C Michotte1, M Nonis1, D Bergeron2, J Cessna2, R Fitzgerald2, L Pibida2, B Zimmerman2, A Fenwick3, K Ferreira3, J Keightley3, I Da Silva4.
Abstract
In 2016, comparisons of activity measurements of 18F and 64Cu using the Transfer Instrument of the International Reference System (SIRTI) took place at the National Institute of Standards and Technology (NIST, USA). This is the first SIRTI comparison for 64Cu. Ampoules containing about 27 kBq of 18F and 100 kBq of 64Cu solutions were measured in the SIRTI for about 5 and 1.5 half-lives, respectively. The NIST standardized the activity in the ampoules by ionization chamber measurements traceable to 4π(LS)β-γ anticoincidence measurements. The comparisons, identifiers BIPM.RI(II)-K4.F-18 and BIPM.RI(II)-K4.Cu-64, are linked to the corresponding BIPM.RI(II)-K1.F-18 and BIPM.RI(II)-K1.Cu-64 comparisons and degrees of equivalence with the respective key comparison reference values have been evaluated. The NIST replaces its earlier degree of equivalence for 18F obtained in the frame of the CCRI(II)-K3.F-18 comparison in 2001.Entities:
Year: 2017 PMID: 33100400 PMCID: PMC7579768
Source DB: PubMed Journal: Metrologia ISSN: 0026-1394 Impact factor: 3.157