Florian A Schmid1, Dominic Gascho2, Niklaus Zoelch2, Jenny A Prange1, Giovanni Colacicco3, Daniel Eberli4.
Abstract
BACKGROUND: The injection of muscle precursor cells (MPC) into the external urinary sphincter muscle (EUS) is a promising therapeutic option for regenerative treatment of stress urinary incontinence (SUI). The objective of the present project was to conduct a pre-clinical trial to investigate the feasibility and accuracy of ultrasound (US) guided, transurethral injections into the EUS of female cadavers.Entities:
Keywords: Female; MRI; Transurethral injection; Ultrasound-guided; Urinary sphincter
Mesh:
Year: 2020 PMID: 33097063 PMCID: PMC7583166 DOI: 10.1186/s12894-020-00719-x
Source DB: PubMed Journal: BMC Urol ISSN: 1471-2490 Impact factor: 2.264
Fig. 1a Anatomy of the lower female pelvis in 3D US with transversal view and b MRI T2 sequence in axial view. c Setting before transurethral injection of nano-iron particles. Extracorporeal setup, where vaginal US probe and urethral therapy shaft are linked together. d Needle puncture and injection are performed through the therapy shaft
Fig. 2a US image in sagittal orientation during transurethral injection. b MRI image in axial orientation after injection of nano-iron particles. c, d Scoring of injection defined on a rating scale from 1 to 4 points according to extinction signal of injected iron nano-particles in MRI images. c Axial orientation with division into right and left, d sagittal orientation with division into anterior and posterior. Diameter and distances in (a) and (b), respectively: 1 = 5 mm, 2 = 10 mm, 3 = 15 mm, 4 > 15 mm
Fig. 3Evaluation of the ratio of signal intensity in the EUS muscle in MRI. Measured differences due to extinction before and after injection of nano-iron particles in Cadaver 3. The red area marks the EUS in all parts of the figure. a Overall, b zoom
Scores per cadaver on the rating scale according to extinction signal in MRI image in axial (left, right) and sagittal (anterior, posterior) orientation
| Cadaver | Axial left | Axial right | Sagittal anterior | Sagittal posterior | Overall mean | Radius (mm) | Area (mm2) | Volume measured in MR (cm3) | SI-ratio |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 4 | 4 | 4 | 4 | 4 | 2.5 | 19.6 | 3.82 | 0.055 |
| 2 | 2 | 4 | 4 | 1 | 2.75 | 5.0 | 78.5 | 0.32 | 0.526 |
| 3 | 3 | 2 | 4 | 3 | 3 | 7.5 | 176.7 | 1.11 | 0.404 |
| Column mean | 3 | 3.33 | 4 | 2.67 | 3.25 | 5 | 91.60 | 1.75 | 0.33 |
Measured radius, area and volume of injection. Signal-intensity ratio before and after injection in the EUS
SI signal-intensity, MR magnetic resonance