Temperature Assessment of a Novel Pulsed Thulium Solid-State Laser Compared with a Holmium:Yttrium-Aluminum-Garnet Laser.

Ralf Petzold1, Rodrigo Suarez-Ibarrola1, Arkadiusz Miernik1.   

Abstract

Purpose: To compare a novel Thulium laser device with the commonly used Holmium:Yttrium-Aluminum-Garnet (Ho:YAG) laser in terms of the in vitro temperatures generated.
Methods: Our study investigated and compared an evaluation model of a solid-state Thulium laser with a Medilas H Solvo 35 Holmium laser device, both by Dornier (Dornier MedTech Laser GmbH, Wessling, Germany). Our in vitro model consisted of a 20 mL test tube placed in a 37°C water bath. Constant irrigation was set at 50 mL/minute with a Reglo Z Digital pump (Cole Parmer, Chicago, IL). Four hundred micrometers of Dornier laser fibers were used. The temperature was measured with a type K thermocouple and a real-time data logger from Pico (PICO Technology, Cambridgeshire, United Kingdom). Power settings between 2 and 30 W were investigated. Each measurement lasted 120 seconds and was repeated five times. The data were evaluated by MATLAB® (The Mathworks, Inc., Natick, MA).
Results: The resulting temperatures were directly proportional to the power supplied. When comparing Holmium with Thulium, we observed maximum deviations of ≤0.82 K in temperatures at 120 seconds. The highest investigated laser power of 30 W yielded maximum temperatures differing by 6.7 K from the initial value. Out of the five comparisons, Thulium showed marginally yet significantly lower end temperatures in four cases and slightly lower cumulative equivalent minutes at 43°C (CEM43) values in three cases.
Conclusion: The Thulium laser resembles the Holmium device in the temperatures generated during in vitro application. An increase in laser power, thus, leads to equivalent increases in temperature that are largely independent of frequency, pulse duration, and single pulse energy. Pulsed Thulium:Yttrium-Aluminum-Garnet (Tm:YAG), Ho:YAG, and Thulium fiber laser seem to share a similar risk profile for patients in terms of temperature development. Intrarenal power outputs exceeding 10 W during clinical application should be compensated by ensuring sufficient irrigation.

Entities:  

Keywords:  Ho:YAG; in vitro study; novel pulsed Tm:YAG laser; temperature development

Mesh:

Substances:

Year:  2020        PMID: 33191783     DOI: 10.1089/end.2020.0803

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  4 in total

Review 1.  A Practical Guide for Intra-Renal Temperature and Pressure Management during Rirs: What Is the Evidence Telling Us.

Authors:  Felipe Pauchard; Eugenio Ventimiglia; Mariela Corrales; Olivier Traxer
Journal:  J Clin Med       Date:  2022-06-15       Impact factor: 4.964

2.  Temperature assessment study of ex vivo holmium laser enucleation of the prostate model.

Authors:  Mehmet Yilmaz; Cäcilia Elisabeth Maria Heuring; Franz Friedrich Dressler; Rodrigo Suarez-Ibarrola; Christian Gratzke; Arkadiusz Miernik; Simon Hein
Journal:  World J Urol       Date:  2022-05-25       Impact factor: 3.661

3.  Can We Identify Patients in Danger of Complications in Retrograde Intrarenal Surgery?-A Retrospective Risk Factors Analysis.

Authors:  Jakub Marek Ratajczak; Taras Hladun; Bartosz Krenz; Krzysztof Bromber; Maciej Salagierski; Michał Marczak
Journal:  Int J Environ Res Public Health       Date:  2022-01-20       Impact factor: 3.390

Review 4.  Moses and Moses 2.0 for Laser Lithotripsy: Expectations vs. Reality.

Authors:  Mariela Corrales; Alba Sierra; Olivier Traxer
Journal:  J Clin Med       Date:  2022-08-18       Impact factor: 4.964