| Literature DB >> 33154626 |
Kjell Gunnar Gundersen1, Richard Potvin2.
Abstract
PURPOSE: To compare uncorrected and best-corrected visual acuity, low contrast acuity, residual refraction and ocular biometry after low cylinder power toric intraocular lens (IOL) or non-toric IOL implantation. PATIENTS AND METHODS: This was a non-interventional comparative study of visual outcomes after uncomplicated cataract or refractive lens exchange surgery with either a low cylinder (Low_Cyl) or non-toric (Non_Toric) IOL of similar design implanted (AcrySof® T2 IQ Toric IOL and AcrySof® IQ IOL). Subjects in both groups had to have been eligible for the low cylinder IOL based on biometry. They had to have uncorrected distance visual acuity (UDVA) of 20/32 (0.2 logMAR) or better at the time of their single diagnostic study visit. Clinical evaluation included the manifest refraction, visual acuity (VA), low contrast VA and ocular biometry.Entities:
Keywords: astigmatism correction; low contrast acuity; low cylinder; toric IOL
Year: 2020 PMID: 33154626 PMCID: PMC7608696 DOI: 10.2147/OPTH.S281178
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Demographics and Preoperative Details
| Low_Toric | Non_Toric | p | |
|---|---|---|---|
| Patients/eyes | 45/51 | 37/43 | 0.71 |
| Male/female | 23/28 | 11/32 | 0.05 |
| Age (Years) | 72 ± 5 (57 to 84) | 72 ± 9 (42 to 82) | 0.93 |
| MRSE (D) | −0.31 ± 2.43 (−4.88 to 3.63) | 0.57 ± 2.19 (−6.00 to 3.63) | 0.07 |
| Average keratometry (D) | 43.68 ± 1.53 (40.21 to 46.95) | 43.74 ± 1.26 (41.41 to 46.41) | 0.82 |
| Corneal astigmatism (D) | 0.63 ± 0.39 (0.07 to 1.46) | 0.75 ± 0.42 (0.10 to 1.45) | 0.16 |
| Anterior corneal astigmatism orientation (WTR/OBL/ATR) | 20/21/10 | 17/21/5 | 0.54 |
| Axial length (mm) | 23.73 ± 1.12 (21.82 to 26.56) | 23.47 ± 090 (21.66 to 25.78) | 0.23 |
| IOL sphere power (D) | 20.8 ± 3.8 (13.0 to 27.0) | 21.5 ± 3.3 (12.5 to 26.5) | 0.45 |
Note: Unless otherwise stated, values are: mean ± standard deviation (range).
Abbreviations: MRSE, mean refraction, spherical equivalent; D, diopter; mm, millimeter; IOL, intraocular lens; WTR, with the rule; OBL, oblique; ATR, against the rule.
Postoperative Refractive Summary by Group
| Low_Toric | Non_Toric | p | |
|---|---|---|---|
| Follow-up time (days) | 522 ± 205 (132 to 791) | 452 ± 268 (35 to 903) | 0.16 |
| MRSE (D) | 0.04 ± 0.38 (−0.63 to 1.00) | 0.18 ± 0.49 (−1.50 to 1.25) | 0.13 |
| Manifest refractive cylinder (D) | 0.31 ± 0.28 (0.00 to 1.00) | 0.53 ± 0.38 (0.0 to 1.25) | < 0.01 |
| Eyes with absolute MRSE ≤ 0.50 D | 40 (75%) | 31 (72%) | 0.48 |
| Eyes with ≤ 0.25 D of cylinder | 33 (65%) | 18 (42%) | 0.03 |
| Eyes with ≤ 0.50 D of cylinder | 43 (84%) | 30 (70%) | 0.09 |
Note: Unless otherwise stated, values are: mean ± standard deviation (range).
Abbreviations: MRSE, mean refraction, spherical equivalent; D, diopter.
Figure 1Residual refractive cylinder by group and preoperative anterior corneal astigmatism orientation.
Postoperative Visual Acuity by Group
| Low_Toric | Non_Toric | p | |
|---|---|---|---|
| Uncorrected distance VA (logMAR) | 0.01 ± 0.12 (−0.18 to 0.34) | 0.07 ± 0.13 (−0.10 to 0.44) | 0.02 |
| Best-corrected distance VA (logMAR) | −0.05 ± 0.06 (−0.20 to 0.14) | −0.03 ± 0.05 (−0.10 to 0.08) | 0.07 |
| Uncorrected distance VA, low contrast photopic (logMAR) | 0.37 ± 0.18 (0.10 to 0.80) | 0.44 ± 0.18 (0.10 to 0.92) | 0.09 |
| Uncorrected distance VA, low contrast mesopic (logMAR) | 0.59 ± 0.13 (0.36 to 0.94) | 0.62 ± 0.16 (0.32 to 1.10) | 0.13 |
| Eyes with 20/20 or better uncorrected VA | 36 (71%) | 22 (51%) | 0.05 |
Note: Unless otherwise stated, values are: mean ± standard deviation (range).
Abbreviations: VA, visual acuity; logMAR, log of the minimum angle of resolution.
Figure 2Uncorrected high contrast VA by group and preoperative anterior corneal astigmatism orientation.
SIA Data by Group
| Low_Toric | Non_Toric | p | |
|---|---|---|---|
| SIA vector magnitude (D) | 0.51 ± 0.29 (0.04 to 1.34) | 0.62 ± 0.37 (0.10 to 1.60) | 0.09 |
| SIA x coordinate (D, double angle) | −0.29 ± 0.40 (−1.28 to 0.63) | −0.15 ± 0.49 (−1.32 to 1.12) | 0.12 |
| SIA y coordinate (D, double angle) | −0.06 ± 0.33 (−0.82 to 0.53) | −0.23 ± 0.48 (−1.32 to 0.85) | 0.05 |
Note: Unless otherwise stated, values are: mean ± standard deviation (range).
Abbreviations: SIA, surgically induced astigmatism; D, diopter.