Ion concentration measurement using synthetic microfluidic papers.

Haruka Kamiya1, Hiroki Yasuga1,2, Norihisa Miki1.   

Abstract

Non-invasive diagnosis on biological liquid samples, such as urine, sweat, saliva, and tears, may allow patients to evaluate their health by themselves. To obtain accurate diagnostic results, target liquid must be precisely sampled. Conventionally, urine sampling using filter paper can be given as an example sampling, but differences in the paper structure can cause variations in sampling volume. This paper describes precise liquid sampling using synthetic microfluidic papers, which are composed of obliquely combined micropillars. Sampling volume accuracy was investigated using different designs and collection methods to determine the optimal design and sample collecting method. The optimized protocol was followed to accurately measure potassium concentration using synthetic microfluidic paper and a commercially available densitometer, which verified the usefulness of the synthetic microfluidic papers for precision sampling.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 33211718      PMCID: PMC7676646          DOI: 10.1371/journal.pone.0242188

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Medical diagnosis that is performed beside a patient and produces results in a rapid manner has been attracting considerable attention, also referred to as Point-of-Care Testing (POCT) [1]. As one of POCT fashions, there is a non-invasive testing performed by a patient oneself through small medical devices. Such devices typically diagnose liquid samples obtained from human bodies such as blood, sweat, saliva, tears and urine. Among them, blood provides more information than the other sample types. Despite the difficulty for patients themselves to access blood without the help of medical professionals, glucose monitoring for diabetic patients is one of the few examples currently used [2, 3]. Sweat also provides some physiological information. Microfluidic devices using colorimetric analysis [4] and biosensors using flexible circuit boards [5] have been developed for measuring metabolites, such as glucose and lactate, and electrolytes, such as sodium and potassium ion, in sweat. Saliva is easy to handle and sample. Several biomarkers associated with mental stress and viral hepatitis are found in saliva [6, 7]. The diagnosis on the saliva depends on the oral cavity that is easily affected by food and drink. Tears have been reported to have a glucose concentration similar to blood, which is refreshed every time a person blinks [8, 9]. Collecting tears is challenging, but sensors for contact lens that will enable continuous monitoring are being developed [10]. Urine tests have been used to detect pregnancy [11], diabetes [12], and cancers [13] because urine sampling is readily accessible. Process of the POCT performed by a patient oneself involves liquid sampling, followed by diagnosis through sensing/measuring. When patients perform the liquid sampling, a method must be simple and, more importantly, able to sample accurately even when patients are unskilled. Given as an example of liquid sampling, sampling urine has been conducted by using a paper strip, which is typically made from filter paper [14, 15]. The filter paper-based strips are low cost, easy to handle and capable of retaining liquid samples by capillary force [16]. However, the cotton fibers that compose the filter papers are randomly intertwined, which may lead to volume variation. Microfabrication can offer structures composed of regular micro-patterns, such as micropillar arrays [17], microchannels [18], and synthetic microfluidic paper (SMP) [19], which are potential in the use for precise sampling. In particular, the SMP which is three-dimensional, regularly ordered microstructures of a photosensitive polymer pumps and maintains liquid volumes with excellent accuracy and repeatability. This paper describes the usefulness of SMP for precise liquid sampling. Initially, we evaluated sampling precision and retention of sampled liquid using the SMP in comparison to conventional filter papers, some part of which were differently analyzed based on our previous results in [20]. Then, we conducted potassium concentration measurement from aqueous solution as shown Fig 1. In these experiments, we supposed urinalysis for a patient of implantable artificial kidney [21]. The implantable artificial kidney requires monitoring of blood potassium concentration of patient because an abnormal blood potassium concentration can cause heart arrhythmias, cardiac arrest, or mental confusion. Blood potassium concentration can be monitored through the urine since the filtrate, has a similar potassium concentration as that in the blood [22].
Fig 1

Conceptual sketch of the ion concentration measurement system.

