Piotr Deptuła1, Dawid Łysik2, Katarzyna Pogoda3, Mateusz Cieśluk1, Andrzej Namiot4, Joanna Mystkowska2, Grzegorz Król5, Stanisław Głuszek6,7, Paul A Janmey8,9, Robert Bucki1. 1. Department of Medical Microbiology and Nanobiomedical Engineering, Medical University of Bialystok, 15-222 Bialystok, Poland. 2. Institute of Biomedical Engineering, Bialystok University of Technology, 15-351 Bialystok, Poland. 3. Institute of Nuclear Physics, Polish Academy of Sciences, PL-31342 Krakow, Poland. 4. Department of Human Anatomy, Medical University of Bialystok, 15-230 Bialystok, Poland. 5. Department of Microbiology and Immunology, Jan Kochanowski University, 25-516 Kielce, Poland. 6. Institute of Medical Sciences, Collegium Medicum, Jan Kochanowski University, 25-369 Kielce, Poland. 7. Clinic for General, Oncologic and Endocrine Surgery, Regional Hospital, 25-736 Kielce, Poland. 8. Institute for Medicine and Engineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States. 9. Departments of Physiology and Physics & Astronomy, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States.
Abstract
In recent years, rheological measurements of cells and tissues at physiological and pathological stages have become an essential method to determine how forces and changes in mechanical properties contribute to disease development and progression, but there is no standardization of this procedure so far. In this study, we evaluate the potential of nanoscale atomic force microscopy (AFM) and macroscopic shear rheometry to assess the mechanical properties of healthy and cancerous human colon tissues. The direct comparison of tissue mechanical behavior under uniaxial and shear deformation shows that cancerous tissues not only are stiffer compared to healthy tissue but also respond differently when shear and compressive stresses are applied. These results suggest that rheological parameters can be useful measures of colon cancer mechanopathology. Additionally, we extend the list of biological materials exhibiting compressional stiffening and shear weakening effects to human colon tumors. These mechanical responses might be promising mechanomarkers and become part of the new procedures in colon cancer diagnosis. Enrichment of histopathological grading with rheological assessment of tissue mechanical properties will potentially allow more accurate colon cancer diagnosis and improve prognosis.
In recent years, rheological measurements of cells and tissues at physiological and pathological stages have become an essential method to determine how forces and changes in mechanical properties contribute to disease development and progression, but there is no standardization of this procedure so far. In this study, we evaluate the potential of nanoscale atomic force microscopy (AFM) and macroscopic shear rheometry to assess the mechanical properties of healthy and cancerous human colon tissues. The direct comparison of tissue mechanical behavior under uniaxial and shear deformation shows that cancerous tissues not only are stiffer compared to healthy tissue but also respond differently when shear and compressive stresses are applied. These results suggest that rheological parameters can be useful measures of colon cancer mechanopathology. Additionally, we extend the list of biological materials exhibiting compressional stiffening and shear weakening effects to humancolon tumors. These mechanical responses might be promising mechanomarkers and become part of the new procedures in colon cancer diagnosis. Enrichment of histopathological grading with rheological assessment of tissue mechanical properties will potentially allow more accurate colon cancer diagnosis and improve prognosis.
Colorectal
carcinomas are one of the most common types of cancers
diagnosed in patients worldwide, and they are a leading cause of death
in both Europe and the United States.[1−3] Although over the past
decade, the incidence and mortality of colorectal cancer have decreased,
20–25% of patients with colon cancer have metastases at the
time of diagnosis and a large percentage (up to 60%) of the others
will develop metastases later. Most of those cases of metastasis are
fatal.[1,4,5] More than 90%
of colorectal carcinomas are adenocarcinomas originating from epithelial
cells of the colorectal mucosa.[6] Smaller
fractions of colorectal carcinomas include neuroendocrine, squamous
cell, adenosquamous, and spindle cell tumors.[6] Histopathology of colon cancers includes accurate assessment of
their origin and pathological stage as well as analysis of surgical
tissue margins.[6] For colorectal carcinomas,
histopathology is the clinical standard technique and an important
final confirmation of diagnosis, which will determine the patients’
therapy, treatment options, and potential outcomes.[6,7] Because
of tissue handling steps required for histopathological grading, limitations
resulting from the tissue shape, sample size, and orientation make
this analysis time-consuming and might give misleading and inconsistent
results.[6−8] Despite significant advances in cancer diagnostic
methods, new early detection techniques that speed up the time of
diagnosis, especially for patients at the early stage of the disease,
are needed. In recent years, more attention is paid not only to morphological
and molecular screening of the tissue samples but also to their mechanical
properties, in order to fully understand the physiological and pathological
processes at the cell and tissue level.[9−17]All physiological processes and their malfunction during disease
development require structural changes that manifest at different
organization levels in organs, tissues, single cells, and cellular
organelles. All of these structures have specific mechanical properties
that can be characterized by rheological parameters. However, relating
structural elements to mechanical properties is especially complex
within biological samples and depends on the internal cell rearrangements
and cellular interactions with molecules that compose the extracellular
matrix (ECM) of stroma (e.g., fibrous proteins, proteoglycans, hyaluronic
acid, chemokines, growth factors). Most of the factors that determine
tissue/cell rheology cannot presently be assessed during the histological
evaluation, even if they tightly connect to the health status of the
organism,[9] and these properties are important
