Gemeda Abebe1,2, Dossegnaw Aragaw1, Mulualem Tadesse1,2.
Abstract
BACKGROUND: Despite its low sensitivity, microscopy remains the main method for the diagnosis of pulmonary tuberculosis in most laboratories in Ethiopia. Few studies have evaluated the performance of light-emitting diode fluorescent microscopy (LED-FM) in bleach-concentrated smear-negative sputum specimens.Entities:
Keywords: Ethiopia; bleach pretreatment; health; light-emitting diode fluorescent microscopy; sputum smear-negative
Year: 2020 PMID: 33102163 PMCID: PMC7565057 DOI: 10.4102/ajlm.v9i1.810
Source DB: PubMed Journal: Afr J Lab Med ISSN: 2225-2002
FIGURE 1Flow diagram of patient recruitment, sample processing and diagnostic test results, Ethiopia, 2014.
Demographic and clinical characteristics of the study participants (n = 194), Ethiopia, 2014.
| Characteristics | Smear-negative pulmonary tuberculosis suspects | ||||||
|---|---|---|---|---|---|---|---|
| All patients ( | Culture-positive ( | Culture-negative ( | |||||
| % | % | % | |||||
| Age, median (IQR) | 38 | 23–55 | 32 | 21–47 | 40 | 26–56 | 0.27 |
| Gender, male | 118 | 60.8 | 20 | 71.4 | 98 | 59.0 | 0.21 |
| Taking antibiotics in last 2 weeks | 111 | 57.2 | 13 | 46.4 | 98 | 59.0 | 0.21 |
| Contact history with tuberculosis patient | 39 | 20.1 | 8 | 28.6 | 31 | 18.7 | 0.23 |
| History of anti-TB treatment | 51 | 26.3 | 5 | 17.9 | 46 | 27.7 | 0.27 |
| Cough > 4 weeks | 98 | 50.5 | 16 | 57.1 | 82 | 49.4 | 0.41 |
| Chest pain | 124 | 63.9 | 18 | 64.3 | 106 | 63.9 | 0.96 |
| Night sweats | 137 | 70.6 | 22 | 78.6 | 115 | 69.3 | 0.32 |
| Fever (> 38.5 °C) | 129 | 66.5 | 20 | 71.4 | 109 | 65.7 | 0.55 |
| Weight loss | 115 | 59.3 | 19 | 67.9 | 96 | 57.8 | 0.31 |
| Shortness of breath | 127 | 66.5 | 16 | 57.1 | 111 | 66.9 | 0.32 |
| Loss of appetite | 135 | 69.6 | 22 | 78.6 | 113 | 68.1 | 0.26 |
| Weakness | 158 | 81.4 | 23 | 82.1 | 135 | 81.3 | 0.91 |
| Haemoptysis | 26 | 13.4 | 6 | 21.4 | 20 | 12.0 | 0.18 |
TB, tuberculosis; IQR, interquartile range.
Diagnostic performance of light-emitting diode fluorescence microscopy compared to reference standard (culture), Ethiopia, 2014.
| Diagnostic accuracy of LED-FM | Value | 95% confidence interval |
|---|---|---|
| Sensitivity | 39.3% | 21.2% – 57.4% |
| Specificity | 93.9% | 90.4% – 97.6% |
| Positive predictive value | 52.4% | 31.0% – 73.7% |
| Negative predictive value | 90.2% | 85.7% – 94.6% |
| Likelihood ratio positive | 6.5 | 3.0 – 13.9 |
| Likelihood ratio negative | 0.65 | 0.48 – 0.87 |
| Pre-test probability | 14.4% | - |
| Positive post-test probability | 52.4% | - |
LED-FM, light-emitting diode fluorescence microscopy.