OBJECTIVES: To test the hypothesis that an accelerated, T1-weighted 3D CAIPIRINHA SPACE sequence with isotropic voxel size offers a similar performance to conventional T1-weighted 2D TSE (turbo spin echo) for the evaluation of bone tumor extent and characteristics.
METHODS: Thirty-four patients who underwent 3-T MRI with 3DT1 (CAIPIRINHA SPACE TSE) and 2DT1 (TSE) were included. Sequence acquisition time was reported. Two radiologists independently evaluated each technique for tumor location, size/length, tumor-to-joint distance, signal intensity, margin/extraosseous extension, and signal-to-noise (SNR) and contrast-to-noise (CNR) ratios.
RESULTS: Tumors were located in long (20/36, 55.5%) and pelvic (16/36, 44.4%) bones. 3DT1 sequence required an average acquisition time of 235 s (± 42 s, range 156-372), while two plane 2DT1 sequences combined (coronal and axial) had an average acquisition time of 381 s (± 73 s, range 312-523). There was no difference in the measurements of tumor length and tumor-to-joint distance (p = 0.95) between 3DT1 and 2DT1 images. Tumors were hypointense (17/36, 47.2% vs 17/36, 47.2%), isointense (12/36, 33.3% vs 12/36, 33.3%), or hyperintense (7/36, 19.4% vs 7/36, 19.4%) on 3DT1 vs 2DT1, respectively. Assessment of tumor margins and extraosseous extension was similar, and there was no difference in tumor SNR or CNR (p > 0.05).
CONCLUSIONS: An accelerated 3D CAIPIRINHA SPACE T1 sequence provides comparable assessments of intramedullary bone tumor extent and similar tumor characteristics to conventional 2DT1 MRI. For the assessment of bone tumors, the isotropic volume acquisition and multiplanar reformation capability of the 3DT1 datasets can obviate the need for 2DT1 acquisitions in multiple planes. KEY POINTS: • 3DT1 offers an equivalent performance to 2DT1 for the assessment of bone tumor characteristics, with faster and higher resolution capability, obviating the need for acquiring 2DT1 in multiple planes. • There was no difference in the measurements of tumor length and tumor-to-joint distance obtained on 3DT1 and 2DT1 images. • There was no difference in signal-to-noise ratio (SNR) or contrast-to-noise ratio (CNR) measures between 3DT1 and 2DT1.

Entities:  

Keywords:  Bone neoplasms; Imaging; Magnetic resonance imaging; Three-dimensional

Year:  2020        PMID: 33179165     DOI: 10.1007/s00330-020-07394-4

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  2 in total

1.  Accuracy of Various MRI Sequences in Determining the Tumour Margin in Musculoskeletal Tumours.

Authors:  Tharani Putta; Sridhar Gibikote; Vrisha Madhuri; Noel Walter
Journal:  Pol J Radiol       Date:  2016-11-16

2.  Chemical shift imaging: preliminary experience as an alternative sequence for defining the extent of a bone tumor.

Authors:  Filippo Del Grande; Ney Tatizawa-Shiga; Sahar Jalali Farahani; Majid Chalian; Laura Marie Fayad
Journal:  Quant Imaging Med Surg       Date:  2014-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.