in maintaining proper functionality of the organ.[10,11,18] Very often, tissues increase their stiffness
during the pathological development and progression of cancer,[19−22] and some types of tumors can even be detected by physical palpation.[23] Cancerous tissues are characterized by an abnormal
altered stroma that facilitates cancer development by providing nutritional
support and imposing a barrier for host defense mechanisms.[24] The tumor stroma consists of fibroblasts, immune
cells, vasculature, fibrillar proteins, and an ECM composed mainly
of collagen I, fibronectin, and elastin as well as hyaluronan and
other glycosaminoglycans.[24,25] Abnormal changes in
stroma include an increase in ECM stiffness and an accumulation of
stress gradients inside the tumor mass.[24,26] Abnormal mechanical
stresses can increase the invasive and metastatic potential and migration
of cancer cells and tissue development.[27−29] Mechanical effects are
likely to be especially important in colon cancer because human colon
cancer-derived cell lines show strong responses to substrate stiffness
and can initiate metastasis-related phenotypes in a stiffness-dependent
manner.[30,31]To analyze the mechanical properties
of soft biological samples,
various techniques, such as atomic force microscopy (AFM), shear rheometry,
micropipette aspiration, optical stretching, and magnetic twisting
cytometry, have been developed.[11] Atomic
force microscopes and shear rheometers, which can quantify the mechanical
properties of soft tissues, are promising tools in terms of cancer
diagnosis and assessment of anticancer treatment effectiveness. AFM
offers the capability of surface imaging at the nanoscale and nanomechanical
characterization by indentation of soft biological materials such
as cells and tissues under physiological conditions.[11,13,14,20,32,33] Not only is
AFM a useful tool to assess the mechanics of biological samples, but
also there are attempts to apply it for routine cancer diagnosis.[13,34] It is also possible to characterize healthy and diseased tissues
in vivo using noninvasive techniques such as shear wave elastography
(SWE) and magnetic resonance elastography (MRE).[35−38] Elastography is an imaging technique
that infers stiffness by examining a tissue’s response to externally
applied mechanical excitation. Excitation leads to the tissue deformation,
which can be measured.[39] Despite their
undeniable advantages, these methods have drawbacks due to the complex
structure of the tissues and the location of the organs away from
the source of excitation and because the results strongly depend on
the excitation frequency, which is often greater than the time scale
to which cells respond.[35−37] A combination of ex vivo rheometric
analysis with in vivo measurements by elastography has the potential
to increase reliability and clinical utility of mechanical measurements.
Recently, the idea of employing shear rheometers to evaluate the mechanical
state of cells and tissues in the clinical routine has gained interest,[14,22,40] but there is a need to develop
and describe useful biomarker(s), or more precisely mechanomarker(s),
that will be a measurement outcome and could support current diagnosis.
Moreover, when it comes to clinical oncology, such mechanomarker(s)
must be well characterized and repeatedly predict the relevant stage
of the disease. The aim of this study was to assess the potential
of atomic force microscopy and shear rheometry to measure mechanical
properties of fresh human colon tissues and to verify if there are
clear mechanical indicators of the cancer state that could support
histopathological scoring and strengthen cancer diagnosis and/or classification.
Experimental Section
Materials and Methods
Tissue Samples
Mechanical properties
of fresh human healthy and cancer colon tissues were tested by atomic
force microscopy and shear rheometry. Biopsy specimens were obtained
from four patients diagnosed with colorectal carcinomas at the Clinical
Department of General Oncological and Endocrinological Surgery, Regional
Hospital, Kielce, Poland. The collection of tissue samples was performed
in accordance with an IRB (no 18/2019 approved by the Bioethics Committee
of the Faculty of Medicine and Health Sciences, Jan Kochanowski University,
Kielce, Poland). All tissue samples were stored in Dulbecco’s
Modified Eagle Medium (DMEM; Sigma-Aldrich, St. Louis, MO) and measured
within 5 h postsurgery. The patient population included both males
and females. A sample of cancerous and healthy tissue (a margin) was
collected from each patient. Healthy tissue was dissected in order
to reach the tumor tissue. All diseased tissue samples were collected
from patients with Stage III cancer, where the tumor had grown to
a specific size. We confirmed the cancer stage using standard histological
procedures. No cancer metastases were found in the patients. Healthy
tissues without physical interaction with the tumor were harvested
from a location in the colon 10 cm away from the tumor mass. Tissue
samples were harvested from the descending and sigmoid colon.
Rheological Characterization
In
the first step of our investigations, small, millimeter-scale samples
obtained from the biopsy of humantumors as well as healthy colon
tissues were measured with a NanoWizard 4 BioScience JPK Instruments
atomic force microscope (AFM) working in the force spectroscopy mode.
Force indentation curves were collected using a silicon nitride cantilever
with a nominal spring constant of 0.62 N/m and measured spring constant
in the range of 0.4–0.6 N/m using the thermal tune method,
with a 4.5 μm diameter polystyrene bead attached. The cantilevers
were manufactured by Novascan Technologies, Inc. (Figure A).
Figure 1
AFM experimental setup:
(A) The main AFM components are a cantilever
with a spherical tip (4.5 μm in diameter bead), a laser source,
a photosensitive photodiode, and a piezoelectric scanner that can
accurately apply compressive force to soft tissues at the nanoscale.
The application of the compressive force measured as a function of
the sample’s position in the Z-direction gives
rise to so-called force vs distance curves (B). The difference between
the cantilever deflection on a stiff glass or hard plastic surface
(blue curve) and the soft, sample (red curve) describes the deformation
of the tissue sample under the load, which allows the determination
of the sample’s modulus of elasticity (Young’s modulus).
