Transseptal puncture for ablation of atrial fibrillation in a patient with an implanted atrial flow regulator: a case report.

Cornelia Biller1, Karlo Filipovic1, Jakob Lüker1, Daniel Steven1.   

Abstract

BACKGROUND: Atrial flow regulator (AFR) (Occlutech, Helsingborg, Sweden) are self-expanding, circular devices. A flexible waist in the centre connects the two discs and has a centrally located shunt. CASE
SUMMARY: We report a case of an 80-year-old woman undergoing a repeat left atrial ablation for persistent atrial fibrillation with an implanted AFR. The AFR was implanted 1 year prior to the procedure for heart failure with preserved ejection fraction as part of the AFR-PRELIEVE trial. A single, fluoroscopy-guided, transseptal puncture was performed infero-posterior to the device, allowing the positioning of the mapping (LASSO® 20 mm, Biosense Webster, Irvine, CA, USA) and ablation (Thermocool Smarttouch SF, CARTO®, Biosense Webster, Irvine, CA, USA) catheter in the left atrium. Three-dimensional mapping (CARTO®, Biosense Webster, Irvine, CA, USA) and left atrial ablation were successfully performed. After the procedure, fluoroscopy and transthoracic echocardiography showed an unchanged device position. DISCUSSION: To our knowledge, this is the first case report of a transseptal puncture in a patient with an implanted AFR. Transseptal puncture in patients with an implanted AFR seems to be safe and feasible. With device diameters of 21-23 mm and based on previous studies on similar devices, transseptal puncture should be possible in most patients, as opposed to puncture through the device.
© The Author(s) 2020. Published by Oxford University Press on behalf of the European Society of Cardiology.

Entities:  

Keywords:  Ablation; Atrial fibrillation; Atrial flow regulator; Case report; Transseptal puncture

Year:  2020        PMID: 33204955      PMCID: PMC7649456          DOI: 10.1093/ehjcr/ytaa242

Source DB:  PubMed          Journal:  Eur Heart J Case Rep        ISSN: 2514-2119


Transseptal puncture and left atrial procedure in patients with implanted atrial flow regulator (AFR) seems to be safe and feasible. The transseptal puncture can be performed fluoroscopy-guided only. The diameter of the AFR device is 21–23 mm, which should allow for enough free septal area for a transseptal puncture in most patients.

Introduction

Atrial flow regulator (AFR) (Occlutech, Helsingborg, Sweden) are self-expanding, circular devices implanted in the interatrial septum as a study treatment for heart failure with preserved ejection fraction. A flexible waist in the centre connects the two discs and has a centrally located shunt. In this case report, we describe a patient undergoing a left atrial ablation procedure for persistent atrial fibrillation with an implanted AFR. AFR in unchanged device position Patient in stable sinus rhythm No other post-procedural complications

