Evaluation of a collapsed-cone convolution algorithm for esophagus and surface mold 192Ir brachytherapy treatment planning.

Luke Eason1, Josh Mason2, Rachel Cooper2, Ganesh Radhakrishna3, Peter Bownes2.   

Abstract

PURPOSE: TG43 does not account for a lack of scatter and tissue and applicator heterogeneities. The advanced collapsed-cone engine (ACE) algorithm available for use in the Oncentra Brachy treatment planning system (Elekta AB, Stockholm, Sweden) can model these conditions more accurately and is evaluated for esophageal and surface mold brachytherapy treatments. METHODS AND MATERIALS: ACE was commissioned for use then compared against TG43 for five esophageal and five surface mold treatment plans. Dosimetric differences between each algorithm were assessed using superimposed comparisons and dose-volume histogram statistics.
RESULTS: Esophagus (6 Gy per fraction): Compared with TG43, ACE demonstrated up to a 0.63% and 0.05 Gy reduction in planning target volume (PTV) V100% and PTV D98, respectively. Lung D2cc and bone D2cc deviated by up to 0.09 Gy and 0.03 Gy, respectively. Lung D0.1 cc and bone D0.1 cc both deviated by up to 0.12 Gy. Surface mold (4.5 Gy per fraction): Compared with TG43, ACE demonstrated up to a 12.5% and 0.18 Gy reduction in PTV V80% and PTV D98, respectively. Bone D2cc and D0.1 cc both reduced by up to 0.2 Gy when modeled with ACE. Increasing mold size laterally increased the dosimetric differences between TG43 and ACE.
CONCLUSIONS: TG43 generally overestimated dose delivered to the target volume and organs at risk for the sites investigated. Dosimetric differences observed for esophageal treatments were minimal; however, surface mold treatments would benefit from the increased dosimetric accuracy offered by ACE. Implementation should be considered for surface mold 192Ir treatment planning, but increased calculation time, additional contouring, and mass density assignment requirements should be scrutinized with regard to their potentially negative impact on current clinical practice.
Copyright © 2020 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  192Ir; Brachytherapy; Esophagus; MBDCA; Surface mold; Treatment planning

Year:  2020        PMID: 33071170     DOI: 10.1016/j.brachy.2020.09.006

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  2 in total

Review 1.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

2.  Dose calculation accuracy in particle therapy: Comparing carbon ions with protons.

Authors:  Sirinya Ruangchan; Hugo Palmans; Barbara Knäusl; Dietmar Georg; Monika Clausen
Journal:  Med Phys       Date:  2021-09-23       Impact factor: 4.506

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.