| Literature DB >> 33191950 |
Christopher J Rowlands1,2, Oliver T Bruns3, Daniel Franke4, Dai Fukamura5, Rakesh K Jain5,6, Moungi G Bawendi4, Peter T C So7,8.
Abstract
The first ever demonstration of temporal focusing with short wave infrared (SWIR) excitation and emission is demonstrated, achieving a penetration depth of 500 µm in brain tissue. This is substantially deeper than the highest previously-reported values for temporal focusing imaging in brain tissue, and demonstrates the value of these optimized wavelengths for neurobiological applications.Entities:
Keywords: fluorescence microscopy; multiphoton microscopy; neurophotonics; quantum dots; temporal focusing
Year: 2019 PMID: 33191950 PMCID: PMC7655118 DOI: 10.1088/1361-6463/ab16b4
Source DB: PubMed Journal: J Phys D Appl Phys ISSN: 0022-3727 Impact factor: 3.207
Figure 1.Temporal focusing illustration. An ultrafast pulse strikes a grating and is dispersed; as a consequence, the pulse is spread out in time. The grating surface is imaged onto the sample, thus recombining the pulse components and restoring the optical pulse. The result is that efficient multiphoton excitation can only occur at the focal plane of the objective, and not elsewhere.
Figure 2.Optical layout. Light from the regenerative amplifier pumps the OPA, which changes the wavelength to 1300 nm. This 1300 nm beam passes through a shutter and beam expander, before striking a grating where it is dispersed. The −1 order from the grating is demagnified onto an intermediate focal plane (marked with a dotted line) before being imaged onto the sample using a tube lens and microscope objective. Fluorescent emission from the sample then passes through the dichroic mirror and optical filters, forming an axially-resolved image on the camera.
Figure 3.Transmission spectrum for the microscope tube lens. Note that the absolute value for transmission cannot be relied upon; measurements were taken using a commercial spectrometer and the fact that the test object is a lens means the light falling on the detector is a function both of the transmission and the focusing properties of the lens. Relative values between wavelengths are expected to be preserved however.
Figure 4.Absorbance and photoluminescence spectra of the QDs.
Figure 5.Example frames illustrating the penetration depth of SWIR temporal focusing microscopy. The sample consists of 1023 nm QDs circulating in the brain vasculature of a mouse. The first appearance of a fluorescent feature occurs between 10 µm and 20 µm, and emission is still detectable at 520 µm. The non-uniform background appeared similar in other images, so is not attributed to background fluorescence; it is more likely due to inhomogeneities in the InGaAs sensor noise characteristics (the sensor has a mean read noise of 70e−), or possibly excitation light passing through the two OD2 filters.