A precise amount of the urine is sampled by a strip made of synthetic microfluidic paper and the potassium ion concentration is measured using a commercially available ion densitometer.

Conceptual sketch of the ion concentration measurement system.

A precise amount of the urine is sampled by a strip made of synthetic microfluidic paper and the potassium ion concentration is measured using a commercially available ion densitometer.

Materials and methods

Chemicals

The following chemicals were used without further purification: isobutyl acetate, acetone, 2-hydroxyethyl methacrylate (HEMA), isopropyl alcohol (IPA), benzophenone, potassium chloride (KCl) (Wako Pure Chemical Industries, Ltd., Japan) and Irgacure819 (BASF Corp., Germany). Ultra-purified water was prepared using a Millipore system (Direct-Q UV3) (Merck KGaA, Germany). A photomask for multidirectional exposure was purchased from Tokyo Process Service Co., Ltd., Japan. Transparent sheets were purchased from Sakae Technical Paper Co., Ltd., Japan. Slide glass (S9111), micro slide glass (S3131), and micro cover glass thickness No. 1 (24×50 mm) were obtained from Matsunami Glass Ind., Ltd., Japan. A potassium ion densitometer (LAQUA twin K-11) was purchased form Horiba, Ltd., Japan. Filter paper used for sampling was Whatman No. 4 (GE Healthcare Japan, Japan). The sampling strip was based on SMP [19]. The SMP was made of off-stoichiometry thiol-ene (OSTE, OSTEMER 220 Litho, Mercene Labs AB, Sweden). The OSTE is a negative photoresist reported previously [23, 24], and is a viscous liquid at room temperature (15°C~25°C). Precursors of OSTE are two monomers, which contains thiol groups and ene (allyl) groups, with a photo-initiator. The thiol-ene reaction triggered by UV exposure induces covalent bonds between the two monomers, resulting in a transparent OSTE polymer [24]. The OSTE contains unreacted thiol groups on the polymer surfaces, which allows surface modification based on polymer-chain grafting [19].

Fabrication process

SMP made of photosensitive polymer, OSTE, is a paper-like substrate composed of micropillar structures that are combined obliquely, as shown in Fig 2(A). The internal structure of the SMP is characterized by the pillar-to-pillar pitch (p), pillar diameter (d), pillar height (h), and pillar angle (θ). The SMP has many pillars combined with each other and is more robust against collapse than the structures containing vertical straight pillars [19]. This feature enables a pillar matrix with a high aspect ratio (= h/d) and a large surface area. These synthetic microfluidic papers allow various flow rates in capillary pumping depending on the structure and have shown smaller deviation than the nitrocellulose-based papers [6].
Fig 2

Internal structure of synthetic microfluidic paper (SMP).

(a) SMP is defined by the pillar-to-pillar pitch (p), pillar diameter (d), pillar height (h) and pillar angle (θ). The value of pitch (p) over diameter (d) represents pitch ratio (r). (b) A photograph of the strip having SMP and a close-up SEM photograph of the SMP.

Internal structure of synthetic microfluidic paper (SMP).

(a) SMP is defined by the pillar-to-pillar pitch (p), pillar diameter (d), pillar height (h) and pillar angle (θ). The value of pitch (p) over diameter (d) represents pitch ratio (r). (b) A photograph of the strip having SMP and a close-up SEM photograph of the SMP. The SMP was fabricated by multi-directional photolithography, as shown in Fig 3. First, a rectangular-shaped frame with 1-mm thick OSTE was prepared (Fig 3A). The OSTE precursor was poured into this frame, and a photomask was set. As shown in Fig 3B, the precursor was exposed to UV light from four directions through the mask using a collimated UV lamp (EMA-400, Union Optical Co., Ltd., Japan) to form obliquely combined micropillars. A stage containing four slanted aluminum mirrors was used for the multidirectional exposure, as shown in Fig 3B. The OSTE was developed in isobutyl acetate or acetone under ultrasonication (FU-10C, TGK, Japan) (Fig 3C). Then, the surface of the micropillars was hydrophilized by photo-grafting of HEMA (Fig 3D). The entire structure was exposed to UV light for 300 s while immersed into IPA containing HEMA (5 wt%) and the two photopolymerization initiators, Irgracure819 (0.5 wt%) and benzophenone (0.5 wt%). Then, the device was developed again using pure IPA for 300 s, followed by drying (Fig 3E). For this study, six types of SMP with a thickness of 1 mm were prepared (Table 1) with various pitches and diameters. For the experiments, a strip containing the SMP was used, as shown in Fig 2(B).
Fig 3

