Radiation dose reduction capabilities of a new C-arm system with optimized hard- and software.

Domagoj Javor1, Johanna Moyses2, Christian Loewe2, Ruediger E Schernthaner2.   

Abstract

PURPOSE: To assess the radiation dose reduction capabilities and the image quality of a new C-arm system in comparison to a standard C-arm system.
METHOD: Prospective, randomized, IRB approved two-arm trial design. 49 consecutive patients with primary or secondary liver cancer were treated with transarterial chemoembolization (TACE) on two different angiography units. 28 patients were treated on a conventional angiography unit B, 21 patients on unit A which provides improved hardware and optimized image processing algorithms. Dose area product (DAP) and fluoroscopy time were recorded. DSA image quality of all procedures was assessed on a four-rank-scale by two independent and blinded readers.
RESULTS: Both cohorts showed no significant differences with regard to patient characteristics, tumor burden and fluoroscopy time. The new system resulted in a statistically significant reduction of cumulative DAP of 72% compared to the old platform (median 76 vs. 269 Gy*cm2). Individually, Fluoro-DAP and DSA-DAP decreased by 48% and 77% (p = 0.012 and p < 0.01), respectively. No statistically significant differences in DSA image quality were found between the two imaging platforms.
CONCLUSIONS: The new C-arm system significantly reduced radiation exposure for TACE procedures without increased radiation time or negative impact on DSA image quality. The combination of optimized hardware and software yields the highest radiation dose reduction and is of utmost importance for patients and interventionalists.
Copyright © 2020 The Authors. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Interventional Radiology; Liver Neoplasms; Radiation Protection; Therapeutic Chemoembolization

Mesh:

Year:  2020        PMID: 33221618     DOI: 10.1016/j.ejrad.2020.109367

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures.

Authors:  Kelly S Wilson-Stewart; Davide Fontanarosa; Eva Malacova; Jamie V Trapp
Journal:  Phys Eng Sci Med       Date:  2022-05-09

2.  Comparison of radiation doses between hepatic artery infusion chemotherapy and transarterial chemoembolization for liver cancer.

Authors:  Hui Yuan; Hailei Lu; Jiahuan Zeng; Yan Zhang; Lujun Shen
Journal:  J Interv Med       Date:  2021-08-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.