AFM experimental setup:
(A) The main AFM components are a cantilever
with a spherical tip (4.5 μm in diameter bead), a laser source,
a photosensitive photodiode, and a piezoelectric scanner that can
accurately apply compressive force to soft tissues at the nanoscale.
The application of the compressive force measured as a function of
the sample’s position in the Z-direction gives
rise to so-called force vs distance curves (B). The difference between
the cantilever deflection on a stiff glass or hard plastic surface
(blue curve) and the soft, sample (red curve) describes the deformation
of the tissue sample under the load, which allows the determination
of the sample’s modulus of elasticity (Young’s modulus).The bead–tissue contact area during the
tests ranged from
7 to 32 μm2, depending on the depth of indentation.
AFM experiments were made maximally 5 h after biopsy, and tissues
were kept in culture medium during the experiments at 37 °C.
Tissues were glued using cyanoacrylate glue onto a Petri dish and
immersed in DMEM for measurements. To account for cantilever bending,
force curves were first recorded on a rigid plastic substrate, and
then, the rigid surface was replaced by the compliant tissue sample.
Indentations were carried out in multiple places on the tissue surface.
Up to 15 maps consisting of 8 × 8 points corresponding to a scan
area of 10 × 10 μm were made for each sample. Indentations
were carried out in multiple places on the tissue surface, in the
central zone of each tumor sample. The difference between the cantilever
deflection on a rigid surface and the compliant tissue sample describes
the deformation of the tissue under the bead load (Figure B). When the force used for
deformation is plotted against the depth of indentation that this
force induced, so-called force-versus-indentation curves can be obtained.
To determine the elastic modulus (i.e., the Young’s modulus),
we fitted the curves to the Hertz contact model for a sphere using
following formula:where E* is the apparent
Young’s modulus:If Esample ≪ Etip (as is true for living
cells), then can be simplified:Esample is the
Young’s modulus of the tissue, and μsample is the Poisson ratio of the sample, related to the compressibility
of the material[10,41] and assumed to be 0.5 for an
incompressible material, as is true for tissues. Histograms of the
distributions of Young’s modulus values for each sample were
prepared, and the mean values for all healthy and cancer tissues along
with standard deviations were calculated.Macroscopic rheometry,
using a HAAKE Rheostress 6000 rheometer
(Thermo Fisher Scientific, Waltham, MA, USA) fitted with a 20 mm diameter
parallel plate system, was used to measure the viscoelasticity of
tissue samples. The height of the tissue slices ranged from 3 to 5
mm in the uncompressed state. Healthy and colon cancer tissues were
cut into disk-shaped samples using a 20 mm diameter steel punch. To
avoid tissue slippage during the measurements, samples were placed
on sand paper (P800) gaskets inside Petri dishes, and dishes were
glued to the rheometer bottom plate (Figure A). Experiments were also made maximally
5 h after biopsy, and tissues were kept at 37 °C during the experiment.
During the test, a sample hood was used to prevent heat loss and water
evaporation from the tissue. The tests were carried out in the deformation
control system, where the deflection of the upper measuring plate
by the angle φ is converted into shear deformation:where r is the radius of
the plate and h is the gap height between the rheometer’s
plates.
Figure 2
Rheological experimental setup: (A) human colon samples were cut
to a diameter of 20 mm and placed on sand paper. Tension and compression
were caused by applying force in a direction perpendicular to the tissue samples. Sample
stress as a function of axial deformation and time, storage modulus
and loss modulus as a function of axial strain, and axial stress and
shear strain as well as changes of phase shift as a function of shear
strain were measured; (B) viscoelastic behavior of the sample (as
a sinusoidal function versus time with phase shift between them);
(C) stresses applied to the sample: shear forces (in combination with
tissue compression, axial stress) were applied by rotating the upper
plate in a direction parallel to the sample.
Rheological experimental setup: (A) human colon samples were cut
to a diameter of 20 mm and placed on sand paper. Tension and compression
were caused by applying force in a direction perpendicular to the tissue samples. Sample
stress as a function of axial deformation and time, storage modulus
and loss modulus as a function of axial strain, and axial stress and
shear strain as well as changes of phase shift as a function of shear
strain were measured; (B) viscoelastic behavior of the sample (as
a sinusoidal function versus time with phase shift between them);
(C) stresses applied to the sample: shear forces (in combination with
tissue compression, axial stress) were applied by rotating the upper
plate in a direction parallel to the sample.The measured torque M corresponds to the stresses
τ in the sample:In dynamic tests, the course of strain over
time can be presented
aswhere γ0 is the amplitude
and ω = 2πf is the angular frequency
(in our studies f = 1 Hz).Viscoelastic tissues
exhibit mechanical behavior somewhere between
that of a purely viscous and a purely elastic material; therefore,
there is a phase lag in the measured stress in relation to the applied
shear strain. The shear stress as a function of time equals:where
δ is the phase lag between stress
and strain.The complex modulus of viscoelasticity G* is the
ratio of stress and strain:and the
modulus of elasticity and viscosity
can be calculated fromwhere G′
is known
as the modulus of elasticity or storage modulus and G″ is known as the viscous or loss modulus.The moduli
are related to the phase lag angle δ by the relation . For an ideally elastic
material, δ
= 0°, and for an ideally viscous material, δ = 90°
(Figure ).Two
kinds of rheological tests were performed. The first consisted
of oscillating shear deformation of the tissue with 2% constant shear
amplitude (and constant frequency of 1 Hz) and simultaneous application
of uniaxial strain, which was applied by changing the distance between
the parallel plates, i.e., gap height, by lowering (compression) or
lifting (extension) the upper plate in the range of 0–40% of
the sample initial height with a 10% increment. During these tests,
the course of normal stress (determined by dividing the recorded normal
force FN by the surface area of the upper
rheometer plate in contact with the tissue) and shear stress as a
function of time was determined. The second test consisted of oscillating
shear deformation of the sample with variable amplitude ranging from
2% to 20% at a frequency (f) of 1 Hz. At the same
time, the tissue was subjected to constant compression in the range
of 0–40%. In this way, the G′ and G″ moduli were obtained as a function of shear deformation
in compressed and uncompressed states that mimic the range of deformations
expected to occur in vivo. For example, gastric volume after a meal
can change by a factor of 60 by unfolding and stretching of the gastric
wall by as much as ∼160%.[42]
Histopathological Evaluation
Healthy
and cancer tissues were subjected to histopathological scoring to
illustrate mucosa and submucosa architecture. Intestines were fixed
in 10% neutral-buffered formalin for 12 h, dehydrated in an automatic
tissue processor (LEICA, TP1020), and embedded using a paraffin dispenser
(Bio Optica DP500). Tissue orientation was chosen such that positions
where AFM indentation was performed were visible on the microscopic
slide, and the entire thickness of the examined tissue could be assessed.
