Literature DB >> 33086208

Microdosimetry of a therapeutic proton beam with a mini-TEPC and a MicroPlus-Bridge detector for RBE assessment.

V Conte1, S Agosteo2,3, A Bianchi1,4,5, D Bolst6, D Bortot2,3, R Catalano7, G A P Cirrone7, P Colautti1, G Cuttone7, S Guatelli6, B James6, D Mazzucconi2,3, A B Rosenfeld6, A Selva1, L Tran6, G Petringa7.   

Abstract

Proton beams are widely used worldwide to treat localized tumours, the lower entrance dose and no exit dose, thus sparing surrounding normal tissues, being the main advantage of this treatment modality compared to conventional photon techniques. Clinical proton beam therapy treatment planning is based on the use of a general relative biological effectiveness (RBE) of 1.1 along the whole beam penetration depth, without taking into account the documented increase in RBE at the end of the depth dose profile, in the Bragg peak and beyond. However, an inaccurate estimation of the RBE can cause both underdose or overdose, in particular it can cause the unfavourable situation of underdosing the tumour and overdosing the normal tissue just beyond the tumour, which limits the treatment success and increases the risk of complications. In view of a more precise dose delivery that takes into account the variation of RBE, experimental microdosimetry offers valuable tools for the quality assurance of LET or RBE-based treatment planning systems. The purpose of this work is to compare the response of two different microdosimetry systems: the mini-TEPC and the MicroPlus-Bridge detector. Microdosimetric spectra were measured across the 62 MeV spread out Bragg peak of CATANA with the mini-TEPC and with the Bridge microdosimeter. The frequency and dose distributions of lineal energy were compared and the different contributions to the spectra were analysed, discussing the effects of different site sizes and chord length distributions. The shape of the lineal energy distributions measured with the two detectors are markedly different, due to the different water-equivalent sizes of the sensitive volumes: 0.85 μm for the TEPC and 17.3 μm for the silicon detector. When the Loncol's biological weighting function is applied to calculate the microdosimetric assessment of the RBE, both detectors lead to results that are consistent with biological survival data for glioma U87 cells. Both the mini-TEPC and the MicroPlus-Bridge detector can be used to assess the RBE variation of a 62 MeV modulated proton beam along its penetration depth. The microdosimetric assessment of the RBE based on the Loncol's weighting function is in good agreement with radiobiological results when the 10% biological uncertainty is taken into account.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 33086208     DOI: 10.1088/1361-6560/abc368

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Microdosimetry performance of the first multi-arrays of 3D-cylindrical microdetectors.

Authors:  Diana Bachiller-Perea; Mingming Zhang; Celeste Fleta; David Quirion; Daniela Bassignana; Faustino Gómez; Consuelo Guardiola
Journal:  Sci Rep       Date:  2022-07-18       Impact factor: 4.996

2.  Mapping the Relative Biological Effectiveness of Proton, Helium and Carbon Ions with High-Throughput Techniques.

Authors:  Lawrence Bronk; Fada Guan; Darshana Patel; Duo Ma; Benjamin Kroger; Xiaochun Wang; Kevin Tran; Joycelyn Yiu; Clifford Stephan; Jürgen Debus; Amir Abdollahi; Oliver Jäkel; Radhe Mohan; Uwe Titt; David R Grosshans
Journal:  Cancers (Basel)       Date:  2020-12-05       Impact factor: 6.639

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.