Literature DB >> 33204183

A Novel Radiofrequency Device to Monitor Changes in Pulmonary Fluid in Dialysis Patients.

Jeffrey J Connaire1, Matthew L Sundermann2, Ramu Perumal2, Charles A Herzog3.   

Abstract

BACKGROUND AND OBJECTIVES: Fluid monitoring is an important management strategy in patients with chronic kidney disease (CKD) and heart failure (HF). The µCor™ Heart Failure and Arrhythmia Management System uses a radiofrequency-based thoracic fluid index (TFI) to track pulmonary edema. During hemodialysis, the acute removal of fluid through ultrafiltration offers a model for measuring a patient's fluid status. The objective of the study was to assess the relationship between the device measured TFI and ultrafiltration volume (UFV). DESIGN SETTING PARTICIPANTS AND MEASUREMENTS: Patients undergoing chronic dialysis with and without heart failure were enrolled in the study. The relationship between TFI and UFV in each individual subject was assessed by calculating the Pearson correlation coefficient (r). The average correlation across all subjects was calculated through the use of the Fisher's z transform. Responder analysis was performed to assess the magnitude of change in TFI before and after dialysis.
RESULTS: Twenty subjects were enrolled in the trial. The mean volume of fluid removal was 3.63 L (SD 0.88 L). The mean correlation based on Fisher's transform was 0.95 CI (0.92-0.99). Responder analysis showed that the mean reduction of TFI after dialysis was 5.5% ± 3.8.
CONCLUSION: The µCor system provides radiofrequency-based measurements of thoracic fluid which correlate well with total body fluid removal in a real-world setting. Fluid management based on the radar-derived TFI may provide benefits to dialysis patients and serves as a potential model for pulmonary edema common to the clinical course of heart failure.
© 2020 Connaire et al.

Entities:  

Keywords:  dialysis; heart failure; total body fluid monitoring; ultrafiltration; wearables

Year:  2020        PMID: 33204183      PMCID: PMC7667146          DOI: 10.2147/MDER.S277159

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


  11 in total

Review 1.  Current state of bioimpedance technologies in dialysis.

Authors:  Peter Kotanko; Nathan W Levin; Fansan Zhu
Journal:  Nephrol Dial Transplant       Date:  2008-01-08       Impact factor: 5.992

2.  Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans.

Authors:  D Bracco; D Thiébaud; R L Chioléro; M Landry; P Burckhardt; Y Schutz
Journal:  J Appl Physiol (1985)       Date:  1996-12

Review 3.  Lung Ultrasonography and Cardiac Surgery: A Narrative Review.

Authors:  Sergey M Efremov; Vsevolod V Kuzkov; Evgenia V Fot; Mikhail Y Kirov; Dmitry N Ponomarev; Roman E Lakhin; Evgenii A Kokarev
Journal:  J Cardiothorac Vasc Anesth       Date:  2020-01-23       Impact factor: 2.628

4.  Detection of pulmonary congestion by chest ultrasound in dialysis patients.

Authors:  Francesca Mallamaci; Francesco A Benedetto; Rocco Tripepi; Stefania Rastelli; Pietro Castellino; Giovanni Tripepi; Eugenio Picano; Carmine Zoccali
Journal:  JACC Cardiovasc Imaging       Date:  2010-06

Review 5.  Fluid management for the prevention and attenuation of acute kidney injury.

Authors:  John R Prowle; Christopher J Kirwan; Rinaldo Bellomo
Journal:  Nat Rev Nephrol       Date:  2013-11-12       Impact factor: 28.314

Review 6.  Proposal for a functional classification system of heart failure in patients with end-stage renal disease: proceedings of the acute dialysis quality initiative (ADQI) XI workgroup.

Authors:  Lakhmir S Chawla; Charles A Herzog; Maria Rosa Costanzo; James Tumlin; John A Kellum; Peter A McCullough; Claudio Ronco
Journal:  J Am Coll Cardiol       Date:  2014-02-12       Impact factor: 24.094

7.  A positive fluid balance is associated with a worse outcome in patients with acute renal failure.

Authors:  Didier Payen; Anne Cornélie de Pont; Yasser Sakr; Claudia Spies; Konrad Reinhart; Jean Louis Vincent
Journal:  Crit Care       Date:  2008-06-04       Impact factor: 9.097

8.  Feasibility of Using Wideband Microwave System for Non-Invasive Detection and Monitoring of Pulmonary Oedema.

Authors:  S Ahdi Rezaeieh; A Zamani; K S Bialkowski; A Mahmoud; A M Abbosh
Journal:  Sci Rep       Date:  2015-09-14       Impact factor: 4.379

9.  The effect of diuresis on extravascular lung water and pulmonary function in acute decompensated heart failure.

Authors:  Steven C Chase; Caitlin C Fermoyle; Courtney M Wheatley; Jacob J Schaefer; Lyle J Olson; Bruce D Johnson
Journal:  ESC Heart Fail       Date:  2018-01-18

10.  Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study.

Authors:  Natasa Reljin; Hugo F Posada-Quintero; Caitlin Eaton-Robb; Sophia Binici; Emily Ensom; Eric Ding; Anna Hayes; Jarno Riistama; Chad Darling; David McManus; Ki H Chon
Journal:  JMIR Med Inform       Date:  2020-08-27
View more