The paraffin blocks were cut on a microtome (Leica RM2125 RTS) for
1 to 2 μm thick sections, dewaxed, irrigated, and stained with
hematoxylin and eosin (Diapath, Italy) using a dyeing machine (Bio
Optica AUS124). Microscope slides were than evaluated using an OLYMPUS
BX53 microscope.
Statistical Analysis
The significance
of differences was determined using the two-tailed Student’s t test. Statistical analyses were performed using OriginPro
9.65 (OriginLab Corporation, Northampton, MA, USA). p < 0.05 was considered to be statistically significant. Results
are the average from all force curves for each patient sample (Figure F). Overall average
values of Young’s modulus are presented as mean ± SD,
where the mean is the average value for the patient from all curves
and SD is standard deviation.
Figure 3
Young’s modulus values obtained for healthy
and cancer tissues
using AFM indentation: (A–D) The Young’s modulus distributions
for healthy and cancer tissue for each patient. The blue column represents
healthy tissues, and the red column represents cancer tissues: (A)
patient no. 1; (B) patient no. 2; (C) patient no. 3; (D) patient no.
4; (E) Young’s modulus distribution for healthy and cancer
tissue for all patients with fitted probability density function of
the log-normal distribution; (F) average Young’s modulus values
for each patient demonstrating the significantly smaller deformability
of cancerous tissues compared to healthy ones.
Young’s modulus values obtained for healthy
and cancer tissues
using AFM indentation: (A–D) The Young’s modulus distributions
for healthy and cancer tissue for each patient. The blue column represents
healthy tissues, and the red column represents cancer tissues: (A)
patient no. 1; (B) patient no. 2; (C) patient no. 3; (D) patient no.
4; (E) Young’s modulus distribution for healthy and cancer
tissue for all patients with fitted probability density function of
the log-normal distribution; (F) average Young’s modulus values
for each patient demonstrating the significantly smaller deformability
of cancerous tissues compared to healthy ones.
Results and Discussion
For many human
diseases including cancers, histopathology, along
with genetic and molecular tests, is the standard procedure confirming
diagnosis and directing therapy that translates into a patient’s
treatment plan. Any additional method(s) supporting this procedure
might be helpful to obtain more accurate diagnosis. Histopathology
is performed using fixed tissues subjected to specific staining that
allow the determination of the morphology of cells and changes in
the tissue architecture, and using immunohistochemistry, the presence
of specific markers is determined. However, information for the characterization
of tissue mechanical properties is lost when the tissue is fixed after
taking the sample. All biological structures at the organization level
starting from cell organelles through whole cells, tissues, and organs
have specific mechanical properties that can be characterized by rheological
parameters, like elasticity or viscosity, and that affect their functions.
Dysfunction of physiological processes during disease development
usually generates changes in these structures that translate into
changes of cell and tissue mechanical properties.[10,11,18] However, these mechanical properties cannot
be accurately determined when histopathology of fixed or frozen material
is performed.Cancer development is usually associated with
a genetic mutation
causing pathological alterations of the cell cycle and invasive motility.
For most cancers, changes in tissue stroma are also important. In
many cases, tissues stiffen during cancer progression.[19,20,22,26] Quick and precise measurements of stiffness and other rheological
parameters characterizing tissue mechanics, so-called mechanomarkers,
might provide a new means to describe tissue pathology. In recent
years, the development of AFM has provided a new method for nanoscale
characterization of a wide spectrum of biomaterials, including human
tissues,[11,13−15,40,43] and stiffness is the key parameter
to be determined. Solid material stiffness is generally defined as
the resistance to deformation caused by the mechanical force after
applying tension, compression, or shear to the material.[11] Stiffness can be quantified by the corresponding
modulus, such as Young’s modulus (elastic modulus), which is
a quantity that measures a material’s resistance to being deformed
elastically when uniaxial stress is applied.[11,44] The elastic modulus of a material is defined as the slope of its
stress–strain curve in the elastic deformation region.[44] Basic laws of mechanics can be applied to study
the physicochemical properties of biological materials, such as human
healthy and diseased tissues.[21,45] In the course of this
study, we assessed the potential of atomic force microscopy and shear
rheometry to determine the rheological properties of healthy tissue
margins and diseased colon tissues to test the hypothesis that rheological
data might be used to complement the histopathological description
of colon cancer tissue. In our work, we have used colloidal AFM tips
as nanoscale indenters together with Hertz contact mechanics to determine
tissues stiffness with high resolution.
