| Literature DB >> 33174973 |
Saulo Maia d'Avila Melo1,2, Marília Ferraz de Oliveira Macedo3, Juliana Silva Santana Pereira2.
Abstract
Entities:
Mesh:
Substances:
Year: 2020 PMID: 33174973 PMCID: PMC7889320 DOI: 10.36416/1806-3756/e20200251
Source DB: PubMed Journal: J Bras Pneumol ISSN: 1806-3713 Impact factor: 2.624
Measurement of SpO2 at three different time points with four different pulse oximeters.a
| Oximeter | Time point | p* | ||
|---|---|---|---|---|
| 30 s | 60 s | 120 s | ||
| ChoiceMMed | 96.38 ± 1.84 (96.06-96.69) | 96.38 ± 1.79 (96.07-96.68) | 96.39 ± 1.88 (96.07-96.71) | 0.988 |
| GO2 Achieve | 95.75 ± 2.46 (95.33-96.17) | 95.78 ± 2.11 (95.42-96.14) | 95.89 ± 2.22 (95.50-96.27) | 0.455 |
| Rossmax SB100 | 96.06 ± 2.20 (95.68-96.44) | 96.07 ± 2.32 (95.67-96.46) | 96.29 ± 2.22 (95.91-96.67) | 0.063 |
| Finger Type & Oximeter | 94.97 ± 2.62 (94.52-95.42) | 95.14 ± 2.65 (94.69-95.60) | 95.22 ± 2.47 (94.79-95.64) | 0.122 |
Values expressed as mean ± SD (95% CI). *ANOVA and Bonferroni test.