Ahnika Kline1, Nicole E Putnam2, Jung-Ho Youn2, Amanda East2, Sanchita Das2, Karen M Frank2, Adrian M Zelazny3.
Abstract
Nasopharyngeal flocked swabs placed in viral transport media (VTM) are the preferred collection methodology for respiratory virus testing. Due to the rapid depletion of available reagents and swabs, we have validated an alternative swab placed in phosphate-buffered saline (PBS) for use in respiratory virus testing in a SARS-CoV-2 real-time polymerase chain reaction assay and a multiplexed respiratory virus panel. We collected nasopharyngeal (NP) swabs and oropharyngeal (OP) swabs from 10 healthy volunteers. Flocked swabs were placed in VTM and alternative swabs in PBS. In this feasibility study, we show that NP collection is better for detection of human material than OP collection, as measured by significantly lower RNase P gene cycle threshold values, and that a Dacron polyester swab in PBS shows equivalent detection of SARS-CoV-2 and RSV to a flocked swab in VTM in contrived specimens. Diluted SARS-CoV-2-positive patient specimens are detectable for up to 72 h at 4 °C. Published by Elsevier Inc.Entities:
Keywords: 2019-nCOV; Alternative reagents; COVID-19; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 33080426 PMCID: PMC7492856 DOI: 10.1016/j.diagmicrobio.2020.115209
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Fig. 1Dacron polyester swab in PBS shows equivalent performance and stability up to 72 h compared to a flocked nylon swab in VTM. Nasopharyngeal and oropharyngeal collections were obtained using a flocked nylon swab placed into VTM (red, circle) or a Dacron polyester swab and placed into PBS (blue, square). Ct values for detection of the human RNase P (RP1) gene were measured by RT-PCR from 10 healthy volunteers at 24 h (light red, light blue) or 72 h (dark red, dark blue). Each of 8 collection condition and time point combinations was tested on 10 healthy volunteers, with paired comparisons shown between flocked swab/VTM and Dacron swab/PBS collection methods. A 3-way ANOVA with Tukey's multiple-comparisons test was performed to compare Ct values between NP and OP collection types in each swab/media combination at 24 and 72 h. Multiple comparisons show that at both 24 and 72 h, the Ct values between NP-VTM and OP-VTM and NP-PBS and OP-PBS differ significantly, with all 4 comparisons giving P < 0.0001. Thus, the collection type (NP versus OP) led to a significant source of variation (P < 0.0001).
PBS shows equivalent performance to VTM and stability at 24 and 72 h for the detection of RSV using the BioFire FilmArray Respiratory Pathogen Panel. NP collections from flocked swabs in VTM and Dacron swabs in PBS (n = 10 each) were spiked with 10×-LOD of RSV to give a final concentration of 20 TCID50/mL. The results for RSV on the Respiratory Pathogen Panel were detected (+) or not detected (−).
| Type | Vehicle/swab | Time point | BioFire FilmArray RP Panel results | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| NP | VTM/flocked | 24 h | − | + | + | + | + | + | + | + | + | + |
| 72 h | + | + | + | + | + | + | − | + | + | + | ||
| NP | PBS/Dacron | 24 h | + | + | + | + | + | + | + | + | + | + |
| 72 h | + | + | + | + | + | + | + | + | + | + | ||
Fig. 2PBS and VTM show equivalent detection of heat-inactivated SARS-CoV-2 and human material. Nasopharyngeal collections were obtained using a flocked swab placed into VTM (solid symbols) or a Dacron polyester swab and placed into PBS (open symbols). Ct values for detection of the N1 (green) and N2 (blue) nucleocapsid gene segments and internal control RP1 (magenta) gene were measured at time 0 (n = 3), 24 h (n = 5), and 72 h (n = 5) from media spiked with heat-inactivated SARS-CoV-2. One sample in the VTM 72-h condition failed the nucleic acid extraction step. Each symbol represents the Ct value from 1 sample, with VTM and PBS paired symbols connected by lines, represented as time 0 (gray dashed line), 24 h (black dotted line), and 72 h (black solid line).
Fig. 3PBS shows similar detection of diluted positive SARS-CoV-2 patient specimens compared to VTM using the SARS-CoV-2 real-time PCR assay. Ten known positive SARS-CoV-2 patient specimens with varying Ct values were diluted 1:10 into VTM from a flocked swab NP collection (green circles) or PBS from a Dacron swab NP collection (blue squares). Paired Ct values from each diluted sample are shown for the N1 and N2 genes at 0 h and 72 h after dilution.