AFM Measurements
AFM testing consisted
of the series of loading–unloading cycles over the tissue’s
surface with a constant force of 1 nN. Figure shows relative values of the Young’s
modulus distributions for 4 patients, where healthy and cancer tissues
were compared.We observed that the values of Young’s
moduli for healthy and diseased tissues using the AFM method are significantly
different and much greater for cancer tissue than for healthy tissue.
These results agree with previous reports[11,13,14,34] that cancer
development and progression are associated with changes in mechano-cellular
phenotype and manifested by changes in tissue stiffness with cancer
tissue being stiffer than the healthy tissue of its origin.Figure A–D
shows distributions of the Young’s modulus values for each
sample. Young’s modulus values of cancer samples are shifted
to higher values, and a significant difference between healthy and
diseased tissues stiffness can be seen. The summary histogram presented
in Figure E shows
the overall results obtained for all samples. The mean Young’s
modulus value and the calculated standard deviation for healthy tissue
is 0.44 ± 0.3 kPa, whereas for cancer tissue, it is 5.80 ±
3.8 kPa (Figure F).
Only one of the healthy tissue samples showed higher stiffness in
comparison to the other healthy samples (patient 1), but still, it
is significantly softer than cancerous tissues (Figure A). This may mean that the process of fibrosis
or neoplastic change could have started earlier than visible changes
in the morphology of tissue cells, especially when histopathological
examination did not confirm the presence of neoplastic changes. Previous
studies clearly indicate that deviations in the stiffness and complex
mechanics of cancerous tissues are closely related to alterations
in the extracellular matrix, which provides structural support for
cells allowing their proliferation, motility, and survival.[46] The distributions of Young’s modulus
values are similar to previously reported data for the stiffness of
humanbreast cancer tissues that support an attempt to provide an
approach for nanomechanical profiling of breast cancer.[34] Stiffness profiles observed in our study for
healthy colon tissue, characterized by a single sharp peak, differ
from the broad distribution of Young’s modulus in colon cancer,
indicative of tissue mechanical heterogeneity in disease. The high
heterogeneity of cancer tissue architecture is reflected in the shape
of the histograms presented in Figure , which are much wider for samples from cancerous than
normal tissues. This difference is also manifested in the error bars
(standard deviations) of the averaged values presented in Figure F. The increase in
tissue stiffness may be associated with extracellular matrix protein
alignment or overexpression, especially of different types of collagens,
increased matrix fibrosis, cross-linking, and vascularization during
cancer progression.[18,26,47] Despite many possible sources of ECM alterations, their overall
contribution to the tissue mechanical properties is significant, and
we propose to use stiffness as a new mechanomarker of colon pathology.On the other hand, we cannot disregard the challenge posed by the
accurate measurement of tissue stiffness that has recently been expressed.[48] It is reasonable to propose that tissue stiffening
is not a simple process that is proportional to the extent of cancer
progression because of the nonlinear mechanical nature of tissues
and the compression stiffening that can arise as pressure gradients
develop in solid tumors. Furthermore, tissue sections from solid tumors
are complex nonlinear materials that can exhibit molecular, cellular,
and architectural alterations, manifested by various mechanical properties
with microranges. When measuring the stiffness map using an atomic
force microscope in a traditional way, we do not know the exact place
where the AFM’s tip contacts the tissue, thus lacking information
about the correlation between the stiffness and local tissue morphology.
The small contact area and relatively low indentation caused by the
AFM tip results in small deformations where only individual cells
or their parts undergo compression. Overall, the AFM strategy, as
a procedure of colon cancer diagnosis, should include a large number
of measurements performed at different locations in order to get a
useful mechanical profile of the examined tissue. As argued clearly
in a previous report,[48] there is currently
no possibility to base the entire colon cancer diagnosis on AFM studies.
Therefore, progress in AFM technology will be required to develop
a method that will offer an accurate test to examine the extraction
of tissue samples collected during biopsy or surgical procedures.
Rheology tests can serve as an additional step to expand histopathological
procedures, especially when facing disputable cases and when microscopic
based diagnosis is unclear. Overall, there is significant potential
for AFM as a device for identification and early cancer grading and
classification.[11,13,48]
Shear Rheometry
In general, rheological
properties of tissues, especially soft ones, might be determined with
higher accuracy using shear rheometry.[14,22,40] In our study, more detailed mechanical properties
of colon cancer tissues subjected to different compression levels
are described. One of the biggest advantage of using shear rheometry
is that it allows the determination of the mechanical properties of
a larger fragment of the 3D tissue and gives average mechanical parameters
for the entire tissue volume, not just for the thin surface layer
of the cells and extracellular matrix that is in contact with the
AFM probe. The large size of the biopsy specimens allows for more
thorough tissue rheological characterization in a strain-controlled
shear rheometer. Figures –6 show the change in axial
stress as a function of axial strain and time, storage modulus (G′) and loss modulus (G″)
as a function of axial strain, and axial stress and shear strain as
well as changes of phase shift as a function of shear strain. Our
first observation using strain rheometry showed, similarly to the
AFM results, that cancer tissues are clearly stiffer than normal ones,
which can be seen as an increase in storage modulus G′ of cancer tissues compared to G′
of normal tissues (G′ for healthy tissue,
1.52 kPa, and G′ for cancer tissue, 9.60 kPa,
for 40% tissue compression). Differences of storage and loss modulus
values between healthy and cancer tissue were clearly evident and
confirm our AFM data and previous studies.[11,13,14,34] Therefore,
we postulate that G′ similarly to Young’s
modulus can be proposed as a promising mechanomarker of colon cancer.
