -46% to -8%), P < .0001; method 3, pre-LVAD to LVAD ratio: 46 of 48, 17 ± 5: 10 ± 1, P < .0001; method 4, LVAD/pre-LVAD index: 46 of 48, 57% (interquartile range, 50% to 73%), P < .0001). Bleeding occurred in 27 of 48 patients. Method 1 demonstrated significantly fewer VWF multimers in bleeders compared with nonbleeders (-11% ± 1% vs -8% ± 1%; P = .01). Other methods did not demonstrate this potentially important clinical relationship.
CONCLUSIONS: A standardized methodology is needed to quantify VWF multimer degradation with mechanical circulatory support devices. Novel method 1 successfully quantified the patient-specific change in VWF multimer length during LVAD support and demonstrated a difference in VWF multimers between bleeders and nonbleeders. Adoption of consensus methodology will assist to standardize patient-specific bleeding risk, inform anticoagulation and antiplatelet therapy, and evaluate LVAD hemocompatibility.
Copyright © 2021 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2020        PMID: 33227272     DOI: 10.1016/j.athoracsur.2020.09.039

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Pathologic von Willebrand factor degradation is a major contributor to left ventricular assist device-associated bleeding: pathophysiology and evolving clinical management.

Authors:  Carlo R Bartoli
Journal:  Ann Cardiothorac Surg       Date:  2021-05

2.  Abnormalities in the Von Willebrand-Angiopoietin Axis Contribute to Dysregulated Angiogenesis and Angiodysplasia in Children With a Glenn Circulation.

Authors:  Carlo R Bartoli; Samson Hennessy-Strahs; Robert D Dowling; J William Gaynor; Andrew C Glatz
Journal:  JACC Basic Transl Sci       Date:  2021-03-22
  2 in total