Fabrication process of synthetic microfluidic paper.

(a) Manufacturing setup; (b) Multidirectional UV exposure to form obliquely combined micro pillars; (c) Development under ultrasonication; (d) Surface hydrophilization; (e) Development and drying.

Table 1

Dimensions of fabricated synthetic microfluidic papers.

No.Diameter (d) [μm]Pitch (p) [μm] (r = Pitch/Diameter)
175112.5 (1.5)
275150 (2.0)
375225 (3.0)
4100150 (1.5)
5100200 (2.0)
6100300 (3.0)

Fabrication process of synthetic microfluidic paper.

(a) Manufacturing setup; (b) Multidirectional UV exposure to form obliquely combined micro pillars; (c) Development under ultrasonication; (d) Surface hydrophilization; (e) Development and drying.

Experiments

Precision sampling experiment

To optimize the design of the SMP and the method for precise sampling, sampling experiments were conducted with six types of strips (Table 1) using 3 methods: (a) sprinkling method, (b) immersion method I, and (c) immersion method II (Fig 4). 6 types of SMPs were prepared in different fabrication condition as in Table 1, and the SMP in each condition was tested using a same paper strip five times (n = 5); sample volume was measured from the weight of the strip before and after sampling. Water was used as the liquid during the experiments. Strip dimensions were 20×10 mm. Detailed procedures for each method are described below.
Fig 4

Precision sampling experiment.

(a) Sprinkling method: Water is sprinkled onto the strip directly. (b) Immersion method I: The strip is immersed for 3 s. (c) Immersion method II: The strip is swirled 2 strokes during a 5-s immersion.

Precision sampling experiment.

(a) Sprinkling method: Water is sprinkled onto the strip directly. (b) Immersion method I: The strip is immersed for 3 s. (c) Immersion method II: The strip is swirled 2 strokes during a 5-s immersion. (a) Sprinkling method: Water was sprinkled onto the strip for 3 s at a rate of 60 mL/min with a smooth flow pump (Tacmina Corp., Japan). This method simulates a patient directly sprinkling urine on a strip to collect a urine sample. (b) Immersion method I: A strip was placed vertically into a water bath, immersed for 3 s, and then removed. This method simulates a patient immersing a strip into urine collected in a container. (c) Immersion method II: A strip was placed vertically into a water bath, moved using 2 strokes in 5 s, and then removed. This simulates a patient immersing a strip into urine and swirling the strip before removing it. All experiments including following experiments were conducted at approximately 23°C and humidity was kept at 50±5%.

Sample retention test

As a patient handles a strip, drops of urine could fall off the strip, leading to measurement errors. Therefore, the ability of the strips to retain the urine sample was investigated. During the experiments, a mechanical impact was applied to the strip, as shown in Fig 5. A strip containing a sample was bent by 10 mm, then, it was released abruptly. The weight before and after release was measured. 6 types of SMPs were prepared in different fabrication condition as in Table 1, and the SMP in each condition was tested using a same paper strip five times (n = 5).
Fig 5

Sample retention test.

A ruler was fixed and aligned with the edge of the strip. Then, the tip of the strip was bent by 10 mm and released. Weight of the strip before and after the release was measured.

