Literature DB >> 33197634

Comparing Occlusive Balloon Performance Using 3-Dimensional Printed Models of Intracranial Aneurysmal Defects.

Joseph Knox1, Ryan Sincic2, David B McCoy2, Colby Sostarich2, Daniel L Cooke2.   

Abstract

OBJECTIVE: Balloon remodeling microcatheters are essential in assisting endovascular coiling of brain aneurysms, but the performance and pressure requirements of different balloon types when used in aneurysmal defects are currently unknown.
METHODS: We used Tinkercad (Autodesk, Montreal, Quebec) to create model vessels with aneurysmal defects and 3-dimensionally printed these designs with polylactic acid using the Ultimaker2 (Ultimaker, Geldermalsen, Netherlands). We constructed a pressurized box capable of reaching physiologic pressures that housed our vessels and then tested compliant remodeling balloons under fluoroscopy from 3 manufacturers: Hyperglide (Medtronic, Minneapolis, Minnesota, USA), Transform (Stryker Neurovascular, Fremont, California, USA), and Scepter C (Microvention, Tustin, California, USA). Each balloon was inflated to a nominal and supranominal point at each defect, and at each inflation the maximum diameter of the balloon and internal balloon pressure were recorded. The Phillips Intellivue (Phillips, Amsterdam, The Netherlands) monitor was adapted for internal balloon pressure monitoring. A multivariate linear regression analysis was performed to model balloon compliance (balloon diameter divided by pressure).
RESULTS: Multivariate regression modeling demonstrated the Scepter balloon had significantly greater compliance compared with both the Hyperglide and Transform balloons (P < 0.001). In addition, we found that Scepter balloons had higher compliance in larger defects compared with the other types of balloons and performance differences based on vessel size.
CONCLUSIONS: Scepter balloons require less pressure compared with their counterparts to adequately deform through model defects, specifically in larger aneurysm necks in smaller vessel diameters. This result could inform operators of optimal balloon type and size when trying to minimize balloon pressure in fragile brain aneurysms.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  3D-printing; Brain aneurysm; Microcatheters; Occlusion balloons

Mesh:

Year:  2020        PMID: 33197634     DOI: 10.1016/j.wneu.2020.11.030

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  1 in total

1.  Evaluation of the differences in pressure applied to the vessel wall by different types of balloon remodeling microcatheters in an experimental model.

Authors:  Mizuka Ikezawa; Takashi Izumi; Masahiro Nishihori; Tetsuya Tsukada; Yosuke Tamari; Yoshio Araki; Kinya Yokoyama; Kenji Uda; Shunsaku Goto; Asuka E Kropp; Takafumi Otsuka; Naoki Kato; Mizuki Nakano; Ryuta Saito
Journal:  Interv Neuroradiol       Date:  2021-09-13       Impact factor: 1.764

  1 in total