Jun Hu1, Guosheng Yang2, Chutima Kranrod1, Kazuki Iwaoka2, Masahiro Hosoda1,3, Shinji Tokonami1.
Abstract
An improved passive CR-39-based direct 222Rn/220Rn progeny detector with 3 detection channels was designed and tested in this study to measure and calculate equilibrium equivalent concentration (EEC) of both 222Rn and 220Rn without the equilibrium factor. A theoretical model was established to calculate the EEC with optimization. Subsequently, an exposure experiment was carried out to test the performance of this detector, and we compared the chamber experiment and the theoretical model by estimating and measuring various parameters. The deposition flux of progeny derived from the prediction agreed well with the value measured in the exposure chamber. The energy-weighted net track density (NTD) measured by this detector is much more reliable to reflect the linear relation between NTD and time-integrated EEC. Since the detector is sensitive to the exposure environmental condition, it is recommended to apply the detector to measure the EEC after its calibration in a typical indoor environment.Entities:
Keywords: 220Rn progeny; 222Rn progeny; CR-39; deposition velocity; equilibrium equivalent concentration
Mesh:
Substances:
Year: 2020 PMID: 33218216 PMCID: PMC7699185 DOI: 10.3390/ijerph17228569
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1The passive CR-39-based direct 222Rn/220Rn progeny detector: (a) a schematic diagram of the progeny detector showing the detectable radionuclides of each channel and (b) a photo of the actual setup of the progeny detector.
The design information of each detector channel.
| Channel | Membrane Thickness | Nuclide | Energy of α-Particle | Track Registration Efficiency, | Deposited Atoms |
|---|---|---|---|---|---|
| TnP | 7.10 | 212Po | 8.785 | 0.063 | 212Pb, 212Bi |
| RnPII | 5.05 | 212Po | 8.785 | 0.212 | 212Pb, 212Bi |
| 214Po | 7.687 | 0.182 | 218Po, 214Pb, 214Bi | ||
| RnPI | 3.25 | 212Po and 212Bi | 8.785 and 6.051 | 0.152 | 212Pb, 212Bi |
| 214Po | 7.687 | 0.249 | 218Po, 214Pb, 214Bi | ||
| 218Po | 6.003 | 0.135 | 218Po |
Figure 2A schematic diagram of the chamber verification system.
The comparison results of the progeny detector and experiment in a 220Rn chamber verification system.
| Model | Experiment | Model | Experiment | ||
|---|---|---|---|---|---|
| Exposure time (h) | 25.5 | 36.9 | |||
| The number concentration (N cm−3) | 27,297 | 30,417 | |||
| GSD, σg | 1.506 | 1.613 | |||
| CMD (μm) | 0.203 | 0.203 | |||
| Track density | Upward | 3.9 | 2.9 | ||
| Vertical | 4.3 | 4.8 | |||
| Deposition velocity | Upward | 0.185 | 0.143 ± 0.029 | 0.243 | 0.193 ± 0.067 |
| Vertical | 0.169 | 0.158 ± 0.040 | 0.238 | 0.217 ± 0.045 | |
| 212Pb | 0.185 | —— | 0.219 | —— | |
| 212Bi | 0.132 | —— | 0.183 | —— | |
| Deposition flux | Upward | 0.081 | 0.081 | 0.055 | 0.069 |
| Vertical | 0.090 | 0.090 | 0.042 | 0.042 | |
| EEC of 220Rn | Upward | 237 ± 88 | 244 ± 108 1 | 148 ± 38 | 155 ± 67 1 |
| Vertical | 263 ± 88 | 188 ± 50 | |||
1 The value was calculated by the average equilibrium equivalent concentration (EEC) of 220Rn over the whole exposure period by grab sampling method.
The comparison results of the progeny detector and experiment in a 222Rn chamber verification system.
| Model | Experiment | Model | Experiment | Model | Experiment | |||
|---|---|---|---|---|---|---|---|---|
| Exposure time (h) | 30 | 70 | 102 | |||||
| Concentration of nuclei number (N cm−3) | 34,528 | 33,833 | 33,278 | |||||
| GSD, σg | 1.714 | 1.761 | 1.768 | |||||
| CMD (μm) | 0.233 | 0.200 | 0.191 | |||||
| Track density | Upward | RnPI | 6.3 ± 1.3 | 16.4 ± 2.5 | 14.9 ± 2.3 | |||
| RnPII | 4.5 ± 1.2 | 6.2 ± 1.6 | 4.9 ± 1.1 | |||||
| Vertical | RnPI | 7.4 ± 1.5 | 13.6 ± 2.1 | 17.2 ± 2.7 | ||||
| RnPII | 1.1 ± 0.8 | 7.1 ± 1.8 | 5.9 ± 1.3 | |||||
| Deposition velocity | Upward | 0.136 | —— | 0.123 | —— | 0.125 | —— | |
| Vertical | 0.131 | —— | 0.117 | —— | 0.119 | —— | ||
| 218Po | 0.290 | —— | 0.248 | —— | 0.230 | —— | ||
| 214Pb | 0.155 | —— | 0.135 | —— | 0.112 | —— | ||
| 214Bi | 0.101 | —— | 0.091 | —— | 0.092 | —— | ||
| Deposition flux | 218Po to RnPI | Upward | 0.041 | 0.051 ± 0.010 | 0.022 | 0.048 ± 0.017 | 0.018 | 0.030 ± 0.016 |
| Vertical | 0.043 ± 0.009 | 0.040 ± 0.014 | 0.035 ± 0.019 | |||||
| 218Po, 214Pb, 214Bi to RnPII | Upward | 0.014 | 0.005 ± 0.004 | 0.011 | 0.014 ± 0.008 | 0.008 | 0.007 ± 0.006 | |
| Vertical | 0.023 ± 0.006 | 0.016 ± 0.009 | 0.009 ± 0.007 | |||||
| EEC of 222Rn | Upward | 1214 ± 322 | 740 ± 27 1 | 730 ±290 | 693 ±16 1 | 758 ± 370 | 686 ± 31 1 | |
| Vertical | 267 ± 187 | 603 ±312 | 630 ± 375 | |||||
1 The value was calculated by the average EEC of 222Rn over the whole exposure period by the continuous portable Si-photodiode detector.
Summary of the deposition velocities of 222Rn and 220Rn progenies in previous studies.
| Reference | Deposition Velocity (m h−1) | Remark | |
|---|---|---|---|
| RnP Detector | TnP Detector | ||
| [ | 8.64 | 19.08 | Fan on, 26 m3 test facility |
| 3.24 | 2.16 | Fan off, 26 m3 test facility | |
| [ | 0.19 ± 0.04 | Dwelling, the same structure of TnP detector as this study | |
| [ | 0.132 ± 0.0036 | 0.075 ± 0.0072 | 22.5 m3 test room |
| 2.37 ± 0.785 | 0.5 m3 chamber | ||
| 0.117 | 0.072 | Dwelling | |
| [ | 0.828 ±0.072 | 150 L chamber, same TnP detector as this study | |
| This study | 0.125 ± 0.007 | 0.178 ± 0.045 | 3.24 m3 chamber |
Figure 3The net track density (NTD) as a function of time-integrated equilibrium equivalent concentration (EEC): (a) the NTD as a function of time-integrated EEC of 222Rn, where the NTD has excluded the contribution of other radionuclides on the CR-39, and (b) the NTD as a function of time-integrated EEC of 220Rn.