Sample retention test.

A ruler was fixed and aligned with the edge of the strip. Then, the tip of the strip was bent by 10 mm and released. Weight of the strip before and after the release was measured.

Comparison between synthetic microfluidic paper and filter paper

To verify the usefulness of the SMP for precision sampling, we compared SMP with d = 75 and p = 150 μm and filter paper in terms of the precision sampling tests by immersion method I and sample retention test. Each strip was cut out into 20×8 mm without any supports and 10 replicates (N = 10) were prepared for these tests. In order to avoid the variation of sampling volume, we used vertically arranged linear actuator that was controlled by microcomputer to pulling strip out of water. Each strip was fixed to a movable stage by stabbing it at the top 1mm by a fishhook-like stainless wire. The pulling speed was set as 10 mm/s. Sample volume was measured from the weight of the strip before and after sampling. After the sampling, sample retention test was conducted sequentially.

Measurement of potassium ion concentration

To see if the SMP can be used for measurement system using a paper strip, we conducted potassium concentration measurement using SMP, supposing urinalysis for a patient of implantable artificial kidney. The implantable artificial kidney being developed by our group has a mechanism whereby the filtrate generated by hemofiltration is conveyed to the bladder, turning into urine [21]. Our previous studies showed that this filtrate is equivalent to blood plasma with respect to waste and electrolytes [21]. Since the blood plasma is mostly composed of water and includes the electrolyte component about 1% and its pH does not deviate significantly from 7, we determined using an aqueous potassium chloride solution as a model urine for potassium measurement in this study. Potassium ion concentration contained in the fabricated strips was measured using a commercially available potassium ion densitometer. The instrument contains an ion selective electrode, which measures specific ion concentration in a sample solution through a potential difference between the ion selective electrode and the reference electrode, as shown in Fig 6(A) [25, 26]. The potassium ion densitometer could accept a sample by direct injection into the instrument sample port or by placing a sample paper immersed with the sample solution into the instrument. Four sampling methods were investigated, as shown in Fig 6(B) (N = 30 for each method): (i) directly injecting the sample solution using a pipette (control experiment); (ii) placing a SMP strip with sample solution into the instrument; (iii) placing a filter-paper strip into the instrument; and (iv) placing a strip (BY046, Horiba, Ltd., Japan) tailored for the densitometer into the instrument. The SMP had d and p values of 75 and 150 μm, respectively. The sample solutions contained potassium ion concentrations of 60, 140, 200, and 360 ppm, which were prepared by dissolving 60, 140, 200, and 360 mg of potassium chloride in 1 L of ultra-purified water, respectively. Less than 10 seconds were taken for each measurement, which resulted from automatic measurement process of the densitometer used.
Fig 6

Measurement of potassium ion concentration.

(a) Measurement mechanism of the densitometer. (b) Measurement methods: (i) Direct injection of the sample solution into the instrument by pipetting; (ii) Placing the synthetic microfluidic paper into the instrument; (iii) Placing the filter paper into the instrument; and (iv) Using the sample strip specific to the densitometer.

Measurement of potassium ion concentration.

(a) Measurement mechanism of the densitometer. (b) Measurement methods: (i) Direct injection of the sample solution into the instrument by pipetting; (ii) Placing the synthetic microfluidic paper into the instrument; (iii) Placing the filter paper into the instrument; and (iv) Using the sample strip specific to the densitometer.

Results and discussion

Precision sampling experiment

Fig 7 shows the coefficient of variation (CV) of sampling volume for each strip, which was calculated as the ratio of the standard deviation to the mean value (n = 5). The actual volume sampled by each paper is shown in S1 Fig. The sprinkling method produced no less than 0.06 in CV for all types of SMP. The CV value for immersion methods Ⅰ and Ⅱ using SMP decreased as pitch ratio increased, while no significant difference was found between the two methods. Among the conditions tested, immersion method I or II using the SMP with (d, p) of (75, 150), (75, 225), and (100, 300) seemed better for sampling liquid when one needs small CV (< 0.05).
Fig 7

Results of precision sampling experiments.

