The influence of respiratory motion on dose distribution in accelerated partial breast irradiation using volumetric modulated arc therapy.

Ryohei Yamauchi1, Norifumi Mizuno2, Tomoko Itazawa2, Jiro Kawamori2.   

Abstract

PURPOSE: Accelerated partial breast irradiation (APBI) is alternative treatment option for patients with early stage breast cancer. The interplay effect on volumetric modulated arc therapy APBI (VMAT-APBI) has not been clarified. This study aimed to evaluate the feasibility of VMAT-APBI for patients with small breasts and investigate the amplitude of respiratory motion during VMAT-APBI delivery that significantly affects dose distribution.
METHODS: The VMAT-APBI plans were generated with 28.5 Gy in five fractions. We performed patient-specific quality assurance using Delta4 phantom under static conditions. We also measured point dose and dose distribution using the ionization chamber and radiochromic film under static and moving conditions of 2, 3 and 5 mm. We compared the measured and calculated point doses and dose distributions by dose difference and gamma passing rates.
RESULTS: A total of 20 plans were generated; the dose distributions were consistent with those of previous reports. For all measurements under static conditions, the measured and calculated point doses and dose distributions showed good agreement. The dose differences for chamber measurement were within 3%, regardless of moving conditions. The mean gamma passing rates with 3%/2 mm criteria in the film measurement under static conditions and with 2 mm, 3 mm, and 5 mm of amplitude were 95.0 ± 2.0%, 93.3 ± 3.3%, 92.1 ± 6.2% and 84.8 ± 7.8%, respectively. The difference between 5 mm amplitude and other conditions was statistically significant.
CONCLUSIONS: Respiratory management should be considered for the risk of unintended dose distribution if the respiratory amplitude is >5 mm.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Accelerated partial breast irradiation; Breast cancer; Intensity modulated radiation therapy; Respiratory motion

Mesh:

Year:  2020        PMID: 33075732     DOI: 10.1016/j.ejmp.2020.09.024

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Residual image registration error by fiducial markers in accelerated partial breast irradiation using C-arm linac: a phantom study.

Authors:  Ryohei Yamauchi; Natsuki Murayoshi; Shinobu Akiyama; Norifumi Mizuno; Tomoyuki Masuda; Tomoko Itazawa; Jiro Kawamori
Journal:  Phys Eng Sci Med       Date:  2022-06-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.