Case presentation

We report a case of an 80-year-old woman undergoing a repeat pulmonary vein isolation for persistent atrial fibrillation with an implanted AFR. The AFR was implanted 1 year prior to the procedure for heart failure with preserved ejection fraction as part of the AFR-PRELIEVE trial. The AFR-PRELIEVE trial sought to assess the effects of transcatheter AFR implantation on symptoms and surrogate parameters of heart failure in patients with reduced or preserved left ventricular ejection fraction. Relevant pre-existing conditions were ischaemic cardiomyopathy, persistent atrial fibrillation, and history of transient ischaemic attack. The first pulmonary vein isolation took place two and a half years prior to the ablation procedure and resulted in sinus rhythms maintenance for 24 months and significant symptom improvement. Relevant medication consisted of a betablocker and oral anticoagulation with Apixaban. The 12-lead electrocardiography at hospital admission showed atrial fibrillation with normal ventricular response at a heart rate of 85 b.p.m. A preprocedural transoesophageal echocardiography (TOE) showed the AFR in an anterosuperior position in the fossa ovalis with sufficient free septal area posterior and inferior to the device (Figure ), as well as a normal left ventricular ejection fraction and absence of left atrial thrombi. Oral anticoagulation was discontinued the day prior to the procedure. The ablation procedure was performed under sedation using propofol, midazolam, and fentanyl. After obtaining triple femoral venous access (TSX™ Fixed Curve Transseptal Sheath, Boston Scientific, Marlborough, MA, USA) through the right femoral vein and positioning of a decapolar coronary sinus catheter, a fluoroscopy-guided single transseptal puncture (TSX™ Transseptal Needle, Boston Scientific) infero-posterior to the AFR was performed (Figure ). Immediately after assuring successful transseptal puncture, a weight-adjusted heparin bolus was administered, with successive boluses every 30 min based on a targeted activated clotting time of >300 s. According to our clinical routine, the ablation catheter (Thermocool Smarttouch SF, CARTO®, Biosense Webster, Irvine, CA, USA) was guided along the wire into the left atrium, omitting the need for a second transseptal puncture. Subsequently, a three-dimensional map (LASSO® 20 mm, Biosense Webster, Irvine, CA, USA) of the left atrium was created (CARTO®, Biosense Webster, Irvine, CA, USA). All pulmonary veins were shown to be isolated, therefore, ablation of complex fractionated atrial electrograms was performed. Since extended catheter ablation only transiently leads to conversion in a left atrial tachycardia, external electrical cardioversion was performed at the end of the procedure. Oral anticoagulation with Apixaban was continued on the same evening. Pre-procedural transoesophageal echocardiography [midoesophageal 120° (A) and 60° (B) views] showing adequate septal area for transseptal puncture inferior and posterior to the atrial flow regulator (*), respectively. Fluoroscopy showing transseptal puncture infero-posterior to the atrial flow regulator device in antero-posterior (A) and left anterior oblique (B) views as well as mapping and ablation catheters positioned in the left atrium (C). Fluoroscopy at the end of the procedure as well as transthoracic echocardiography performed on the next day showed an unchanged device position (Figure ). Significant pericardial effusion was excluded and no other post-procedural complications were observed during the hospital stay. Post-procedural transthoracic echocardiography (four-chamber view) showing the atrial flow regulator in loco typico.

Discussion

Transseptal puncture is routinely performed in left atrial procedures. To our knowledge, so far no data exist on transseptal puncture in patients with an implanted AFR. The AFR is a self-expanding, double-disc, circular device made of nitinol wire mesh. A flexible waist in the centre connects the two discs and has a centrally located shunt. The devices are produced in two sizes with transverse diameters of 21 and 23 mm, and shunt sizes of 8 mm and 10 mm, respectively. Our patient had a device diameter of 21 mm and a shunt size of 8 mm, an opening judged too small for direct positioning of both catheters through the shunt, while still allowing adequate manoeuvrability. The transseptal rather than ‘transdevice’ route into the left atrium was also felt to be safer concerning device dislodgement. Previously, other case reports and studies showed the safety of transseptal punctures in challenging patients. Gloeckler et al. reported a successful transseptal puncture through an Amplatzer Atrial Septal Occluder for left atrial appendage closure using fluoroscopy guidance only, with subsequent closure of the access hole with a small atrial septal occluder (ASO). Li et al. presented a case series of nine patients with implanted ASOs undergoing left atrial ablation for atrial fibrillation. Alike in the case presented, in six patients, left atrial access was obtained performing a double puncture through the septum infero-posterior to the device. In the remaining three patients a single puncture through the device with subsequent balloon dilatation was performed. Most of the punctures were fluoroscopy-guided only. Sang et al. presented a case series of 16 patients with implanted ASOs, undergoing ablation for atrial fibrillation. Preprocedural imaging using three-dimensional computed tomography (CT) was performed in all patients and used to determine possible entry points. A single puncture was performed through the septum in 11 patients, thereof infero-posterior to the device (as in our case) in eight patients and infero-anterior to the device in three patients. A single puncture was performed through the device in the rest of the patients, with subsequent balloon dilatation. Transseptal punctures were successful in all patients with no puncture-related complications. We showed that transseptal puncture in a patient with AFR is safe and feasible using fluoroscopy guidance only. Li et al. reported that transseptal access through the ASO may be necessary in patients with a device diameter >26 mm, whereas Sang et al. reported successful transseptal puncture in ASOs as large as 32 mm in favourable cases. The diameter of the AFR device is 21–23 mm, which should allow for enough free septal area for a transseptal puncture in most patients. Our chosen entry point was infero-posterior to the device, which was also the most frequently used entry point in the ASO studies. We used preprocedural TOE to help planning the transseptal approach. In our centre, we do not routinely use TOE imaging for transseptal puncture as fluoroscopy only guidance allows successful puncture in most patient with low complication rates, although TOE is available as backup in the setting of difficult anatomy. In this case, the transseptal puncture was easily performed using fluoroscopy only, so no TOE guidance was required.