The horizontal axis represents sampling strip type and the vertical axis represents coefficient of variation (CV) [–]. The number of trials n using a same paper strip is 5.

Results of precision sampling experiments.

The horizontal axis represents sampling strip type and the vertical axis represents coefficient of variation (CV) [-]. The number of trials n using a same paper strip is 5.

Sample retention test

Fig 8 shows the falling ratio (R), which represents the ratio of the falling amount (V) to sampling volume (V), as described in Eq (1).
Fig 8

Results of sample retention tests.

The horizontal axis represents strip type, while the vertical axis represents falling ratio [%]. Error bars represent standard deviation. The number of trials n using same paper strip is 5.

Results of sample retention tests.

The horizontal axis represents strip type, while the vertical axis represents falling ratio [%]. Error bars represent standard deviation. The number of trials n using same paper strip is 5. The error bars represent standard deviation. The SMP with a pitch rate of 3.0 had a greater R value than other conditions. It is deduced SMP with a large pitch ratio were susceptible to mechanical disturbance. The SMP with (d, p) of (75, 112.5), (75, 150), (100, 150), and (100, 200) had the similar R value. It is deduced the SMP were less susceptible to mechanical disturbance, resulting in lower variations in sampling volume. As shown in S1 Table, it is certain that there is a significant difference between these parameters ((d, p) of (75, 112.5), (75, 150), (100, 150), and (100, 200)) and the others, when an α-level of 0.05 was used. On the basis on the results of the precision sampling experiments and sample retention test, we decided the use of the SMP with (d, p) of (75, 150) as a sampling strip for the following experiments. Fig 9(A) shows the coefficient of variation (CV) of sampling volume for the SMP and the filter paper, which was calculated as the ratio of the standard deviation to the mean value (N = 10). The actual volume sampled by each paper is shown in S2 Fig. The CV value of the SMP was slightly greater than that of the filter paper. However, we performed F-test on the two normalized data set (ratio of sampled mass over its mean mass) of the SMP and the filter paper by “ftest” function in Excel, and, when an α-level of 0.05 was used, no significant difference was found between them (P = 0.382 > α). Fig 9(B) shows the R. The error bars represent standard deviation. T-test performed on the two groups by “ttest” function in Excel showed no significant difference between the SMP and the filter paper (P = 0.078 > α), when an α-level of 0.05 was used. On the other hand, F-test performed on the two groups showed that there is a significant difference between variation for the two samples (P = 0.0056 < α), when an α-level of 0.05 was used.
Fig 9

Results of precision sampling tests (a), and sample retention tests (b), comparing synthetic microfluidic paper (SMP) and filter paper. The number of trials N using different paper strip every time is 10.

Results of precision sampling tests (a), and sample retention tests (b), comparing synthetic microfluidic paper (SMP) and filter paper. The number of trials N using different paper strip every time is 10. In summary of the comparison between the SMP and the filter paper, they were found to have comparable intra-assay precision in water sampling and retention, but robustness to mechanical disturbances was shown more in the SMP than the filter paper. However, there is a big difference between the SMP and the filter paper: the former was fabricated manually in a laboratory but the latter was done by mass-production process. Actually, the mass variation of the SMP and the filter paper was 0.078 and 0.033 in CV. Automated process or reduction of manual process, if possible, are expected to decrease variation of mass in SMP and consequently variation of sampling volume. From a future perspective, sampling liquid by SMP has challenges in terms of evaluation for long-term or multi-person use. In this work, the usability of SMP for precise sampling has been confirmed in short-term period. It is also important to know long-term capability to perform precise sampling as a reference for a shelf-life. To get longer-term capability, optimization of polymer formulation for the SMP is one of the issues to be considered. Another challenge to be addressed is multi-operator tests. Depending on the application destination, it is necessary to test with a wide range of subjects who have different age and presence or absence of disease.

