Literature DB >> 33050715

Impact of Laser Fiber Diameter and Irrigation Fluids on Induced Bubble Stream Dynamics with Thulium Fiber Laser: An In Vitro Study.

Frédéric Panthier1,2,3,4, Steeve Doizi1,2,3, Cyril Gorny3, Laurent Berthe3, Olivier Traxer1,2.   

Abstract

Objectives: The Thulium Fiber Laser (TFL) is studied as an alternative to the holmium:yttrium-aluminium-garnet (Ho:YAG) laser for lithotripsy, with the advantage of an induced bubble stream (IBS). This in vitro study compared the TFL's IBS with 150- and 272 μm-core-diameter laser fiber (CDF) and in four irrigant fluids.
Methods: A TFL of 50 W (IPG Photonics©) and 150 and 272 μm-CDF (Boston Scientific©) were used, comparing nine energies (in the range from 0.025 to 4 J). An experimental setup consisted of a vertically disposed fiber in a cuvette filled with saline, iodinated contrast agent (IOA), human urine, or deionized water (DW) at ambient temperature. High-speed imaging of three consecutive IBS was performed to determine the influence of energy on their maximum length (ML; μm), width (MW; μm), and duration (MD; μs). Fibers were cleaved with ceramic scissors between each experience.
Results: The IBS had higher ML and MW and MD with 150CDF than 272CDF. Maximum pulse rate for 150CDF and 272CDF was 2182 and 2000 Hz, respectively. Every maximum power was higher than the technological limit of TFL (>50 W). At equal energy density, 150CDF was associated with lower dimensions and durations. The IBS had higher maximum dimensions in IOA compared with saline solution (SS). Compared with DW and urine, IBS in IOA were longer beyond 500 mJ. Over 25 mJ, IBS were thinner in DW, urine, and SS. The IBS in DW, urine, and SS had similar maximum dimensions. The IBS's duration was higher in IOA compared with DW, urine, and SS, meaning a lower theoretical maximum pulse rate and power in IOA.
Conclusion: Lasering with 150CDF fits with lower pulse energies-higher pulse rates settings than 272CDF, such as fine dusting mode. In IOA, Induced Bubbles Streams present higher dimensions and durations than in other studied fluids, related to its higher viscosity. Safety distance and pulse rate should be increased and decreased, respectively.

Entities:  

Keywords:  Thulium Fiber Laser; endourology; induced bubble streams; laser; lithotripsy

Mesh:

Substances:

Year:  2020        PMID: 33050715     DOI: 10.1089/end.2020.0766

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  2 in total

1.  Initial clinical experience with the thulium fiber laser from Quanta System: First 50 reported cases.

Authors:  Alba Sierra; Mariela Corrales; Merkourios Kolvatzis; Olivier Traxer
Journal:  World J Urol       Date:  2022-07-21       Impact factor: 3.661

2.  Laser Fiber Displacement Velocity during Tm-Fiber and Ho:YAG Laser Lithotripsy: Introducing the Concept of Optimal Displacement Velocity.

Authors:  Frederic Panthier; Thibault Germain; Cyril Gorny; Laurent Berthe; Steeve Doizi; Olivier Traxer
Journal:  J Clin Med       Date:  2021-12-29       Impact factor: 4.241

  2 in total