Figure 4
Rheological
properties of the healthy and cancer tissues: (A) stress
(normal force/sample surface) as a function of axial strain; (B) G′ as a function of axial strain (compression); (C,
D) G′ and G″ as a
function of axial stress. Average values for all samples. Blue, healthy
tissue; red, cancer tissue. (E) G″/G′ as a function of axial stress. For both healthy
and cancer samples, the compression stiffening effect is visible,
but cancer tissues have significantly higher storage modulus and stiffen
to a greater extent when compressive force is applied. Cancer tissue
in compression reacts by increasing elasticity more prominently compared
to healthy tissue. At increasing compression, samples become more
elastic and less dissipative, especially cancer tissues.
Figure 6
Axial stress exerted by the tissues at different levels
of uniaxial
compression, from 0 up to 40% with 10% increments. The amount of stress
for healthy (blue) and cancer (red) tissues was calculated by dividing
the normal force (FN) registered by the
force transducer by the sample surface area in contact with the rheometer
plate. (A) Stress–relaxation for cancer and healthy tissues;
(B) stress–relaxation normalized to the initial stress just
before each compression step; (C) stress–relaxation course
for cancer tissue in 20% compression with a two-component stress–relaxation
model: σ(t) = 1 – σ∞(1 + e–), where σ(t) is the reduced relaxation function
and σ∞ and τ are the fitting constants
for the equilibrium modulus and relaxation time constant, respectively;
(D) exponential decay time (t) for cancer and healthy
tissues at different levels of compression.[56]
Rheological
properties of the healthy and cancer tissues: (A) stress
(normal force/sample surface) as a function of axial strain; (B) G′ as a function of axial strain (compression); (C,
D) G′ and G″ as a
function of axial stress. Average values for all samples. Blue, healthy
tissue; red, cancer tissue. (E) G″/G′ as a function of axial stress. For both healthy
and cancer samples, the compression stiffening effect is visible,
but cancer tissues have significantly higher storage modulus and stiffen
to a greater extent when compressive force is applied. Cancer tissue
in compression reacts by increasing elasticity more prominently compared
to healthy tissue. At increasing compression, samples become more
elastic and less dissipative, especially cancer tissues.Figure shows
cancer
and healthy tissues’ mechanical response under different levels
of compression. Figure A shows how the axial stress increases in compression, and Figure B shows parallel
results for the shear storage modulus G′. Figure C,D compares the
shear storage and loss moduli (G′, G″) for normal and cancer tissues, respectively,
and Figure E shows
how the ratio of the two moduli changes in compression.For
the normal sample, the stress–strain relationship is
linear and the apparent Young’s modulus, calculated from the
slope, is 1.9 kPa, consistent with the value obtained by AFM. For
the cancerous tissue, the stress–strain plot is clearly nonlinear
and the local slope increases with increasing strain, a feature that
is characteristic of other fibrotic diseased tissues.[22] Although it is possible to create a regression line with
the coefficient of determination (R2)
similar for both healthy and cancer tissues, the stress values at
40% compression for cancer tissue clearly stand out from the rest.
From the initial slope at strains below 20%, where the tissue response
is approximately linear, the calculated Young’s modulus for
cancer tissues is 18.6 kPa, again consistent with the values measured
by AFM, although somewhat higher, which likely reflects the higher
strains of the macroscopic measurement and the dominance of the stiffest
regions of the sample in the macroscopic response. The order of magnitude
difference between Young’s modulus for normal and diseased
tissues is similar for both means of measurement. The mechanical response
of cancer tissues to compressive stress is more pronounced when compared
to the healthy margin tissue, and it increases with the increasing
level of tissue compression. This observation confirms the values
of G′ as a function of axial deformation.
Additionally, a higher storage modulus for cancer tissue and its increase
with compression were observed. This is also in agreement with results
previously published.[14,22,41,49] Tissue strengthening during compression
was described previously[40] as a compression
stiffening effect.Although plots of axial stress and G′ vs
axial strain are nonlinear for cancer samples, plots of G′ and G″ vs axial stress are linear
for both normal and cancer samples. Storage (G″)
and loss (G″) moduli as functions of axial
stress are shown in Figure C,D. Although both G′ and G″ rise with increasing axial stress, the slopes
of these plots are different. At increasing compression, the human
colon samples become more elastic and less dissipative, as the ratio G″/G′ decreases, as seen
by the smaller slope for G″ than for G′ and also in Figure E. This feature is more evident in cancer tissues.The observed changes in mechanical response of cancer tissues under
axial and tangential forces might allow for the identification of
specific mechanomarkers and, in particular, stiffness, as quantified by a uniaxial or shear elastic modulus, might be
a new mechanomarker of colon cancers and potentially other colon pathology.
On the basis of our results, we confirm that tissue shear
dissipation might also be a new effective marker of cancer
distinct from the changes in magnitude of the elastic modulus. Recent
studies of brain tumors using magnetic resonance elastography suggest
the potential of a dissipative feature of tissue rheology as a new
marker in tissue pathology.[14,50−52]In contrast to the increase in G′
with
increasing uniaxial strain, neither G′ nor G″ increase with increasing shear strain. Figure Column A shows plots
of normalized G′ as a function of increasing
shear strain. G′ decreases with increasing
shear strain for all tissues. Similar shear weakening of liver tissue
has been observed in an earlier study.[22] The colon tissues were subjected to oscillating shear strain in
the range of 2–20% in the uncompressed state and at 20% and
40% uniaxial compression. If the tissue is modeled as a composite
of a fibrous extracellular matrix and cells within the matrix mesh,
the shear modulus is determined by the resistance of the matrix fibers,
the cells, and the interface between them. Reference (53) shows how volume conserving
cells restrict the possible movements of matrix fibers and eliminate
some of the mechanisms by which fibrous networks stiffen in shear.