Measurement of potassium ion concentration

Potassium ion concentration in four types of strips was measured using a commercially available densitometer. Fig 10(A) shows the relation between measured potassium ion concentration and the reference values for each measurement method. Fig 10(B) shows the relative error η, which is described by Eq 2. The value of C represents the measurement result while C is the reference concentration of the sample solution. Standard solution potassium ion concentration [ppm] represents the concentration of the sample solution.
Fig 10

(a) Potassium ion concentration in different sampling papers and (b) Relative error for potassium ion concentration measurements. (a) The horizontal axis represents standard solution potassium ion concentration [ppm] and the vertical axis represents measured potassium ion concentration [ppm]. (b) The horizontal axis represents standard solution potassium ion concentration [ppm], the vertical axis represents relative error [%], and error bars represent standard deviation. For SMP, the number of trials n using a same strip is 30 and for other paper the number of trials N using different paper every time is 30.

(a) Potassium ion concentration in different sampling papers and (b) Relative error for potassium ion concentration measurements. (a) The horizontal axis represents standard solution potassium ion concentration [ppm] and the vertical axis represents measured potassium ion concentration [ppm]. (b) The horizontal axis represents standard solution potassium ion concentration [ppm], the vertical axis represents relative error [%], and error bars represent standard deviation. For SMP, the number of trials n using a same strip is 30 and for other paper the number of trials N using different paper every time is 30. For SMP, the relative error η was 0.38 to 2.2%, with good accuracy. Because the relative error in the control experiment was −1.2 to 0.48%, measurement using SMP had accuracy comparable to the control experiment. For measurement using filter paper, the relative error rates were −17 to −25%. For the sampling sheet, we observed the tendency that the measured value was greater than the control experiments. We speculate that the large relative error of concentration measured using filter paper results from the negative charge of the fiber that is the raw material of the filter paper [27]. The results showed that the SMP was better than the filter paper and the sampling sheet provided by the densitometer manufacturer. This result suggests that the SMP is applicable to precision sampling of filtrate of a patient of implantable artificial kidney, followed by the potassium concentration measurement.

Conclusion

The usefulness of SMPs for the precise collection of liquid samples (such as urine) for ion concentration measurement was investigated. Precise and consistent sampling volumes were achieved with the SMPs being comparable to filter paper when sampling was done by immersion into the liquid. The robustness against mechanical disturbance was high when the SMP with small pitch ratio was used. In the concentration measurement, the strip of SMP was more useful than the filter paper, and the sampling sheet provided by the densitometer manufacturer.

P-value derived from T-test on sampling results using synthetic microfluidic paper.

(TIF) Click here for additional data file.

Mass per unit capacity of water sampled by synthetic microfluidic paper (SMP) in precision sampling experiments.

(TIF) Click here for additional data file.

Mass of water sampled by synthetic microfluidic paper (SMP) and filter paper.

