Gert Marais1,2, Michelle Naidoo1,2, Nei-Yuan Hsiao1,2, Ziyaad Valley-Omar1,3, Heidi Smuts1,2, Diana Hardie1,2.
Abstract
The SARS-CoV-2 pandemic has resulted in shortages of both critical reagents for nucleic acid purification and highly trained staff as supply chains are strained by high demand, public health measures and frequent quarantining and isolation of staff. This created the need for alternate workflows with limited reliance on specialised reagents, equipment and staff. We present here the validation and implementation of such a workflow for preparing samples for downstream SARS-CoV-2 RT-PCR using liquid handling robots. The rapid sample preparation technique evaluated, which included sample centrifugation and heating prior to RT-PCR, showed a 97.37% (95% CI: 92.55-99.28%) positive percent agreement and 97.30% (95% CI: 90.67-99.52%) negative percent agreement compared to nucleic acid purification-based testing. This method was subsequently adopted as the primary sample preparation method in the Groote Schuur Hospital Virology Diagnostic Laboratory in Cape Town, South Africa.Entities:
Mesh:
Substances:
Year: 2020 PMID: 33079951 PMCID: PMC7575110 DOI: 10.1371/journal.pone.0241029
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
RT-PCR reaction preparation.
| Volume per reaction (μl) | |
|---|---|
| 11.1 | |
| 6 | |
| 6 | |
| 2.4 | |
| 1.5 | |
| 3 | |
| 30 |
Contingency table used for positive and negative percent agreement with NA purification-based testing calculation.
| Positive SARS-CoV-2 | Negative SARS-CoV-2 | |
|---|---|---|
| Abbott RealTime SARS-CoV-2 Assay or Seegene AllplexTM 2019-nCoV Assay | Abbott RealTime SARS-CoV-2 Assay or Seegene AllplexTM 2019-nCoV Assay | |
| NA Purification | NA Purification | |
| 111 | 2 | |
| 3 | 72 |
Fig 1Comparison of target Ct values after RSP and NucliSENS® easyMag NA purification.
The Ct values for the SARS-CoV-2 (A) Envelope (E), (B) RNA-dependent RNA-polymerase (RdRp) and (C) Nucleocapsid (N) gene targets are shown for samples tested with the Allplex™ 2019-nCoV assay after NucliSENS® easyMag® NA purification and RSP. The difference in generated Ct values was found to be statistically significant in each case with a P value of <0.0001 as determined by paired t-test.
Fig 2Comparison of target Ct and CN values after RSP and testing with the Abbott RealTime SARS-CoV-2 assay.
The Ct values for the SARS-CoV-2 Envelope (E), RNA-dependent RNA-polymerase (RdRp) and Nucleocapsid (N) gene targets are shown for samples tested with the Allplex™ 2019-nCoV assay after RSP and CN values after testing with the Abbott RealTime SARS-CoV-2 assay. A plotted CN or Ct value of 40 indicates that detectable amplification did not occur. The Abbott assay CN values are assay specific and not directly comparable to Ct values, but are shown to demonstrate the performance of the spectrum of selected samples.
Relative analytical sensitivity assessment.
| Dilution | Replicates | Seegene AllplexTM 2019-nCoV Assay | Seegene AllplexTM 2019-nCoV Assay |
|---|---|---|---|
| RSP Method | NA Purification | ||
| Percentage of Samples Positive | Percentage of Samples Positive | ||
| 24 | 100% | Not done | |
| 24 | 95.8% | Not done | |
| 24 | 70.8% | Not done | |
| 24 | 58.3% | Not done | |
| 24 | 41.7% | Not done | |
| 10 | Not done | 100% | |
| 24 | 33.3% | Not done | |
| 10 | Not done | 100% | |
| 10 | Not done | 90% | |
| 10 | Not done | 70% | |
| 10 | Not done | 60% | |
| 10 | Not done | 30% |