At the same time, the contacts between the cell and the matrix are
a combination of stable and dynamic bonds. At large strains, some
of these bonds detach, and this effect leads to softening. However,
if the strain is not so large as to damage the tissue, the stable
bonds remain and, then, the shear strain is removed; the sample can
recover its initial state, and the dynamic bonds can reform in the
relaxed state.
Figure 5
Rheological properties of healthy and cancer tissues under
shear
strain: Column A, the normalized storage modulus (G′) as a function of shear strain for different degrees of
axial compression (0%, 20%, and 40%). The decrease of G′ with increasing shear strain depicts strain weakening and
is greater for cancer tissues (red dots). Column B, the phase shift
angle as a function of shear strain for different degrees of axial
compression. Cancerous tissues became significantly more dissipative
at increased shear strain (Column B). G′ was
normalized to the minimum shear strain value.
Rheological properties of healthy and cancer tissues under
shear
strain: Column A, the normalized storage modulus (G′) as a function of shear strain for different degrees of
axial compression (0%, 20%, and 40%). The decrease of G′ with increasing shear strain depicts strain weakening and
is greater for cancer tissues (red dots). Column B, the phase shift
angle as a function of shear strain for different degrees of axial
compression. Cancerous tissues became significantly more dissipative
at increased shear strain (Column B). G′ was
normalized to the minimum shear strain value.Shear strain softening of the tissue samples occurs both with and
without superimposed uniaxial compression. The strongest effect of
shear strain softening was visible for cancer tissue. These results
agree with reports of the mechanical behavior of other tissues under
different levels of shear strain.[22] Stiffening
under axial compression and a decrease of G′
with increasing shear strain is not a universal feature of soft materials
and is not observed for purified extracellular matrices such as fibrin
gel and collagen gel,[49,54] which exhibit strain-stiffening
in shear and softening in compression.[55] The pronounced shear strain softening of cancerous tissue even in
the uncompressed state can also be considered as a possible mechanomarker
of colon pathology. The phase shift angle δ between the oscillation
curves is related to the ratio G″/G′ and is a measure of viscous dissipation within
the material. For a phase shift angle equal to 0°, the material
exhibits ideal elastic behavior, while for 90°, ideal viscous
flow is observed. Figure Column B shows δ as a function of shear strain for
uncompressed as well as 20% and 40% uniaxially compressed tissue samples.
While healthy tissues did not change their dissipative response significantly
with increased shear strain, cancerous tissues became significantly
more dissipative at increased shear strain. In the range of 2–20%
shear stress, the phase shift angle for healthy tissue ranges from
15° (for small strains) to 20° (this decreases with increasing
compression), which indicates a relatively elastic response. In the
case of cancerous tissue, this angle increases from 12.5° for
small deformations up to 41° with 20% shear strain. This change
indicates that tumor tissue becomes more viscous with increasing shear
deformation, and this observation suggests that the phase shift angle
δ can be considered as a new mechanomarker.When the normal
force registered by the force transducer is measured,
the compressive stress exerted by the tissue during compression can
be calculated. Figure shows compressive stress as a function of
time when shear strain was set to 2% over the whole course of an experiment
in which tissues were rapidly compressed at 10% increments. The compressive
stress in cancer tissues (red dots) is much higher than in healthy
tissues (blue dots) at every level of compression. For 10% compression,
the maximum stress in healthy tissue is 252 Pa and in cancer tissue,
2030 Pa, while at 40% compression, the maximum compressive stress
is ∼2 kPa in healthy tissues and ∼21 kPa in cancer tissues.
Both tissues relax with time. A similar phenomenon was seen in ref (22) for healthy and fibrotic
rat liver tissues. The forces exerted in compression are much higher
for ratcancer tissues, but a larger relaxation of the stress has
not been observed.Axial stress exerted by the tissues at different levels
of uniaxial
compression, from 0 up to 40% with 10% increments. The amount of stress
for healthy (blue) and cancer (red) tissues was calculated by dividing
the normal force (FN) registered by the
force transducer by the sample surface area in contact with the rheometer
plate. (A) Stress–relaxation for cancer and healthy tissues;
(B) stress–relaxation normalized to the initial stress just
before each compression step; (C) stress–relaxation course
for cancer tissue in 20% compression with a two-component stress–relaxation
model: σ(t) = 1 – σ∞(1 + e–), where σ(t) is the reduced relaxation function
and σ∞ and τ are the fitting constants
for the equilibrium modulus and relaxation time constant, respectively;
(D) exponential decay time (t) for cancer and healthy
tissues at different levels of compression.[56]Figure C shows
that the stress caused by subsequent lowering of the upper rheometer
plate, after reaching the maximal value, decreases in a nonlinear
way over time. Individual relaxation curves were fitted to a simple
stress–relaxation model σ(t) = 1 –
σ∞(1 + e–), where σ(t) is the
reduced relaxation function and σ∞ and τ
are the fitting constants for the equilibrium modulus and relaxation
time constant, respectively.[56] In compression,
healthy tissues relax slightly more slowly at modest levels of compression
(Figure D). These
observations of compressive stress–relaxation over time do
not provide a clear discrimination between normal and cancerous tissues.The response of tissues under compressive, tensile, and shear force
have recently been modeled by systems based on biopolymers in which
cell-like inclusions are embedded.[53] Understanding
the rheological properties of tissue models can help us to identify
mechanisms by which tissue stiffness is altered in disease and to
assess how these changes lead to cellular dysfunction. The work[53] of van Oosten et al. presents multiple experimental
setups with a combination of nonlinear polymer networks with and without
small elastic particles that mimic cells inside the tissue, and interestingly,
this approach shows that the mechanical behavior of native tissues
cannot be reproduced by biopolymer networks or by particle systems
alone.[53] This study shows that tissue rheology
arises closely from an interaction between the polymer network and
volume-conserving cells within the network. This is especially consequential
for tumor promotion, which is associated with uncontrolled multiplication
of mutated cells, their migratory and invasive potential in response
to external loads, pressure gradients, and changes in ECM mechanoarchitecture.[57,58] While mechanical testing of such complex 3D systems is possible
with AFM and is sufficient to distinguish between healthy and diseased
tissue, bulk rheometer testing seems to be a faster and simpler method
for rheological testing that could support histopathology or even
intraoperative decisions. Although there are other possibilities to
assess tissue neoplastic conditions, such as measurements of the density
of the samples,[59] which are in addition
to standard histopathological methods, rheological properties of cancerous
tissues and comparing them to healthy tissues will provide a broader
and more accurate view of tissue changes in cancer progression. The
rheological examination of tissues removed during surgery is possible,
and with miniaturization of this technique, it might be possible even
if a small tissue volume is obtained.[56] This work provides evidence that rheological examination of tissues
may be a part of new procedures to describe tissue pathology.