(TIF) Click here for additional data file. 9 Jan 2020 PONE-D-19-32085 Ion concentration measurement using synthetic microfluidic papers; a new technology of precise sampling PLOS ONE Dear Miss Kamiya, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. We would appreciate receiving your revised manuscript by Feb 23 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'. Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Albert J. Fornace Jr, MD Academic Editor PLOS ONE Additional Editor Comments: The authors need to address the concerns of both reviewers. An administrative issue needs to be addressed about how closely this submission is related to a previously published article in the proceedings of the 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), https://doi.org/10.1109/EMBC.2018.8513267. Can you confirm that the present work has been written as an independent article, the related works have been adequately discussed by the authors, and that the separation into more than one article has not compromised the robustness of the statistical analysis (e.g., by removing required adjustments for multiple hypothesis testing). For further information on submission guidelines on related manuscripts, please see http://journals.plos.org/plosone/s/submission-guidelines#loc-related-manuscripts. Journal requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ 3. Please amend either the title on the online submission form (via Edit Submission) or the title in the manuscript so that they are identical. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: No ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know Reviewer #2: No ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors present a very clear and simple comparison paper that nicely addresses ion concentration measurement using synthetic microfluidic papers. To better comprehend that information presented the following questions/comments should be addressed: 1) Please describe conditions of testing in more detail, as temperature, humidity, drying time, measurement time can all effect ion concentration measurements using microfluidic paper. The possible condition limitations are important to a medical setting. 2) Please address the appropriateness of using your sample size (n=5). Reviewer #2: Authors present in this manuscript an interesting microfluidic-based point-of-care device whose paper properties seem to improve sample volume recovery and ion measurement in large samples such as urine. However, additional experiments and precision in methodology would be required to sustain these promising assertions. Major points: - Overall, it is difficult to know how many replicates have been used for each experiment. Authors should clarify this point in experiments section and add n numbers in legend of each figures when applicable. In “precision sampling experiment” in experiments section, authors claimed “six samples (N=5, for each) were collected for each method…” but it is unclear. Power analysis is not provided, but I would suggest authors to perform each condition with at least 10-20 replicates. - Intra- and inter-assay variation could also be an important parameter to assess. I would suggest authors to perform precision sampling and sample retention test experiments on (1) same lot of microfluidic paper and same operator over a short period of time (intra-assay) and (2) different lot over an extended period of time performed by multiple operators (inter-assay). - For the precision sample experiment, in addition to the coefficient of variation, it would be interesting to also show the actual volume collected by each strip/filter paper (by including errors bars and calculating difference between each method). - Line 195, please discuss/provide an explanation on why the synthetic microfluidic paper with (d, p) of (75, 150), (75, 225), and (100, 300) was more suitable for producing consistent sampling volume than others. - Line 206-208, “the values of the synthetic microfluidic papers with a pitch rate of 1.5 and 2.0 and (d, p) of (75, 112.5), (75, 150), (100, 150), and (100, 200) were approximately the same as that of filter paper” and “the synthetic microfluidic paper with a pitch ratio of 3.0 had a greater value”. Is it significant? Authors should calculate significance of difference between strip themselves and filter. - Time between collection and analysis is also a parameter that can widely change (especially when performed directly by patient) and induce important variation. Therefore, authors should specify the period of time between sample collection and measurements they performed. In addition, I would also suggest authors to evaluate potassium concentration at different time after collection in order to assess if microfluidic strip can result in better sample preservation than filter paper. - Authors assessed the interest of their device in a context of “urinalysis” for a patient of implantable artificial kidney. Although it is surprising that authors do not use urine sample being given the easy access, they claimed that they used water to test their device. What is the pH of this solution? pH can impact ion measurement and it would have been a minimum to use a solution with a pH equal to urine (~6). Even more interesting, potassium measurement could have been tested with solution with different pH (4-8). - Line 237, authors claimed “This result suggests that the synthetic microfluidic papers are applicable for precision sampling of urine for the potassium concentration measurement.” This seems overstated since authors did not assess potassium concentration in urine. In addition, sample nature (water spiked with K+?) needs to be specified in Materials and methods section. - Sensitivity could also have been evaluated by measuring the lowest achievable potassium concentration with the different methods. Minor points: - In introduction (line 52), please specify what “unprecise sampling” means (volume variation, stability, concentration variation, etc.?) - In introduction (line 66), please remove “or urine” - In “fabrication process” in Materials and methods section, authors claimed that “The synthetic microfluidic paper … is more robust against collapse than the structures containing vertical straight pillars”. Reference (or data) should be included to support this affirmation. - Please specify how CV has been calculated ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. 6 Oct 2020 Our response to the two reviewers’ comments on the previously submitted manuscript for PLOS ONE the Response to Reviewers file. In the revised manuscripts, these comments are reflected in the main text and the electronic supporting information. Submitted filename: Response to Reviews_PLOSONE_2nd_Kamiya_20200928.docx Click here for additional data file. 29 Oct 2020 Ion concentration measurement using synthetic microfluidic papers PONE-D-19-32085R1 Dear Dr. Kamiya, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Both reviewers and I  recommend acceptance.  Minor corrections are needed for misspelling errors: e.g. Fig 4: Immersion method (immerision) and Suppl Fig 1: Y axis title: capacity (capasity). Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Albert J. Fornace Jr, MD Academic Editor PLOS ONE Additional Editor Comments (optional): Recommend acceptance. Minor corrections are needed for misspelling errors: e.g. Fig 4: Immersion method (immerision) and Suppl Fig 1: Y axis title: capacity (capasity). Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Authors have adequately addressed all comments and have provided extra trials giving the experiments the robustness that was advised. Reviewer #2: All comments have been addressed. Minor corrections are misspelling errors. Ex: Fig 4: Immersion method (immerision) Suppl Fig 1: Y axis title: capacity (capasity) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No 9 Nov 2020 PONE-D-19-32085R1 Ion concentration measurement using synthetic microfluidic papers Dear Dr. Kamiya: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Albert J. Fornace Jr Academic Editor PLOS ONE
  21 in total

