[Comparison of three medical goggle sterilizing approaches].

Qun Wang1, Junjun Mo1, Fang Huang1, Ying Pu1, Bei Lyu1.   

Abstract

OBJECTIVE: To compare three sterilizing methods for reusable medical goggles.
METHODS: A total of 180 medical goggles of the same brand and same model were randomly divided into three groups. In group A the goggles were first soaked with 2000 mg/L chlorine-containing disinfectant and then cleaned manually; goggles in other two groups were sterilized using pre-programmed automatic spray cleaning and disinfection machine, the disinfection program was set to 90 ℃ for 5 min in group B and 70 ℃ for 30 min in group C. The quality of the sterilization was monitored by visual inspection with luminous magnifying glass and residual protein detection assay. User satisfaction on cleanliness of medical goggles, clarity of mirror surface and suitability of elastic bands was investigated with questionnaire survey.
RESULTS: The qualification rates verified by visual inspection were 82.4%, 84.6%and 98.3%in group A, B and C, respectively, the qualification rate in group C was significantly higher than those in group B and group C (all P<0.05). The qualification rates verified by residual protein detection assay were 96.7%, 100.0%and 100.0%in group A, B and C, respectively (P>0.05). A total of 54 questionnaires were submitted for the survey. The satisfaction rates were 100.0%, 90.7%and 94.4% for cleanliness of medical goggles, clarity of mirror surface and suitability of elastic bands, respectively.
CONCLUSIONS: Machinery sterilization set 70 ℃ for 30 min has better cleaning and sterilizing effects for reusable medical goggles.

Entities:  

Keywords:  Cleaning; Coronavirus disease 2019; Medical goggle; Satisfaction degree; Severe acute respiratory syndrome coronavirus 2; Sterilization

Mesh:

Year:  2020        PMID: 33210488      PMCID: PMC8800676          DOI: 10.3785/j.issn.1008-9292.2020.10.10

Source DB:  PubMed          Journal:  Zhejiang Da Xue Xue Bao Yi Xue Ban        ISSN: 1008-9292


  1 in total

1.  2019-nCoV transmission through the ocular surface must not be ignored.

Authors:  Cheng-Wei Lu; Xiu-Fen Liu; Zhi-Fang Jia
Journal:  Lancet       Date:  2020-02-06       Impact factor: 79.321

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.