Histopathological Evaluation
All
the samples measured by AFM and rheometry underwent histopathological
evaluation. Figure shows representative images of stained healthy colon tissues.
Figure 7
Representative
images of stained colon tissues: (A, C) tumor tissues;
(B, D) healthy colon tissues. (A, B) Sample sections from AFM; (C,
D) Sections from a sample measured in the rheometer. The morphology
of healthy colon tissue shows surface epithelium, mucosa with colon
crypts, goblet cells, lamina propria, and muscularis mucosa and submucosa.
The crypts open to the surface epithelium in this cross section, and
some of the crypts appear partially or below the surface. A section
of cancer tissue from AFM (A) shows the infiltration of adenocarcinoma
with necrosis in the submucosa together with the desmoplastic reaction
and microinflammation. A section of cancer tissue from the rheometer
(C) shows the infiltration of adenocarcinoma with a fibrous inflammation
reaction. The tissue is damaged due to the destructive nature of the
rheometer tests at large strains (red scale bar, 100 μm; blue
scale bar, 200 μm). Large arrows indicate the places from which
the data was collected.
Representative
images of stained colon tissues: (A, C) tumor tissues;
(B, D) healthy colon tissues. (A, B) Sample sections from AFM; (C,
D) Sections from a sample measured in the rheometer. The morphology
of healthy colon tissue shows surface epithelium, mucosa with colon
crypts, goblet cells, lamina propria, and muscularis mucosa and submucosa.
The crypts open to the surface epithelium in this cross section, and
some of the crypts appear partially or below the surface. A section
of cancer tissue from AFM (A) shows the infiltration of adenocarcinoma
with necrosis in the submucosa together with the desmoplastic reaction
and microinflammation. A section of cancer tissue from the rheometer
(C) shows the infiltration of adenocarcinoma with a fibrous inflammation
reaction. The tissue is damaged due to the destructive nature of the
rheometer tests at large strains (red scale bar, 100 μm; blue
scale bar, 200 μm). Large arrows indicate the places from which
the data was collected.These images confirm
that the tissue samples have normal morphology
with the mucosal submucosa layer and the muscularis. Histopathological
evaluation confirmed neoplastic changes in all colon cancer tissues,
with typical changes such as cancer cell infiltration, desmoplastic
reactions, and inflammation. The heterogeneity of cancer tissue structure,
as observed in histopathological images, manifests strongly in the
AFM measurements, where local mechanical properties of the tissue
surface can be measured, whereas bulk rheology measurements using
strain rheometry reflects averaged properties of the whole tissue
sample.
Conclusions
The
analysis of data collected using atomic force microscopy and
shear rheometry revealed that colon cancer tissues have different
mechanical properties compared to the healthy margin of the tissue.
Significantly higher Young’s and shear modulus values for the
cancer samples were observed. This difference was more pronounced
after separating the Young’s modulus, shear storage, and loss
modulus values. Overall, more pronounced compression stiffening of
colon cancer tissue samples was observed. A combination of histopathological
and mechanical tests directed to assess different mechanomarkers might
place tissue rheology as a complementary procedure in the advanced
diagnosis of colon cancer.
Authors: D Krndija; H Schmid; J-L Eismann; U Lother; G Adler; F Oswald; T Seufferlein; G von Wichert Journal: Oncogene Date: 2010-03-08 Impact factor: 9.867
Authors: Renping Zhao; Xiangda Zhou; Essak S Khan; Dalia Alansary; Kim S Friedmann; Wenjuan Yang; Eva C Schwarz; Aránzazu Del Campo; Markus Hoth; Bin Qu Journal: Front Immunol Date: 2021-08-17 Impact factor: 7.561
Authors: Vladislav M Farniev; Mikhail E Shmelev; Nikita A Shved; Valeriia S Gulaia; Arthur R Biktimirov; Alexey Y Zhizhchenko; Aleksandr A Kuchmizhak; Vadim V Kumeiko Journal: Biomedicines Date: 2022-07-19