1.  Patterned paper as a platform for inexpensive, low-volume, portable bioassays.

Authors:  Andres W Martinez; Scott T Phillips; Manish J Butte; George M Whitesides
Journal:  Angew Chem Int Ed Engl       Date:  2007       Impact factor: 15.336

2.  Simple telemedicine for developing regions: camera phones and paper-based microfluidic devices for real-time, off-site diagnosis.

Authors:  Andres W Martinez; Scott T Phillips; Emanuel Carrilho; Samuel W Thomas; Hayat Sindi; George M Whitesides
Journal:  Anal Chem       Date:  2008-04-11       Impact factor: 6.986

3.  Development of a nanotechnology-based dialysis device.

Authors:  Yoshihiko Kanno; Norihisa Miki
Journal:  Contrib Nephrol       Date:  2012-05-08       Impact factor: 1.580

Review 4.  Paper-based microfluidic point-of-care diagnostic devices.

Authors:  Ali Kemal Yetisen; Muhammad Safwan Akram; Christopher R Lowe
Journal:  Lab Chip       Date:  2013-05-08       Impact factor: 6.799

5.  Tear glucose detection of hyperglycemia.

Authors:  A R Gasset; L E Braverman; M C Fleming; R A Arky; B R Alter
Journal:  Am J Ophthalmol       Date:  1968-03       Impact factor: 5.258

6.  A Precise Sampling Strip with Microstructures.

Authors:  H Kamiya; T Ota; H Yasuga; N Miki
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2018-07

7.  Beyond PDMS: off-stoichiometry thiol-ene (OSTE) based soft lithography for rapid prototyping of microfluidic devices.

Authors:  Carl Fredrik Carlborg; Tommy Haraldsson; Kim Öberg; Michael Malkoch; Wouter van der Wijngaart
Journal:  Lab Chip       Date:  2011-08-01       Impact factor: 6.799

Review 8.  Pregnancy tests: a review.

Authors:  T Chard
Journal:  Hum Reprod       Date:  1992-05       Impact factor: 6.918

Review 9.  Salivary cortisol as a biomarker in stress research.

Authors:  Dirk H Hellhammer; Stefan Wüst; Brigitte M Kudielka
Journal:  Psychoneuroendocrinology       Date:  2008-12-18       Impact factor: 4.905

Review 10.  Point-of-care testing (POCT): Current techniques and future perspectives.

Authors:  Peter B Luppa; Carolin Müller; Alice Schlichtiger; Harald Schlebusch
Journal:  Trends Analyt Chem       Date:  2011-03-21       Impact factor: 12.296