Yuxiang Wu1,2, Jinzhi Yu2, Feng Li2, Jianlin Li1, Zhiqiang Shen2.
Abstract
In this study, a rapid and sensitive immunoassay method has been established based on calibration curve implanted enzyme-linked immunosorbent assay (C-ELISA) for the simultaneously quantitative determination of aflatoxin B1, deoxynivalenol and zearalenone in cereal samples, soybean and peanut. The C-ELISA avoids using the standard substances during the detection. The principle of the C-ELISA is to implant the optimized standard curve data into the matched analysis software which can make data processing more convenient and faster. The implanted calibration curve software was programmed with C plus plus. In the new immunoassay system for aflatoxin B1, deoxynivalenol and zearalenone, their linear detection ranges were from 0.03~0.81, 1.00~27.00 and 5.00~135.00 ng/g, respectively. Recovery rates from spiked samples ranged from 85% to 110% with the intra-assay coefficients of variation under 5%. Compared with HPLC method, the new method showed consistence in all the observed contents of the three mycotoxins in real samples. The new method can rapidly and reliably high throughput simultaneously screen for multiplex mycotoxins.Entities:
Keywords: C plus plus programming language; analysis software; calibration curve implanted ELISA; high throughput detection; multiplex mycotoxins
Mesh:
Substances:
Year: 2020 PMID: 33202953 PMCID: PMC7696722 DOI: 10.3390/toxins12110718
Source DB: PubMed Journal: Toxins (Basel) ISSN: 2072-6651 Impact factor: 4.546
Figure 1The optimal mycotoxin calibration curves for implanting in the analysis software.
The optimal mycotoxin calibration curves for the three mycotoxins in the detection assay.
| Analytes | Standard Curve | IC50 | LOD/(ng/g) | Working Range/ |
|---|---|---|---|---|
| AFB1 | yAFB1 = −45.851x + 96.820, R2 = 0.991 | 0.093 | 0.03 | 0.03~0.81 |
| ZEN | y ZEN = −47.158x + 25.200, R2 = 0.987 | 3.38 | 1.00 | 1.00~27.00 |
| DON | yDON = −43.846x − 6.077, R2 = 0.962 | 17.70 | 5.00 | 5.00~135.00 |
The results measured by the C-ELISA in spiked cereal samples.
| Analytes | Spiked Concentration | CV (%) | 0.1 mol/L Tris-HCL | 0.1 mol/L PBS | 0.1 mol/L PB | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Recovery (%) | False Positive Rate (%) | False Negative Rate (%) | Recovery (%) | False Positive Rate (%) | False Negative Rate (%) | Recovery (%) | False Positive Rate (%) | False Negative Rate (%) | |||
|
| 100 | 0.9~1.1 | 65~95 | 0 | 11 | 85~115 | 1 | 0 | 83~116 | 6 | 0 |
| 300 | 68~97 | 85~114 | 85~115 | ||||||||
| 600 | 65~99 | 81~119 | 81~119 | ||||||||
|
| 200 | 1.3~1.7 | 68~96 | 0 | 15 | 85~116 | 2 | 0 | 81~115 | 9 | 0 |
| 1000 | 67~99 | 84~115 | 82~116 | ||||||||
| 2000 | 65~99 | 81~119 | 85~119 | ||||||||
|
| 1 | 5.1~5.2 | 63~93 | 0 | 18 | 85~119 | 1 | 0 | 85~115 | 7 | 0 |
| 5 | 68~99 | 81~115 | 85~119 | ||||||||
| 20 | 66~100 | 85~115 | 81~114 | ||||||||
The specificity evaluation.
| Analytes | IC50 | S | Analytes | IC50 | S | Analytes | IC50 | S |
|---|---|---|---|---|---|---|---|---|
| AFB1 | 0.093 | 100 | ZEN | 3.38 | 100 | DON | 18.89 | 100 |
| AFB2 | 0.76 | 12.2 | ZEA | 27.7 | 12.2 | 3A-DON | 20.67 | 91.4 |
| AFG1 | 0.87 | 10.7 | ZAA | 42.3 | 7.9 | 15-ADON | 23.65 | 79.9 |
| AFG2 | 0.98 | 9.5 | DON | >5000 | <0.1 | AFB1 | >10000 | <0.1 |
| AFM1 | 1.75 | 5.3 | T2 | >5000 | <0.1 | AFM1 | >10000 | <0.1 |
| AFM2 | 1.97 | 4.7 | OTA | >5000 | <0.1 | AFG1 | >10000 | <0.1 |
| ZEN | >1000 | <0.01 | AFB1 | >5000 | <0.1 | ZEN | >10000 | <0.1 |
| DON | >1000 | <0.01 | AFM1 | >5000 | <0.1 | T2 | >10000 | <0.1 |
| OTA | >1000 | <0.01 | AFG1 | >5000 | <0.1 | OTA | >10000 | <0.1 |
Comparison between C-ELISA and HPLC.
| Sample | C-ELISA (μg/kg) | HPLC(μg/kg) | ||||
|---|---|---|---|---|---|---|
| ZEN | DON | AFB1 | ZEN | DON | AFB1 | |
| S2 | ND | ND | 5.6 ± 0.07 | ND | ND | 6.2 ± 0.02 |
| S4 | 423.2 ± 0.08 | ND | ND | 412.2 ± 0.19 | ND | ND |
| S5 | ND | 956.2 ± 0.11 | ND | ND | 935.3 ± 0.25 | ND |
| S8 | 263.2 ± 0.12 | ND | ND | 265.2 ± 0.13 | ND | ND |
| S15 | ND | ND | 4.6 ± 0.03 | ND | ND | 4.2 ± 0.06 |
| S23 | 412.3 ± 0.16 | ND | ND | 423.2 ± 0.16 | ND | ND |
| S29 | ND | ND | 3.2 ± 0.08 | ND | ND | 3.6 ± 0.07 |
| S1, S3, S6, S7, S9~S14, S16~S22 S24~S28, S30 | ND | ND | ND | ND | ND | ND |
Each was determined with 3 repeats. ND not detectable.
Figure 2The interface of the analysis software implanted with calibration curves.