Ear-EEG for sleep assessment: a comparison with actigraphy and PSG.

Yousef Rezaei Tabar1, Kaare B Mikkelsen2, Mike Lind Rank3, Martin Christian Hemmsen3, Marit Otto4, Preben Kidmose2.   

Abstract

PURPOSE: To assess automatic sleep staging of three ear-EEG setups with different electrode configurations and compare performance with concurrent polysomnography and wrist-worn actigraphy recordings.
METHODS: Automatic sleep staging was performed for single-ear, single-ear with ipsilateral mastoid, and cross-ear electrode configurations, and for actigraphy data. The polysomnography data were manually scored and used as the gold standard. The automatic sleep staging was tested on 80 full-night recordings from 20 healthy subjects. The scoring performance and sleep metrics were determined for all ear-EEG setups and the actigraphy device.
RESULTS: The single-ear, the single-ear with ipsilateral mastoid setup, and the cross-ear setup performed five class sleep staging with kappa values 0.36, 0.63, and 0.72, respectively. For the single-ear with mastoid electrode and the cross-ear setup, the performance of the sleep metrics, in terms of mean absolute error, was better than the sleep metrics estimated from the actigraphy device in the current study, and also better than current state-of-the-art actigraphy studies.
CONCLUSION: A statistically significant improvement in both accuracy and kappa was observed from single-ear to single-ear with ipsilateral mastoid, and from single-ear with ipsilateral mastoid to cross-ear configurations for both two and five-sleep stage classification. In terms of sleep metrics, the results were more heterogeneous, but in general, actigraphy and single-ear with ipsilateral mastoid configuration were better than the single-ear configuration; and the cross-ear configuration was consistently better than both the actigraphy device and the single-ear configuration.
© 2020. Springer Nature Switzerland AG.

Entities:  

Keywords:  Ambulatory sleep monitoring; Ear-EEG; Long-term sleep monitoring; Outpatient sleep monitoring

Mesh:

Year:  2020        PMID: 33219908     DOI: 10.1007/s11325-020-02248-1

Source DB:  PubMed          Journal:  Sleep Breath        ISSN: 1520-9512            Impact factor:   2.816


  2 in total

Review 1.  International classification of sleep disorders-third edition: highlights and modifications.

Authors:  Michael J Sateia
Journal:  Chest       Date:  2014-11       Impact factor: 9.410

2.  Accurate whole-night sleep monitoring with dry-contact ear-EEG.

Authors:  Kaare B Mikkelsen; Yousef R Tabar; Simon L Kappel; Christian B Christensen; Hans O Toft; Martin C Hemmsen; Mike L Rank; Marit Otto; Preben Kidmose
Journal:  Sci Rep       Date:  2019-11-14       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.