Conclusion

Transseptal puncture in patients with an implanted AFR seems to be safe and feasible and can be performed fluoroscopy-guided only. Preprocedural imaging (CT or TOE) can help to plan the approach. The diameter of the AFR device is 21–23 mm, which should allow for enough free septal area for a transseptal puncture in most patients.

Lead author biography

Cornelia Biller is a clinical fellow at the Department of Electrophysiology at the University Hospital of Cologne, Germany. Her current research focus is on atrial fibrillation and device therapy.

Supplementary material

Supplementary material is available at European Heart Journal - Case Reports online. Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data. Consent: The author/s confirm that written consent for submission and publication of this case report including image(s) and associated text has been obtained from the patient in line with COPE guidance. Conflict of interest: none declared. Click here for additional data file.
March 2017Initial diagnosis of persistent atrial fibrillation - electrical cardioversion
November 2017Pulmonary vein isolation
November 2018Heart failure with preserved ejection fraction
November 2018Implantation atrial flow regulator (AFR) as part of AFR-PRELIEVE trial
November 2019Recurrence of atrial fibrillation
March 2020Repeat left atrial ablation - transseptal puncture with implanted AFRResult of the procedure:

AFR in unchanged device position

Patient in stable sinus rhythm

No other post-procedural complications

  4 in total

1.  The AFR-PRELIEVE trial: a prospective, non-randomised, pilot study to assess the Atrial Flow Regulator (AFR) in heart failure patients with either preserved or reduced ejection fraction.

Authors:  Christina Paitazoglou; Ramazan Özdemir; Roman Pfister; Martin W Bergmann; Jozef Bartunek; Teoman Kilic; Alexander Lauten; Alexander Schmeisser; Mehdi Zoghi; Stefan Anker; Horst Sievert; Felix Mahfoud
Journal:  EuroIntervention       Date:  2019-08-29       Impact factor: 6.534

2.  Safety and feasibility of transseptal puncture for atrial fibrillation ablation in patients with atrial septal defect closure devices.

Authors:  Xuping Li; Erik Wissner; Masashi Kamioka; Hisaki Makimoto; Peter Rausch; Andreas Metzner; Shibu Mathew; Andreas Rillig; Roland Richard Tilz; Alexander Fürnkranz; Qi Chen; Qingyong Zhang; Qiming Liu; Shenghua Zhou; Karl-Heinz Kuck; Feifan Ouyang
Journal:  Heart Rhythm       Date:  2013-11-14       Impact factor: 6.343

3.  Transseptal Puncture Through Amplatzer Atrial Septal Occluder for Left Atrial Appendage Closure.

Authors:  Steffen Gloekler; Samera Shakir; Bernhard Meier
Journal:  JACC Cardiovasc Interv       Date:  2017-11-13       Impact factor: 11.195

4.  Transseptal puncture and catheter ablation of atrial fibrillation in patients with atrial septal occluder: initial experience of a single centre.

Authors:  Cai-Hua Sang; Jian-Zeng Dong; De-Yong Long; Rong-Hui Yu; Rong Bai; Mohamed Salim; Ri-Bo Tang; Man Ning; Chen-Xi Jiang; Nian Liu; Song-Nan Li; Song-Nan Wen; Jia-Hui Wu; Ke Chen; Ying-Wei Chen; Chang-Sheng Ma
Journal:  Europace       Date:  2018-09-01       Impact factor: 5.214

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.