| Literature DB >> 33070848 |
Huisi Yang1, Jing Bao1, Yanli Qi1, JiaYing Zhao1, Yian Hu1, Weixuan Wu1, Xicheng Wu1, Daidi Zhong2, Danqun Huo3, Changjun Hou4.
Abstract
Herein, a three-dimensional graphene wall (GWs) and nano-Cu2O modified carbon fiber paper (CFP) electrode were used to develop a disposable and sensitive non-enzymatic glucose sensor. This sensing interface of GWs/Cu2O consists of an interlaced CFP on which intercrossed graphene walls (GWs) were vertically tethered in situ by radio frequency plasma enhanced chemical vapor deposition (RF-PECVD), and Cu2O nanoparticles (NPs) were evenly grew on the 3D GWs layer and skeleton through the complete thermal decomposition of copper acetate (Cu (CH3COO)2) at high temperature. The CFP/GWs/Cu2O shows a large specific surface area and rich solution diffusion channels, which can expose more catalytic active sites without Nafion fixation film, thus greatly improving the electrocatalytic performance of this glucose sensor. The CFP/GWs/Cu2O sensor shows excellent catalytic performance to glucose with a linear detection range of 0.5 μM-5166 μM, LOD of 0.21 μM, and response time <4 s. This kind of disposable and sensitive electrode can capable of controlling uniform growth and accurate quantification, and has great development potential in the field of medical detection and the commercialization of wearable sensors.Entities:
Keywords: CFP; Cu(2)O; Disposable electrode; Glucose; Non-enzymatic
Mesh:
Substances:
Year: 2020 PMID: 33070848 DOI: 10.1016/j.aca.2020.08.010
Source DB: PubMed Journal: Anal Chim Acta ISSN: 0003-2670 Impact factor: 6.558