| Literature DB >> 33071278 |
Tian-Tian Chang1, Ya-Nan Feng2, Yi Zhu3, Chun-Long Liu4, Xue-Qiang Wang1, Zhi-Jie Zhang2.
Abstract
BACKGROUND Achilles tendinopathy commonly occurs in specific regions of the tendon, and Achilles tendon stiffness can be related to local pathological changes in the tendon. The MyotonPRO is a new handheld device that conveniently assesses stiffness of muscles and tendons. This study aimed to 1) evaluate the intra- and inter-rater reliability of stiffness measurements of the Achilles tendon at different ankle positions, 2) investigate the modulation of stiffness at different ankle joint angles, and 3) examine the differences between 2 regions of Achilles tendon stiffness. MATERIAL AND METHODS Thirty healthy young adults (15 men and 15 women) participated in this study. The regional Achilles tendon stiffness at 0 cm (AT-0) and 6 cm (AT-6) above the tendon insertion were evaluated by the MyotonPRO in the neutral position and 10° dorsiflexion of the ankle joint. Measurements of stiffness were taken by 2 raters on the first day and 5 days later. The stiffness data were compared by repeated measures analysis of variance (ANOVA). RESULTS The intra- and inter-rater reliability of stiffness measurements at AT-0 and AT-6 for each ankle position were good (all intraclass correlation coefficients >0.84). A significant modulation of Achilles tendon stiffness was obtained at different ankle joint angles (P<0.05). Stiffness at AT-0 was higher than at AT-6 (P<0.05) in both positions. CONCLUSIONS These results suggest the MyotonPRO reliably assessed Achilles tendon stiffness and monitors its modulation, and tendon stiffness increased with ankle dorsiflexion. Stiffness was also nonuniform along the length of the tendon.Entities:
Mesh:
Year: 2020 PMID: 33071278 PMCID: PMC7583434 DOI: 10.12659/MSM.926407
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 1Photograph of the splint.
Inter-and intra-operator reliability of Achilles tendon at the neutral ankle neutral. ICC, intraclass correlation coefficients; CI, confidence interval; SEM (N/m), the standard error of measurement; MDC (N/m), the minimal detectable change; A1B, inter-rater reliability; A1A2, intra-rater reliability.
| Ankle neutral position | AT-0 | AT-6 | ||||
|---|---|---|---|---|---|---|
| Mean±SD | SEM | MDC | Mean±SD | SEM | MDC | |
| Test A1 | 1201.28±309.86 | 56.57 | 156.80 | 829.35±116.61 | 21.29 | 59.01 |
| Test B | 1193.05±335.74 | 61.30 | 169.91 | 837.67±112.27 | 20.50 | 56.82 |
| Test A2 | 1198.30±337.61 | 61.64 | 170.86 | 833.20±108.54 | 19.82 | 54.94 |
| ICCA1B (95% CI) | 0.93 (0.85–0.97) | 0.89 (0.75–0.95) | ||||
| ICCA1A2 (95% CI) | 0.92 (0.83–0.96) | 0.93 (0.86–0.97) | ||||
Inter-and intra-operator reliability of Achilles tendon at the ankle 10° dorsiflexion. ICC, intraclass correlation coefficients; CI, confidence interval; SEM(N/m), the standard error of measurement; MDC(N/m), the minimal detectable change; A1B, inter-rater reliability; A1A2, intra-rater reliability.
| Ankle dorsi-flexion 10° | AT-0 | AT-6 | ||||
|---|---|---|---|---|---|---|
| Mean±SD | SEM | MDC | Mean±SD | SEM | MDC | |
| Test A1 | 1409.73±373.63 | 68.22 | 189.10 | 969.26±108.60 | 19.83 | 54.97 |
| Test B | 1399.07±334.26 | 61.03 | 171.72 | 947.07±115.02 | 21.00 | 58.21 |
| Test A2 | 1392.07±338.23 | 61.75 | 171.16 | 956.97±120.55 | 22.00 | 60.98 |
| ICCA1B (95% CI) | 0.94 (0.87–0.97) | 0.85 (0.69–0.93) | ||||
| ICCA1A2 (95% CI) | 0.92 (0.82–0.96) | 0.87 (0.74–0.94) | ||||
Figure 2Bland-Altman plots of intra- and inter-rater reliability of Achilles tendon stiffness. (A, B) Intra- and inter-rater reliability of AT-0 stiffness at ankle neutral position. (C, D) Intra and inter-rater reliability of AT-6 stiffness at ankle neutral position. (E, F) Intra- and inter-rater reliability of AT-0 stiffness at ankle dorsiflexion 10°. (G, H) Intra- and inter-rater reliability of AT-6 stiffness at ankle dorsiflexion 10°.
Figure 3The mean stiffness of AT-0 and AT-6 at ankle joint 0° (deep blue) and dorsiflexion 10° (light blue). The difference between AT-0 and AT-6 mean stiffness at 0° and dorsiflexion 10°. *** P<0.001.